Nuclear Magnetic Resonance Imaging in Patients With Cardiac Pacing Devices

被引:23
作者
Buendia, Francisco [1 ]
Sanchez-Gomez, Juan M. [1 ]
Sancho-Tello, Maria J. [1 ]
Olaguee, Jose [1 ]
Osca, Joaquin [1 ]
Cano, Oscar [1 ]
Arnau, Miguel A. [1 ]
Igual, Begona [1 ]
机构
[1] Hosp Univ La Fe, Unidad Arritmias, Valencia, Spain
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 2010年 / 63卷 / 06期
关键词
Pacemakers; Magnetic resonance imaging; Safety; IMPLANTABLE-CARDIOVERTER-DEFIBRILLATORS; AMERICAN-COLLEGE; MR SAFETY; RADIOLOGY; PACEMAKERS;
D O I
10.1016/S0300-8932(10)70167-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Currently, nuclear magnetic resonance imaging is contraindicated in patients with a pacemaker or implantable cardioverter-defibrillator. This study was carried out because the potential risks in this situation need to be clearly defined. This prospective study evaluated clinical and electrical parameters before and after magnetic resonance imaging was performed in 33 patients (five with implantable cardioverter-defibrillators and 28 with pacemakers). In these patients, magnetic resonance imaging was considered clinically essential. There were no clinical complications. There was a temporary communication failure in two cases, sensing errors during imaging in two cases, and a safety signal was generated in one pacemaker at the maximum magnetic resonance frequency and output level. There were no technical restrictions on imaging nor were there any permanent changes in the performance of the cardiac pacing device.
引用
收藏
页码:735 / 739
页数:5
相关论文
共 12 条
  • [1] Is magnetic resonance imaging safe in cardiac pacemaker recipients?
    Del Ojo, JL
    Moya, F
    Villalba, J
    Sanz, O
    Pavón, R
    Garcia, D
    Pastor, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04): : 274 - 278
  • [2] Non-invasive imaging - Contraindications to magnetic resonance imaging
    Dill, T.
    [J]. HEART, 2008, 94 (07) : 943 - 948
  • [3] Outcome of magnetic resonance imaging (MRI) in selected patients with implantable cardioverter defibrillators (ICDs)
    Gimbel, JR
    Kanal, E
    Schwartz, KM
    Wilkoff, BL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04): : 270 - 273
  • [4] American college of radiology white paper on MR safety: 2004 update and revisions
    Kanal, E
    Borgstede, JP
    Barkovich, AJ
    Bell, C
    Bradley, WG
    Etheridge, S
    Felmlee, JP
    Froelich, JW
    Hayden, J
    Kaminski, EM
    Lester, JW
    Scoumis, EA
    Zaremba, LA
    Zinninger, MD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (05) : 1111 - 1114
  • [5] Safety of magnetic resonance imaging in patients with cardiovascular devices - An American Heart Association scientific statement from the committee on diagnostic and interventional cardiac catheterization, council on clinical cardiology, and the council on cardiovascular radiology and intervention
    Levine, Glenn N.
    Gomes, Antoinette S.
    Arai, Andrew E.
    Bluemke, David A.
    Flamm, Scott D.
    Kanal, Emanuel
    Manning, Warren J.
    Martin, Edward T.
    Smith, J. Michael
    Wilke, Norbert
    Shellock, Frank S.
    [J]. CIRCULATION, 2007, 116 (24) : 2878 - 2891
  • [6] Clinical applications of cardiovascular magnetic resonance imaging
    Marcu, Constantin B.
    Beek, Aernout M.
    van Rossum, Albert C.
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 2006, 175 (08) : 911 - 917
  • [7] Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla
    Martin, ET
    Coman, JA
    Shellock, FG
    Pulling, CC
    Fair, R
    Jenkins, K
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) : 1315 - 1324
  • [8] Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable- cardioverter defibrillators at 1.5 tesla
    Nazarian, Saman
    Roguin, Ariel
    Zviman, Menekhem M.
    Lardo, Albert C.
    Dickfeld, Timm L.
    Calkins, Hugh
    Weiss, Robert G.
    Berger, Ronald D.
    Bluemke, David A.
    Halperin, Henry R.
    [J]. CIRCULATION, 2006, 114 (12) : 1277 - 1284
  • [9] SAFETY OF PATIENTS WITH MEDICAL DEVICES DURING APPLICATION OF MAGNETIC-RESONANCE METHODS
    POHOST, GM
    BLACKWELL, GG
    SHELLOCK, FG
    [J]. ANNALS OF THE NEW YORK ACADEMY OF SCIENCES-SERIES, 1992, 649 : 302 - 312
  • [10] Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices
    Roguin, Ariel
    Schwitter, Juerg
    Vahlhaus, Christian
    Lombardi, Massimo
    Brugada, Josep
    Vardas, Panos
    Auricchio, Angelo
    Priori, Silvia
    Sommer, Torsten
    [J]. EUROPACE, 2008, 10 (03): : 334 - 344