Capture of instantaneous temperature in oscillating flows: Use of constant-voltage anemometry to correct the thermal lag of cold wires operated by constant-current anemometry

被引:10
作者
Berson, Arganthael [1 ]
Poignand, Gaelle [2 ]
Blanc-Benon, Philippe [2 ]
Comte-Bellot, Genevieve [2 ]
机构
[1] Queens Univ, Queens RMC Fuel Cell Res Ctr, Kingston, ON K7L 5L9, Canada
[2] Ecole Cent Lyon, CNRS, UMR 5509, Lab Mecan Fluides & Acoust, F-69134 Ecully, France
关键词
anemometry; flow measurement; fluid oscillations; heat transfer; turbulence; HOT-WIRE; GRID TURBULENCE; BOUNDARY-LAYER; TIME-CONSTANT; VELOCITY; SCALAR;
D O I
10.1063/1.3274155
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new procedure for the instantaneous correction of the thermal inertia of cold wires operated by a constant-current anemometer is proposed for oscillating flows. The thermal inertia of cold wires depends both on the wire properties and on the instantaneous incident flow velocity. Its correction is challenging in oscillating flows because no relationship between flow velocity and heat transfer around the wire is available near flow reversal. The present correction procedure requires neither calibration data for velocity nor thermophysical or geometrical properties of the wires. The method relies on the splitting of the time lag of cold wires into two factors, which are obtained using a constant-voltage anemometer in the heated mode. The first factor, which is intrinsic to the wire, is deduced from time-constant measurements performed in a low-turbulence flow. The second factor, which depends on the instantaneous flow velocity, is acquired in situ. In oscillating flows, data acquisition can be synchronized with a reference signal so that the same wire is alternatively operated in the cold mode by a constant-current anemometer and in the heated mode by a constant-voltage anemometer. Validation experiments are conducted in an acoustic standing-wave resonator, for which the fluctuating temperature field along the resonator axis is known independently from acoustic pressure measurements, so that comparisons can be made with cold-wire measurements. It is shown that despite the fact that the wire experiences flow reversal, the new procedure recovers accurately the instantaneous temperature of the flow.
引用
收藏
页数:8
相关论文
共 21 条
[1]  
[Anonymous], 2007, Handbook of experimental fluid mechanics
[2]   DETERMINATION OF TIME CONSTANTS OF COLD WIRES [J].
ANTONIA, RA ;
BROWNE, LWB ;
CHAMBERS, AJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (09) :1382-1385
[3]  
BERSON A, 2007, THESIS ECOLE CENTRAL
[4]   A strategy to eliminate all nonlinear effects in constant-voltage hot-wire anemometry [J].
Berson, Arganthael ;
Blanc-Benon, Philippe ;
Comte-Bellot, Genevieve .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (04)
[5]  
Blanc-Benon P, 2008, AIP CONF PROC, V1022, P359, DOI 10.1063/1.2956230
[6]   A FULLY COMPENSATED HOT COLD WIRE ANEMOMETER SYSTEM FOR UNSTEADY-FLOW VELOCITY AND TEMPERATURE-MEASUREMENTS [J].
BREMHORST, K ;
GRAHAM, LJW .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1990, 1 (05) :425-430
[7]   INFLUENCE OF END CONDUCTION ON SENSITIVITY TO STREAM TEMPERATURE-FLUCTUATIONS OF A HOT-WIRE ANEMOMETER [J].
BREMHORST, K ;
GILMORE, DB .
INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 1978, 21 (02) :145-154
[8]   Lead-resistance effects in a constant voltage anemometer [J].
Comte-Bellot, G ;
Weiss, J ;
Béra, JC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (06) :2075-2081
[9]  
COMTEBELLOT G, 1998, HDB FLUID DYNAMICS, P1
[10]   DYNAMIC HOT-WIRE ANEMOMETER CALIBRATION USING AN OSCILLATING FLOW [J].
ELGER, DF ;
ADAMS, RL .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1989, 22 (03) :166-172