Comprehensive Echocardiographic Assessment of Normal Mitral Medtronic Hancock II, Medtronic Mosaic, and Carpentier-Edwards Perimount Bioprostheses Early after Implantation

被引:19
作者
Blauwet, Lori A.
Malouf, Joseph F.
Connolly, Heidi M.
Hodge, David O.
Evans, Katie N.
Herges, Regina M.
Sundt, Thoralf M., III
Miller, Fletcher A., Jr. [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN 55905 USA
关键词
Doppler; Echocardiography; Mitral; Prosthesis; VELOCITY INTEGRAL RATIO; PROSTHESIS PERFORMANCE INDEX; DOPPLER-ECHOCARDIOGRAPHY; HEMODYNAMIC PERFORMANCE; VALVE-REPLACEMENT; PORCINE BIOPROSTHESIS; CLINICAL-PERFORMANCE; CONTINUITY EQUATION; TIME; MISMATCH;
D O I
10.1016/j.echo.2010.03.031
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Normal Doppler-derived hemodynamic data for mitral valve bioprostheses are limited. Methods: To establish parameters for identifying normal function for each of the 3 types of bioprostheses examined, we conducted a comprehensive, retrospective, two-dimensional, and Doppler echocardiographic assessment of 179 patients who underwent implantation of the Medtronic Hancock II or the Medtronic Mosaic (Medtronic, Inc, Minneapolis, MN) porcine mitral valve bioprosthesis or the Carpentier-Edwards Perimount (Edwards Lifesciences LLC, Irvine, CA) bovine pericardial mitral valve bioprosthesis. Results: All bioprostheses were normal by clinical examination, intraoperative transesophageal echocardiography, and postoperative transthoracic echocardiography. Regardless of valve type and body surface area, the pressure half-time was < 124 ms in all patients. Mean gradient < 9.5 mm Hg, mitral E velocity < 2.6 m/s, mitral valve prosthesis time-velocity integral < 69 cm, and ratio of the mitral valve prosthesis time-velocity integral to the left ventricular outflow tract time-velocity integral < 3.4 were recorded in nearly all patients. Conclusion: These cutoff values (mean + 2 standard deviation) are specific, but not sensitive, for identifying mitral valve prosthesis dysfunction. Prostheses with hemodynamic values that are higher than these cutoff values are likely dysfunctional, but in select cases, mitral valve prosthesis dysfunction may be present even when hemodynamic values are lower than these thresholds. (J Am Soc Echocardiogr 2010; 23: 656-66.)
引用
收藏
页码:656 / 666
页数:11
相关论文
共 20 条
[1]   DOPPLER-ECHOCARDIOGRAPHIC ASSESSMENT WITH THE CONTINUITY EQUATION OF ST-JUDE MEDICAL MECHANICAL PROSTHESES IN THE MITRAL-VALVE POSITION [J].
BITAR, JN ;
LECHIN, ME ;
SALAZAR, G ;
ZOGHBI, WA .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (04) :287-293
[2]   Doppler echocardiography of 79 normal CarboMedics mitral prostheses: A comprehensive assessment including time-velocity integral ratio and prosthesis performance index [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Herges, Regina M. ;
Sundt, Thoralf M., III ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2007, 20 (10) :1125-1130
[3]   Doppler Echocardiography of 240 Normal Carpentier-Edwards Duraflex Porcine Mitral Bioprostheses: A Comprehensive Assessment Including Time Velocity Integral Ratio and Prosthesis Performance Index [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Evans, Katie N. ;
Herges, Regina M. ;
Sundt, Thoralf M., III ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (04) :388-393
[4]   CLINICAL AND HEMODYNAMIC ASSESSMENT OF THE HANCOCK-II BIOPROSTHESIS [J].
DAVID, TE ;
ARMSTRONG, S ;
SUN, Z .
ANNALS OF THORACIC SURGERY, 1992, 54 (04) :661-668
[5]   VALIDATION AND APPLICATIONS OF MITRAL PROSTHETIC VALVULAR AREAS CALCULATED BY DOPPLER ECHOCARDIOGRAPHY [J].
DUMESNIL, JG ;
HONOS, GN ;
LEMIEUX, M ;
BEAUCHEMIN, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (22) :1443-1448
[6]   VALVE PROSTHESIS HEMODYNAMICS AND THE PROBLEM OF HIGH TRANSPROSTHETIC PRESSURE-GRADIENTS [J].
DUMESNIL, JG ;
YOGANATHAN, AP .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 :S34-S38
[7]   European experience with the Mosaic bioprosthesis [J].
Eichinger, WB ;
Botzenhardt, F ;
Gunzinger, R ;
Kemkes, BM ;
Sosnowski, A ;
Maïza, D ;
Coto, EO ;
Bleese, N .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (02) :333-339
[8]  
Fernandes V, 2002, AM J CARDIOL, V89, P704, DOI 10.1016/S0002-9149(01)02343-8
[9]   Short-term hemodynamic performance of the mitral Carpentier-Edwards PERIMOUNT pericardial valve [J].
Firstenberg, MS ;
Morehead, AJ ;
Thomas, JD ;
Smedira, NG ;
Cosgrove, DM ;
Marchand, MA .
ANNALS OF THORACIC SURGERY, 2001, 71 (05) :S285-S288
[10]  
Fradet G, 2004, J HEART VALVE DIS, V13, P239