Usefulness of remote monitoring for the early detection of back-up mode in implantable cardioverter defibrillators

被引:2
|
作者
Guedon-Moreau, Laurence [1 ,2 ]
Finat, Loic [1 ]
Klein, Cedric [1 ]
Kouakam, Claude [1 ]
Marquie, Christelle [1 ]
Klug, Didier [1 ,2 ]
Potelle, Charlotte [1 ]
Ninni, Sandro [1 ,2 ]
Brigadeau, Francois [1 ]
Mirabel, Xavier [3 ]
Lacroix, Dominique [1 ,2 ]
机构
[1] Lille Univ Hosp, Dept Cardiovasc Med, Heart & Lung Inst, F-59037 Lille, France
[2] Lille Univ, Fac Med, F-59045 Lille, France
[3] Ctr Oscar Lambret, Ctr Canc, F-59000 Lille, France
关键词
Remote monitoring; Implantable cardioverter defibrillator; Telemedicine; Back-up mode; Radiotherapy; HEART-FAILURE PATIENTS; DEVICE FOLLOW-UP; MANAGEMENT; ICD; TRIAL; TIME; LEAD;
D O I
10.1016/j.acvd.2020.11.008
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. - Reversion of an implantable cardioverter defibrillator (ICD) to back-up mode degrades the operating capabilities of the device, puts patients at risk and requires rapid intervention by a manufacturer's technician. Aim. - To illustrate the usefulness of remote monitoring of ICDs for the early detection of reversion to back-up mode. Methods. - In our centre, all patients implanted with an ICD, with or without resynchronisation, were offered remote monitoring as soon as the technology became available. Alerts triggered by the remote monitoring system were included prospectively in a register. During a mean follow-up of 5.7 +/- 1.3 years, a total of 1594 patients with an ICD (441 with resynchronisation function) followed with remote monitoring were included in the register. Results. - Among 15,874 alerts, only 10 were related to a reversion to back-up mode. Among those, seven reversions were caused by radiotherapy, two were fake events and one was caused by magnetic resonance imaging. Except for the two fake events, the eight other patients had an emergency admission for the resetting and reprogramming of their ICD. None of the reversion to back-up mode alerts was followed by a clinical alert (i.e. a shock alert) before the ICD problem was resolved. Conclusions. - Reversion to back-up mode is a very rare event, accounting for 0.06% of total alerts; remote monitoring facilitates the early detection of this critical event to resolve the problem faster than the next scheduled follow-up. Remote monitoring can prevent serious dam-age to the patient and avoids systematic ambulatory control of the ICD after each radiotherapy session. (c) 2021 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:287 / 292
页数:6
相关论文
共 50 条
  • [31] Home monitoring of implantable cardioverter-defibrillators: interpretation reliability of the second-generation "IEGM Online" system
    Naegele, Herbert
    Lipoldova, Jolana
    Oswald, Hanno
    Klein, Gunnar
    Elvan, Arif
    Vester, Ernst
    Bauer, Wolfgang
    Bondke, Hansjuergen
    Reif, Sebastian
    Daub, Claudia
    Menzel, Frank
    Schrader, Juergen
    Zach, Goeran
    EUROPACE, 2015, 17 (04): : 584 - 590
  • [32] Validation of an Organizational Management Model of Remote Implantable Cardioverter-Defibrillator Monitoring Alerts
    Guedon-Moreau, Laurence
    Finat, Loic
    Boule, Stephane
    Wissocque, Ludivine
    Marquie, Christelle
    Brigadeau, Francois
    Kouakam, Claude
    Mondesert, Blandine
    Kacet, Salem
    Klug, Didier
    Lacroix, Dominique
    CIRCULATION-CARDIOVASCULAR QUALITY AND OUTCOMES, 2015, 8 (04): : 403 - 412
  • [33] Patient perspective and safety of remote monitoring of implantable cardioverter-defibrillators in the Polish Nationwide Multicenter Registry: the Medtronic CareLink network evaluation
    Maciag, Aleksander
    Mitkowski, Przemyslaw
    Mazurek, Michal
    Kazmierczak, Jaroslaw
    Nowak, Krzysztof
    Grabowski, Marcin
    Oreziak, Artur
    Kempa, Maciej
    Bacior, Bogumila
    Gepner, Katarzyna
    Chmielewska-Michalak, Lidia
    Lenarczyk, Radoslaw
    Kiedrowicz, Radoslaw
    Fuglewicz, Artur
    Cacko, Andrzej
    Szwed, Hanna
    KARDIOLOGIA POLSKA, 2020, 78 (11) : 1115 - 1121
  • [34] Automatic home monitoring of implantable cardioverter defibrillators
    Nielsen, Jens Cosedis
    Kottkamp, Hans
    Zabel, Markus
    Aliot, Etienne
    Kreutzer, Ulrich
    Bauer, Alexander
    Schuchert, Andreas
    Neuser, Hans
    Schumacher, Burghard
    Schmidinger, Herwig
    Stix, Guenter
    Clementy, Jacques
    Danilovic, Dejan
    Hindricks, Gerhard
    EUROPACE, 2008, 10 (06): : 729 - 735
  • [35] Remote monitoring of pacemakers and implantable cardioverter-defibrillators, safety aspects and study in cardiac implant patients on EMF exposure
    Plawiak-Mowna, Anna
    Krawczyk, Andrzej
    PRZEGLAD ELEKTROTECHNICZNY, 2011, 87 (12B): : 137 - 140
  • [36] Cost-effectiveness of remote monitoring of implantable cardioverter-defibrillators in France: a meta-analysis and an integrated economic model derived from randomized controlled trials
    Sequeira, Saannya
    Jarvis, Christopher, I
    Benchouche, Akram
    Seymour, Jerome
    Tadmouri, Abir
    EUROPACE, 2020, 22 (07): : 1071 - 1082
  • [37] Trends in utilization and spending on remote monitoring of pacemakers and implantable cardioverter-defibrillators among Medicare beneficiaries
    Holtzman, Jessica N.
    Wadhera, Rishi K.
    Choi, Eunhee
    Zhao, Tianyi
    Secemsky, Eric A.
    Fraiche, Ariane M.
    Shen, Changyu
    Kramer, Daniel B.
    HEART RHYTHM, 2020, 17 (11) : 1917 - 1921
  • [38] Remote monitoring for follow-up of patients with implantable cardiac devices
    Oliveira, Mario
    Cunha, Pedro Silva
    da Silva, Nogueira
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2013, 32 (03) : 185 - 190
  • [39] Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems
    Werner Jung
    Andreas Rillig
    Ralf Birkemeyer
    Tomislav Miljak
    Udo Meyerfeldt
    Journal of Interventional Cardiac Electrophysiology, 2008, 23 : 73 - 85
  • [40] Follow-up of implantable cardioverter-defibrillators: Face-to-face or remote?