Voltammetric Determination of Amoxicillin Using a Reduced Graphite Oxide Nanosheet Electrode

被引:26
作者
Pham, Thi Hai Yen [1 ]
Mai, Thi Trang [2 ]
Nguyen, Hoang Anh [1 ]
Chu, Thi Thu Hien [3 ]
Vu, Thi Thu Ha [1 ,4 ]
Le, Quoc Hung [1 ]
机构
[1] Vietnam Acad Sci Technol, Inst Chem, 18 Hoang Quoc Viet, Hanoi 100000, Vietnam
[2] VNU Univ Sci Hanoi, 19 Le Thanh Tong, Hanoi 100000, Vietnam
[3] Natl Univ Civil Engn NUCE, Dept Chem, 55 Giai Phong, Hanoi 100000, Vietnam
[4] Univ Sci & Technol Hanoi USTH, Vietnam Acad Sci & Technol, 18 Hoang Quoc Viet, Hanoi 100000, Vietnam
关键词
MOLECULARLY IMPRINTED POLYMER; GRAPHENE OXIDE; CARBON NANOTUBES; ELECTROCHEMICAL DETECTION; URIC-ACID; AQUATIC ENVIRONMENT; TRACE DETERMINATION; GOLD NANOPARTICLES; ANTIMICROBIAL USE; ASCORBIC-ACID;
D O I
10.1155/2021/8823452
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A reduced graphite oxide nanosheet electrode (RGOnS) was prepared as a sensor for amoxicillin (AMX) detection, an antibiotic commonly used in the livestock farm, by the square-wave adsorptive stripping voltammetry technique. Graphite oxide with nanosheet shape was produced from a graphite electrode by a chronoamperometry process at 5 V and then an electrochemical reduction process was carried out to form RGOnS with restored long-range conjugated networks and better conductivity. The electrodes were characterized by SEM, EDX, and FTIR spectroscopy. The RGOnS electrode prepared at an optimal reduction potential of -1 V for 120 s exhibits a larger electrochemical active surface area, and the obtained oxidation signal of AMX is approximately ten times higher than that of the pristine graphite electrode. The analytical conditions such as the pH of electrolyte and accumulation time were optimized. The calibration curve built under the optimal conditions provided a good linear relationship in the range of AMX concentration from 0.5-80 mu M with the correlation coefficient of 0.9992. The limit of detection was calculated as 0.193 mu M. Satisfactory results are obtained from the detection of the AMX in different samples using the prepared electrode.
引用
收藏
页数:12
相关论文
共 59 条
[1]  
Ahammad A.J.S., 2019, Biomedical Applications of Graphene and 2D Nanomaterials, P249, DOI [10.1016/B978-0-12-815889-0.00012-X, DOI 10.1016/B978-0-12-815889-0.00012-X]
[2]   Removal of caffeine, nicotine and amoxicillin from (waste)waters by various adsorbents. A review [J].
Anastopoulos, Ioannis ;
Pashalidis, Ioannis ;
Orfanos, Alexios G. ;
Manariotis, Ioannis D. ;
Tatarchuk, Tetiana ;
Sellaoui, Lotfi ;
Bonilla-Petriciolet, Adrian ;
Mittal, Alok ;
Nunez-Delgado, Avelino .
JOURNAL OF ENVIRONMENTAL MANAGEMENT, 2020, 261
[3]  
[Anonymous], 1999, Paediatr Child Health, V4, P321
[4]   Environmentally friendly preparation of exfoliated graphite [J].
Asghar, H. M. A. ;
Hussain, S. N. ;
Sattar, H. ;
Brown, N. W. ;
Roberts, E. P. L. .
JOURNAL OF INDUSTRIAL AND ENGINEERING CHEMISTRY, 2014, 20 (04) :1936-1941
[5]   Trace determination of P-lactam antibiotics in environmental aqueous samples using off-line and on-line preconcentration in capillary electrophoresis [J].
Bailon-Perez, M. L. ;
Garcia-Campana, A. M. ;
Cruces-Blanco, C. ;
Iruela, M. del Olmo .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1185 (02) :273-280
[6]  
Bard A. J., 2000, ELECTROCHEMICAL METH, V2nd
[7]   Rapid and on-site simultaneous electrochemical detection of copper, lead and mercury in the Amazon river [J].
Bernalte, Elena ;
Arevalo, Sebastian ;
Perez-Taborda, Jaime ;
Wenk, Jannis ;
Estrela, Pedro ;
Avila, Alba ;
Di Lorenzo, Mirella .
SENSORS AND ACTUATORS B-CHEMICAL, 2020, 307
[8]   Pharmaceuticals in the aquatic environment - a comparison of risk assessment strategies [J].
Bound, JP ;
Voulvoulis, N .
CHEMOSPHERE, 2004, 56 (11) :1143-1155
[9]   Electrochemical detection of glucose at physiological pH using gold nanoparticles deposited on carbon nanotubes [J].
Branagan, David ;
Breslin, Carmel B. .
SENSORS AND ACTUATORS B-CHEMICAL, 2019, 282 :490-499
[10]   Amoxicillin on polyglutamic acid composite three-dimensional graphene modified electrode: Reaction mechanism of amoxicillin insights by computational simulations [J].
Chen, Chenglong ;
Lv, Xuchu ;
Lei, Wu ;
Wu, Yi ;
Feng, Shasha ;
Ding, Yong ;
Lv, Jingjing ;
Hao, Qingli ;
Chen, Shen-Ming .
ANALYTICA CHIMICA ACTA, 2019, 1073 :22-29