X-ray tomography using a CMOS area detector

被引:1
作者
Brunetti, A.
Cesareo, R.
机构
[1] Univ Sassari, Dept Math & Phys, I-07100 Sassari, Italy
[2] Ist Nazl Fis Nucl, Sez Cagliari, Cagliari, Italy
关键词
flat panel; imaging; tomography;
D O I
10.1016/j.nimb.2007.02.067
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A flat panel based on CMOS technology represents a valid alternative to other kinds of flat panels and to ccd detectors for X-ray imaging. Although the spatial resolution of the ccd sensors is better than that of a CMOS sensor, the last has a larger sensitive-area and it can work at room temperature reaching a dynamic performance comparable to that of a cooled ccd sensor. Other kinds of flat panels, such as TFT screen are much more expensive and they have lower spatial resolution and higher noise than the CMOS detector. In this paper, an application of the CMOS sensor to X-ray tomography is described. Preliminary results are reported and discussed. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:485 / 489
页数:5
相关论文
共 10 条
  • [1] [Anonymous], 1988, PRINCIPLES COMPUTERI
  • [2] BONSE U, 2001, DEV XRAY TOMOGRAPH 3, V4503
  • [3] BONSE U, 2004, DEV XRAY TOMOGRAPH 4, V5535
  • [4] Computed tomography in the fatigue test analysis of an asphaltic mixture
    Braz, D
    da Motta, LMG
    Lopes, RT
    [J]. APPLIED RADIATION AND ISOTOPES, 1999, 50 (04) : 661 - 671
  • [5] Visualization of monomer and polymer inside porous stones by using X-ray tomography
    Brunetti, A
    Princi, E
    Vicini, S
    Pincin, S
    Bidali, S
    Mariani, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 222 (1-2) : 235 - 241
  • [6] 90° Compton and Rayleigh measurements and imaging
    Cesareo, R
    Balogun, F
    Brunetti, A
    Borlino, CC
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 339 - 342
  • [7] ATTENUATION COEFFICIENTS AND TOMOGRAPHIC MEASUREMENTS FOR SOIL IN THE ENERGY-RANGE 10-300 KEV
    CESAREO, R
    DEASSIS, JT
    CRESTANA, S
    [J]. APPLIED RADIATION AND ISOTOPES, 1994, 45 (05) : 613 - 620
  • [8] Internal elemental microanalysis combining x-ray fluorescence, Compton and transmission tomography
    Golosio, B
    Simionovici, A
    Somogyi, A
    Lemelle, L
    Chukalina, M
    Brunetti, A
    [J]. JOURNAL OF APPLIED PHYSICS, 2003, 94 (01) : 145 - 156
  • [9] HAMAMATSU, 2005, KACC1097E06 DN
  • [10] A large-area CMOS imager as an X-ray detector for synchrotron radiation experiments
    Yagi, N
    Yamamoto, M