An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma

被引:6
|
作者
Zhang, Qiaoxuan [1 ,2 ]
Han, Liqiao [1 ]
Zheng, Songbai [3 ]
Ouyang, Fen [1 ]
Wu, Xiaobin [1 ]
Yan, Jun [1 ]
Zhan, Min [1 ]
Ke, Peifeng [1 ]
Zhuang, Junhua [1 ]
Huang, Xianzhang [1 ]
机构
[1] Guangzhou Univ Chinese Med, Affiliated Hosp 2, Dept Lab Med, Guangzhou 510120, Guangdong, Peoples R China
[2] Guangzhou Univ Chinese Med, Clin Med Coll 2, Guangzhou 510120, Guangdong, Peoples R China
[3] Fujian Huayin Med Lab Ctr, Xiamen 361101, Fujian, Peoples R China
关键词
Aldosterone; Reference measurement procedure; Tandem mass spectrometry; Clinical laboratory measurements;
D O I
10.1007/s00216-021-03405-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Accurate quantitation of aldosterone is clinically important in standardized testing for primary aldosteronism. The results are often variable when performed by clinical immunoassays. To standardize and ensure the accuracy of clinical systems, reference measurement procedures (RMPs) with higher metrological order are required. A simple and reliable isotope dilution LC-IDMS/MS-based measurement procedure for human plasma aldosterone has been developed. This method involved plasma spiked with a deuterium-labelled internal standard, equilibrated for 0.5 h, and extracted by liquid-liquid extraction (LLE) without derivatization. Aldosterone and its structural analogues were baseline separated with a C18-packed UHPLC column with gradient elution within 7 min. The signal intensity variability and measurement imprecision were reduced by bracketing calibration during plasma aldosterone value assignment. The limit of detection (LoD) was 19.4 pmol/L with a signal-to-noise ratio (S/N) > 3. The lowest limit of quantification (LLoQ) was 27.7 pmol/L (S/N > 10 and CV < 10.0%). LLE was performed with 1 mL of n-hexane/ethyl acetate (3:2, v/v), and the extraction recovery was determined to be 92.15 +/- 3.54%. The imprecisions were <= 3.18% for samples at 124.8, 867.0, and 2628.5 pmol/L. The recoveries were 98.11-101.61%. The relative bias between this candidate RMP and the established RMP was 2.76-1.89%. The linearity response ranged from 27.7 to 2774.4 pmol/L with R-2 = 0.999. The method performance met the requirements of RMPs (<= 5% total CV and <= 3% bias). Furthermore, the developed method was applied to evaluate immunoassays through 41 patient sample comparisons. The calibration and measurement capability (CMC) of this method were also evaluated by measuring these samples. The candidate RMP can serve as an accurate reference baseline for routine methods and can be used for value assignment for reference materials.
引用
收藏
页码:4471 / 4481
页数:11
相关论文
共 50 条
  • [21] An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of gabapentin in human serum and plasma
    Salzmann, Linda
    Wild, Janik
    Singh, Neeraj
    Schierscher, Tobias
    Liesch, Franziska
    Bauland, Friederike
    Geistanger, Andrea
    Risch, Lorenz
    Geletneky, Christian
    Seger, Christoph
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (11) : 1955 - 1966
  • [22] An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of primidone in human serum and plasma
    Schierscher, Tobias
    Salzmann, Linda
    Singh, Neeraj
    Fischer, Vanessa
    Kobel, Anja
    Bauland, Friederike
    Geistanger, Andrea
    Risch, Lorenz
    Geletneky, Christian
    Seger, Christoph
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (07) : 1327 - 1338
  • [23] Evaluation and clinical application of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for serum theophylline
    Du, Yuanyuan
    Yan, Qiaofang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Zhang, Pengwei
    Yan, Jun
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (22) : 4897 - 4906
  • [24] An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of methotrexate in human serum and plasma
    Engel, Anett
    Ruhe, Lena
    Singh, Neeraj
    Wright, Jo Anne
    Liesch, Franziska
    Bauland, Friederike
    Ostermann, Annika I. I.
    Sumalowitsch, Tamara
    Schweinsberg, Vincent J. T.
    Geistanger, Andrea
    Hegel, Johannes Kolja
    Geletneky, Christian
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (11) : 1917 - 1929
  • [25] Development and evaluation of a candidate reference measurement procedure for the determination of progesterone in human serum using isotope-dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    ANALYTICAL CHEMISTRY, 2006, 78 (18) : 6628 - 6633
  • [26] A reference measurement procedure for amino acids in blood using isotope dilution ultra-performance liquid chromatography-tandem mass spectrometry
    Kim, Juok
    Tran, Thi Thanh Huong
    Hong, Seon-Pyo
    Jeong, Ji-Seon
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2017, 1055 : 72 - 80
  • [27] Isotope-Dilution Liquid Chromatography-Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum
    Botelho, Julianne Cook
    Shacklady, Christopher
    Cooper, Hans C.
    Tai, Susan S. -C.
    Van Uytfanghe, Katleen
    Thienpont, Linda M.
    Vesper, Hubert W.
    CLINICAL CHEMISTRY, 2013, 59 (02) : 372 - 380
  • [28] Simultaneous measurement of amiodarone and desethylamiodarone in human plasma and serum by stable isotope dilution liquid chromatography-tandem mass spectrometry assay
    Kuhn, Joachim
    Goetting, Christian
    Kleesiek, Knut
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2010, 51 (01) : 210 - 216
  • [29] A simple and reliable measurement procedure for determination of glycocholic acid in human serum by isotope-dilution liquid chromatography-tandem mass spectrometry
    Wang, Ahui
    Xu, Yuzhu
    Ou, Yangfen
    Han, Liqiao
    Zhang, Qiaoxuan
    Zhan, Min
    Liu, Hongcan
    Ke, Peifeng
    Yan, Jun
    Huang, Xianzhang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2021, 464
  • [30] Development and evaluation of a reference measurement procedure for the determination of estradiol-17β in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry
    Tai, SSC
    Welsh, MJ
    ANALYTICAL CHEMISTRY, 2005, 77 (19) : 6359 - 6363