Does the ProSeal laryngeal mask airway prevent aspiration of regurgitated fluid?

被引:142
作者
Keller, C
Brimacombe, J [1 ]
Kleinsasser, A
Loeckinger, A
机构
[1] Cairns Base Hosp, Dept Anaesthesia & Intens Care, Cairns 4870, Australia
[2] Leopold Franzens Univ, Dept Anaesthesia & Intens Care Med, Innsbruck, Austria
基金
中国国家自然科学基金;
关键词
D O I
10.1097/00000539-200010000-00046
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In this randomized, cross-over cadaver study, we determined whether a new airway device, the ProSeal laryngeal mask airway (PLMA; Laryngeal Mask Company, Henley-on-Thames, UK), prevents aspiration of regurgitated fluid. We studied five male and five female cadavers (6-24 h postmortem). The infusion set of a pressure/controlled, continuous flow pump was inserted into the upper esophagus and ligated into place. Esophageal pressure (EP) was increased in 2-cm H2O increments. This was performed without an airway device (control) and over a range of cuff volumes (0-40 mL) for the classic laryngeal mask airway (LMA), the PLMA with the drainage tube clamped (PLMA clamped) and unclamped (PLMA unclamped). The EP at which fluid was first seen with a fiberoptic scope in the hypopharynx (control), above or below the cuff, or in the drainage tube, was noted. Mean EP at which fluid was seen without any airway device was 9 (range 8-10) cm H2O. EP at which fluid was seen was always higher for the PLMA clamped and LMA compared with the control tall, P < 0.0001). The mean EP at which fluid was seen for the PLMA unclamped was similar to the control at 10 (range 8-13) cm H2O. For the PLMA unclamped, fluid appeared from the drainage tube in all cadavers at 10-40 mt cuff volume and in 8 of 10 cadavers at zero cuff volume. Mean EP at which fluid was seen above the cuff was similar for the PLMA clamped and LMA at 0-30 mt cuff volume, but was higher for PLMA clamped at 40-mL cuff volume (81 vs 48 cm H2O, P = 0.006). Mean EP at which fluid was seen below the cuff was similar at 0-10 mt cuff volume, but was higher for the PLMA clamped at 20, 30, and 40 mt cuff volume (62, 68, 73 vs 46, 46, 46 cm H2O, respectively, P < 0.04). For the PLMA clamped and the LMA, fluid appeared simultaneously above and below the cuff at all cuff volumes. We concluded that in the cadaver model, the correctly placed PLMA allows fluid in the esophagus to bypass the pharynx and mouth when the drainage tube is open. Both the LMA, and PLMA with a closed drainage tube, attenuate Liquid flow between the esophagus and pharynx. This may have implications for airway protection in unconscious patients.
引用
收藏
页码:1017 / 1020
页数:4
相关论文
共 8 条
[1]   Prevention of tracheal aspiration in a patient with a high risk of regurgitation using a new double-lumen gastric laryngeal mask airway [J].
Agro, F ;
Brain, A ;
Gabbrielli, A ;
Alloni, R ;
Pellegrino, P ;
Cataldo, R ;
Costamagna, G ;
Carassiti, M .
GASTROINTESTINAL ENDOSCOPY, 1997, 46 (03) :257-258
[2]  
Brimacombe J, 1996, ANAESTHESIA, V51, P602
[3]   The ProSeal laryngeal mask airway - A randomized, crossover study with the standard laryngeal mask airway in paralyzed, anesthetized patients [J].
Brimacombe, J ;
Keller, C .
ANESTHESIOLOGY, 2000, 93 (01) :104-109
[4]   Cricoid pressure [J].
Brimacombe, JR ;
Berry, AM .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1997, 44 (04) :414-425
[5]   The influence of the tonsillar gag on efficacy of seal, anatomic position, airway patency, and airway protection with the flexible laryngeal mask airway:: A randomized, cross-over study of fresh adult cadavers [J].
Brimacombe, JR ;
Keller, C ;
Gunkel, AR ;
Pühringer, F .
ANESTHESIA AND ANALGESIA, 1999, 89 (01) :181-186
[6]   Do laryngeal mask airway devices attenuate liquid flow between the esophagus and pharynx?: A randomized, controlled cadaver study [J].
Keller, C ;
Brimacombe, J ;
Rädler, C ;
Pühringer, F .
ANESTHESIA AND ANALGESIA, 1999, 88 (04) :904-907
[7]   Segmental cervical spine motion during orotracheal intubation of the intact and injured spine with and without external stabilization [J].
Lennarson, PJ ;
Smith, D ;
Todd, MM ;
Carras, D ;
Sawin, PD ;
Brayton, J ;
Sato, Y ;
Traynelis, VC .
JOURNAL OF NEUROSURGERY, 2000, 92 (02) :201-206
[8]   REGURGITATION AND ESOPHAGEAL RUPTURE WITH CRICOID PRESSURE - A CADAVER STUDY [J].
VANNER, RG ;
PRYLE, BJ .
ANAESTHESIA, 1992, 47 (09) :732-735