Instrument to Enhance Visualization of Descemet Membrane During Graft Preparation for DMEK Surgery

被引:2
作者
Menzel-Severing, Johannes [1 ]
Salla, Sabine [2 ]
Plum, Wolfgang Joachim [2 ]
Tourtas, Theofilos [1 ]
Fuchsluger, Thomas [1 ]
Schloetzer-Schrehardt, Ursula [1 ]
Kruse, Friedrich Eduard [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Ophthalmol, Schwabachanlage 6, D-91054 Erlangen, Germany
[2] Rhein Westfal TH Aachen, Dept Ophthalmol, Aachen, Germany
关键词
cornea; Descemet membrane; DMEK; endothelial keratoplasty; corneal transplantation; ENDOTHELIAL KERATOPLASTY; TRANSPLANTATION; RISK;
D O I
10.1097/ICO.0000000000000805
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:Descemet membrane (DM) endothelial keratoplasty has improved outcomes of corneal transplantation in patients with corneal endothelial disease. However, the procedure has been criticized for jeopardizing donor tissue during graft preparation. Standardization of this procedure may provide a way toward minimizing tissue loss. For this purpose, we propose the use of a novel tool.Methods:Computerized numerical control milling was used to create a blunt instrument, which was used to remove endothelial cells within a defined area in the periphery of donor corneas. Trypan blue was used to stain denuded DM. Graft preparation was continued as per our standard protocol. Transmission electron microscopy was performed on the treated area, and endothelial cell counts were obtained.Results:Use of the modified procedure resulted in delineation of a peripheral band of denuded DM, which readily stained with trypan blue. This provided increased visibility of DM during subsequent steps. Transmission electron microscopy confirmed that no structural deficits of DM were induced. Mean endothelial cell loss (SD) at 24 hours after preparation was 63 (+/- 130) cells per square millimeter in the group prepared with the use of the new instrument (n = 7), versus 116 (+/- 107) cells per square millimeter in the group prepared without the new instrument (n = 7; P = 0.45).Conclusions:The device presented here enhances visualization of DM during creation of the peripheral margin for subsequent lifting of the margin and stripping of the graft. This may increase success rates and shorten preparation times and learning periods for DM preparation. DM ultrastructure and endothelial cells were not negatively affected.
引用
收藏
页码:904 / 907
页数:4
相关论文
共 50 条
  • [21] Influence of Descemet Membrane Endothelial Keratoplasty Graft Preparation Patterns on Postoperative Clinical Outcome
    Schrittenlocher, Silvia
    Matthaei, Mario
    Rokohl, Alexander C.
    Franklin, Jeremy
    Bachmann, Bjoern
    Cursiefen, Claus
    CORNEA, 2023, 42 (08) : 940 - 945
  • [22] Lamellar Dissection Technique for Descemet Membrane Endothelial Keratoplasty Graft Preparation
    Tan, Tien-En
    Devarajan, Kavya
    Seah, Xin-Yi
    Lin, Shu-Jun
    Lim, Gary Swee
    Cajucom-Uy, Howard Yu
    Ang, Marcus Han Nian
    Tan, Donald Tiang Hwee
    CORNEA, 2020, 39 (01) : 23 - 29
  • [23] The need of rebubbling in case of small graft detachments after Descemet Membrane Endothelial Keratoplasty (DMEK)
    Kramer, Nane
    Unterlauft, Jan Darius
    Girbardt, Christian
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2023, 33 (03) : 1347 - 1353
  • [24] Incidence and Outcomes of Cystoid Macular Edema after Descemet Membrane Endothelial Keratoplasty (DMEK) and DMEK Combined with Cataract Surgery
    Ching, Geoffrey
    Covello, Albert T.
    Bae, Steven S.
    Holland, Simon
    McCarthy, Martin
    Ritenour, Rusty
    Iovieno, Alfonso
    Yeung, Sonia N.
    CURRENT EYE RESEARCH, 2021, 46 (05) : 678 - 682
  • [25] Endothelium-in versus endothelium-out for Descemet membrane endothelial keratoplasty graft preparation and implantation
    Parekh, Mohit
    Ruzza, Alessandro
    Ferrari, Stefano
    Ahmad, Sajjad
    Kaye, Stephen
    Ponzin, Diego
    Romano, Vito
    ACTA OPHTHALMOLOGICA, 2017, 95 (02) : 194 - 198
  • [26] 3D Visualization System in Descemet Membrane Endothelial Keratoplasty (DMEK): A Six-Month Comparison with Conventional Microscope
    Morelli, Alberto
    Ferrandina, Rosangela
    Favuzza, Eleonora
    Cennamo, Michela
    Mencucci, Rita
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (15)
  • [27] Standardizing Descemet Membrane Endothelial Keratoplasty Graft Preparation Method in the Eye Bank-Experience of 527 Descemet Membrane Endothelial Keratoplasty Tissues
    Parekh, Mohit
    Baruzzo, Mattia
    Favaro, Elisa
    Borroni, Davide
    Ferrari, Stefano
    Ponzin, Diego
    Ruzza, Alessandro
    CORNEA, 2017, 36 (12) : 1458 - 1466
  • [28] Yogurt Technique for Descemet Membrane Endothelial Keratoplasty Graft Preparation: Early Clinical Outcomes
    Moramarco, Antonio
    Romano, Vito
    Modugno, Rocco Luigi
    Coco, Giulia
    Viola, Pietro
    Fontana, Luigi
    CORNEA, 2023, 42 (01) : 27 - 31
  • [29] A novel device to visualize Descemet membrane during donor preparation for Descemet membrane endothelial keratoplasty
    Fogla, Rajesh
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2021, 69 (06) : 1609 - 1613
  • [30] Descemet's Membrane Endothelial Keratoplasty (DMEK) : Over-stripping the graft bed promotes donor adherence
    Straiko, Michael
    Terry, Mark
    Talajic, Julia
    Davis-Boozer, David
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)