Validation of a high-throughput in vitro alkaline elution/rat hepatocyte assay for DNA damage

被引:9
作者
Gealy, Robert
Wright-Bourque, Jennifer L.
Kraynak, Andrew R.
McKelvey, Troy W.
Barnum, John E.
Storer, Richard D. [1 ]
机构
[1] Merck Res Labs, Dept Lab Sci & Invest Toxicol, West Point, PA 19486 USA
[2] Merck Res Labs, Dept Vaccine & Biol Res, West Point, PA 19486 USA
关键词
alkaline elution; automation; cytotoxicity; genotoxicity; hepatocyte; validation;
D O I
10.1016/j.mrgentox.2007.01.005
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
In vitro alkaline elution is a sensitive and specific short term assay which measures DNA strand breakage in a mammalian test system (primary rat hepatocytes). This lab has previously demonstrated the performance of the assay with known genotoxic and nongenotoxic compounds. The methodology employed has relatively low sample throughput and is labor-intensive, requiring a great deal of manual processing of samples in a format that is not amenable to automation. Here, we present an automated version of the assay. This high-throughput alkaline elution assay (HT-AE) was made possible through 3 key developments: (1) DNA quantitation using PicoGreen (R) and OliGreen (R) fluorescent DNA binding dyes; (2) design and implementation of a custom automation system; and (3) reducing the assay to a 96-well plate format. The assay can now be run with 5-50 mg of test compound. HT-AE was validated in a similar manner as the original assay, including assessment of non-genotoxic and non-carcinogenic compounds and evaluation. of cytotoxicity to avoid confounding effects of toxicity-associated DNA degradation. The validation test results from compounds of known genotoxic potential were used to set appropriate criteria to classify alkaline elution results for genotoxicity. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:49 / 63
页数:15
相关论文
共 50 条
  • [31] Automated high-throughput Vibrio fischeri assay for (eco)toxicity screening: Application to ionic liquids
    Pinto, Paula C. A. G.
    Costa, Susana P. F.
    Lima, Jose L. F. C.
    Saraiva, M. Lucia M. F. S.
    ECOTOXICOLOGY AND ENVIRONMENTAL SAFETY, 2012, 80 : 97 - 102
  • [32] Development and validation of a high-throughput high-performance liquid chromatographic assay for the determination of caffeine in food samples using a monolithic column
    Tzanavaras, Paraskevas D.
    Themelis, Demetrius G.
    ANALYTICA CHIMICA ACTA, 2007, 581 (01) : 89 - 94
  • [33] A HIGH-THROUGHPUT CLINICAL ASSAY FOR TESTING DRUG FACILITATION OF EXPOSURE THERAPY
    Rodebaugh, Thomas L.
    Levinson, Cheri A.
    Lenze, Eric J.
    DEPRESSION AND ANXIETY, 2013, 30 (07) : 631 - 637
  • [34] Development of a Novel, Automated High-Throughput Device for Performing the Comet Assay
    Karbaschi, Mahsa
    Ji, Yunhee
    Mujawar, Mubarak A.
    Mendoza, Mario
    Marquez, Juan S.
    Sonawane, Apurva
    Shah, Pratikkumar
    Ross, Chris
    Bhansali, Shekhar
    Cooke, Marcus S.
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2023, 24 (08)
  • [35] Quality Assessment and Validation of High-Throughput Sequencing for Grapevine Virus Diagnostics
    Soltani, Nourolah
    Stevens, Kristian A.
    Klaassen, Vicki
    Hwang, Min-Sook
    Golino, Deborah A.
    Al Rwahnih, Maher
    VIRUSES-BASEL, 2021, 13 (06):
  • [36] Synthetic images of high-throughput microscopy for validation of image analysis methods
    Lehmussola, Antti
    Ruusuvuori, Pekka
    Selinummi, Jyrki
    Rajala, Tiina
    Yli-Harja, Olli
    PROCEEDINGS OF THE IEEE, 2008, 96 (08) : 1348 - 1360
  • [37] Validation of High-Throughput Genotoxicity Assay Screening Using γH2AX In-Cell Western Assay on HepG2 Cells
    Khoury, Laure
    Zalko, Daniel
    Audebert, Marc
    ENVIRONMENTAL AND MOLECULAR MUTAGENESIS, 2013, 54 (09) : 737 - 746
  • [38] High-throughput analysis of DNA interstrand crosslinks in human peripheral blood mononuclear cells by automated reverse FADU assay
    Debiak, Malgorzata
    Panas, Alicja
    Steinritz, Dirk
    Kehe, Kai
    Buerkle, Alexander
    TOXICOLOGY, 2011, 280 (1-2) : 53 - 60
  • [39] Validation of a Pseudovirus Neutralization Assay for Severe Acute Respiratory Syndrome Coronavirus 2: A High-Throughput Method for the Evaluation of Vaccine Immunogenicity
    Cai, Zhaohui
    Kalkeri, Raj
    Wang, Mi
    Haner, Benjamin
    Dent, Dominic
    Osman, Bahar
    Skonieczny, Paul
    Ross, Jeremy
    Feng, Sheau-Line
    Cai, Rongman
    Zhu, Mingzhu
    Cloney-Clark, Shane
    Plested, Joyce S.
    MICROORGANISMS, 2024, 12 (06)
  • [40] Automation of the ELISpot assay for high-throughput detection of antigen-specific T-cell responses
    Almeida, Coral-Ann M.
    Roberts, Steven G.
    Laird, Rebecca
    McKinnon, Elizabeth
    Ahmed, Imran
    Pfafferott, Katja
    Turley, Joanne
    Keane, Niamh M.
    Lucas, Andrew
    Rushton, Ben
    Chopra, Abha
    Mallal, Simon
    John, Mina
    JOURNAL OF IMMUNOLOGICAL METHODS, 2009, 344 (01) : 1 - 5