Optimized spiral imaging for measurement of myocardial T2 relaxation

被引:70
作者
Foltz, WD
Al-Kwifi, O
Sussman, MS
Stainsby, JA
Wright, GA
机构
[1] Sunnybrook & Womens Coll, Hlth Sci Ctr, Imaging Res Program, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[3] GE Med Syst, ASL West, Menlo Pk, CA USA
关键词
myocardial T2; spinal imaging; oximetry; BOLD-MRI;
D O I
10.1002/mrm.10467
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Microcirculation oxygen levels and blood volumes should be reflected in measurements of myocardial T-2 relaxation. This work describes the optimization of a spiral imaging strategy for robust myocardial T-2 measurement to minimize the standard deviation of T-2 measurement (sigmaT(2)). Theoretical and experimental studies of blurring at muscle/blood interfaces enabled the derivation of parameter sets which reduce sigma T-2 to the level of 5%. T-2 relaxation mapping within healthy volunteers provided estimation of residual sigmaT(2) within the optimized technique. The standard deviation in T-2 measurement across regions of interest (ROIs) in different locations is about 9%. The standard deviation in T-2 measurement in an ROI across different time points is about 5%. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:1089 / 1097
页数:9
相关论文
共 30 条
[1]   OXYGENATION IN THE RABBIT MYOCARDIUM - ASSESSMENT WITH SUSCEPTIBILITY-DEPENDENT MR-IMAGING [J].
ATALAY, MK ;
FORDER, JR ;
CHACKO, VP ;
KAWAMOTO, S ;
ZERHOUNI, EA .
RADIOLOGY, 1993, 189 (03) :759-764
[2]  
Bauer WR, 1999, MAGN RESON MED, V41, P51, DOI 10.1002/(SICI)1522-2594(199901)41:1<51::AID-MRM9>3.0.CO
[3]  
2-G
[4]   CORONARY ANGIOGRAPHY WITH MAGNETIZATION-PREPARED T-2 CONTRAST [J].
BRITTAIN, JH ;
HU, BS ;
WRIGHT, GA ;
MEYER, CH ;
MACOVSKI, A ;
NISHIMURA, DG .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) :689-696
[5]   STUDIES OF GD-DTPA RELAXIVITY AND PROTON-EXCHANGE RATES IN TISSUE [J].
DONAHUE, KM ;
BURSTEIN, D ;
MANNING, WJ ;
GRAY, ML .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (01) :66-76
[6]  
FOLTZ W, 1998, P 6 SCI M ISMRM SYDN, V3, P2168
[7]   T-2 accuracy on a whole-body imager [J].
Foltz, WD ;
Stainsby, JA ;
Wright, GA .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :759-768
[8]   Vasodilator response assessment in porcine myocardium with magnetic resonance relaxometry [J].
Foltz, WD ;
Huang, H ;
Fort, S ;
Wright, GA .
CIRCULATION, 2002, 106 (21) :2714-2719
[9]  
Foltz WD, 1999, MAGN RESON MED, V42, P837, DOI 10.1002/(SICI)1522-2594(199911)42:5<837::AID-MRM3>3.0.CO
[10]  
2-8