Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol

被引:1
作者
Huang, Qi-Fang [1 ]
Wang, Jie [1 ]
Sheng, Chang-Sheng [1 ]
Zhang, Na-Na [1 ]
Li, Yan [1 ]
Wang, Ji-Guang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Ctr Epidemiol Studies & Clin Trials, Shanghai 200030, Peoples R China
关键词
blood pressure determination; blood pressure monitor; European Society of Hypertension International Protocol; validation studies;
D O I
10.1097/MBP.0b013e328339e28c
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure monitor KD-5915 (ANDON Health, Tianjin, China) for home blood pressure monitoring according to the International Protocol. Method Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese individuals using a mercury sphygmomanometer (two observers) and the KD-5915 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and the remaining 18 participants in phase 2 of the validation study. Data analysis was performed using the ESHIP Analyzer. Results The KD-5915 device successfully passed phase 1 of the validation study with a number of absolute differences between the device and observers within 5, 10, and 15 mmHg for at least 27 of 45, 34 of 45, and 43 of 45 measurements, respectively. The device also achieved the targets for phase 2.1, with 66 of 99, 81 of 99, and 95 of 99 device-observer differences within 5, 10, and 15 mmHg, respectively, for systolic blood pressure, and with 79 of 99, 95 of 99, and 99 of 99 differences within 5, 10, and 15 mmHg, respectively, for diastolic blood pressure. In phase 2.2, 23, and 27 participants had at least two of the three device-observer differences within 5 mmHg (required >= 22) for systolic blood pressure and diastolic blood pressure, respectively. Conclusion The ANDON upper arm blood pressure monitor KD-5915 has passed the International Protocol requirements, and hence can be recommended for use at home in adults. Blood Press Monit 15: 232-234 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:232 / 234
页数:3
相关论文
共 6 条
[1]  
*ASS ADV MED INSTR, 2002, ANS AAMI SP
[2]   Assessment of the validation of blood pressure monitors: a statistical reappraisal [J].
Friedman, Bruce A. ;
Alpert, Bruce S. ;
Osborn, David ;
Prisant, L. Michael ;
Quinn, David E. ;
Seller, John .
BLOOD PRESSURE MONITORING, 2008, 13 (04) :187-191
[3]   European Society of hypertension recommendations for conventional, ambulatory and home blood pressure measurement [J].
O'Brien, E ;
Asmar, R ;
Beilin, L ;
Imai, Y ;
Mallion, JM ;
Mancia, G ;
Mengden, T ;
Myers, M ;
Padfield, P ;
Palatini, P ;
Parati, G ;
Pickering, T ;
Redon, J ;
Staessen, J ;
Stergiou, G ;
Verdecchia, P .
JOURNAL OF HYPERTENSION, 2003, 21 (05) :821-848
[4]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[5]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[6]   European Society of Hypertension International Protocol for the validation of blood pressure monitors: a critical review of its application and rationale for revision [J].
Stergiou, George S. ;
Karpettas, Nikos ;
Atkins, Neil ;
O'Brien, Eoin .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :39-48