Validation of planar and tomographic radionuclide ventriculography by a dynamic ventricular phantom

被引:6
作者
De Bondt, P
Vandenberghe, S
De Mey, S
Segers, P
De Winter, O
De Sutter, J
Van de Wiele, C
Verdonck, P
Dierckx, RA
机构
[1] Univ Hosp Ghent, Div Nucl Med, B-9000 Ghent, Belgium
[2] Univ Ghent, Hydraul Lab, Ghent, Belgium
[3] Univ Hosp Ghent, Dept Cardiol, B-9000 Ghent, Belgium
关键词
validation; tomographic radionuclide ventriculography; dynamic phantom;
D O I
10.1097/00006231-200307000-00006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Although there is increasing interest in the automatic processing of tomographic radionuclide ventriculography (TRV) studies, validation is mainly limited to a comparison of TRV results with data from planar radionuclide ventriculography (PRV) or gated perfusion. single photon emission computed tomography (SPECT). The aim of this study was to use a dynamic physical cardiac phantom to validate the ejection fraction (EF) and volumes from PRV and TRV studies. A new dynamic left ventricular phantom was constructed and used to obtain 21 acquisitions in the planar and tomographic mode. The directly measured volumes and EFs of the phantom during the acquisitions were considered as the gold standard for comparison with TRV and PRV. EFs were calculated from PRV by background-corrected end-diastolic and end-systolic frames. Volumes and EFs were calculated from TRV by region growing with different lower thresholds to search for the optimal threshold. EF from PRV correlated significantly with the real EF (r = 0.94, P = 0.00). The optimal threshold value for volume calculation from TRV in 336 cases was 50% (r = 0.98, P = 0.00) yielding the best slope after linear regression. When considering these calculated end-diastolic and end-systolic volumes, EF correlated well (r=0.99, P=0.00) with the real EF, and this correlation was significantly (P = 0.04) higher than that of the EF from PRV. Our experiments prove that EF measured by TRV yields more accurate results compared with PRV in dynamic cardiac phantom studies. ((C) 2003 Lippincott Williams Wilkins).
引用
收藏
页码:771 / 777
页数:7
相关论文
共 15 条
[1]  
Chin BB, 1997, J NUCL MED, V38, P942
[2]   TOMOGRAPHIC GATED BLOOD POOL RADIONUCLIDE VENTRICULOGRAPHY - ANALYSIS OF WALL MOTION AND LEFT-VENTRICULAR VOLUMES IN PATIENTS WITH CORONARY-ARTERY DISEASE [J].
CORBETT, JR ;
JANSEN, DE ;
LEWIS, SE ;
GABLIANI, GI ;
NICOD, P ;
FILIPCHUK, NG ;
REDISH, GA ;
AKERS, MS ;
WOLFE, CL ;
RELLAS, JS ;
PARKEY, RW ;
WILLERSON, JT .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (02) :349-358
[3]   Updated imaging guidelines for nuclear cardiology procedures, part 1 [J].
DePuey, EG ;
Garcia, EV .
JOURNAL OF NUCLEAR CARDIOLOGY, 2001, 8 (01) :G3-G58
[4]  
FABER TL, 1991, J NUCL MED, V32, P2311
[5]  
FEARNOW EC, 1985, J NUCL MED, V26, P1042
[6]   INVIVO QUANTITATION USING SPECT OF RADIOPHARMACEUTICAL UPTAKE BY HUMAN MENINGIOMAS [J].
FRONT, D ;
IOSELEVSKY, G ;
FRENKEL, A ;
EVENSAPIR, E ;
ISRAEL, O ;
KOLODNY, GM ;
FEINSUD, M .
RADIOLOGY, 1987, 164 (01) :93-96
[7]   SPECT VOLUME QUANTITATION - INFLUENCE OF SPATIAL-RESOLUTION, SOURCE SIZE AND SHAPE, AND VOXEL SIZE [J].
KING, MA ;
LONG, DT ;
BRILL, AB .
MEDICAL PHYSICS, 1991, 18 (05) :1016-1024
[8]   MEASUREMENT OF ABSOLUTE LEFT-VENTRICULAR VOLUME FROM GATED BLOOD POOL STUDIES [J].
LINKS, JM ;
BECKER, LC ;
SHINDLEDECKER, JG ;
GUZMAN, P ;
BUROW, RD ;
NICKOLOFF, EL ;
ALDERSON, PO ;
WAGNER, HN .
CIRCULATION, 1982, 65 (01) :82-91
[9]   Semi-automatic segmentation of gated blood pool emission tomographic images by watersheds: application to the determination of right and left ejection fractions [J].
Mariano-Goulart, D ;
Collet, H ;
Kotzki, PO ;
Zanca, M ;
Rossi, M .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (09) :1300-1307
[10]  
Pretorius PH, 1997, J NUCL MED, V38, P1528