Ratio measurements in oxygen determinations: wavelength ratiometry, lifetime discrimination, and polarization detection

被引:21
作者
Kostov, Y [1 ]
Rao, G [1 ]
机构
[1] Univ Maryland Baltimore Cty, Dept Chem & Biochem Engn, Baltimore, MD 21250 USA
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2003年 / 90卷 / 1-3期
关键词
optical oxygen sensors; radiometric sensor; polarization sensor;
D O I
10.1016/S0925-4005(03)00040-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work we present ratiometric measurements of oxygen concentrations using a dual-emitting dye. Three different methods for ratio measurements in oxygen sensing are employed. The standard wavelength-ratiometric approach is simple to implement, but the sensor response has significant non-linearity. The frequency discrimination of the lifetimes allows for measurements of dual-emission ratio through a single emission filter. In this case the Stern-Volmer plots are linear. Another option is to use emission polarization. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:139 / 142
页数:4
相关论文
共 10 条