Characterization of Electrical Reconnection Following Pulmonary Vein Isolation Using First- and Second-Generation Cryoballoon

被引:18
|
作者
Koektuerk, Buelent [1 ]
Yorgun, Hikmet [1 ]
Koektuerk, Oezlem [1 ]
Turan, Cem H. [1 ,2 ]
Keskin, Kerim [1 ]
Schoett, Moritz [1 ]
Turan, Ramazan G. [1 ]
Gorr, Eduard [1 ]
Hoppe, Christian [1 ]
Horlitz, Marc [1 ]
机构
[1] Univ Witten Herdecke, Dept Cardiol Electrophysiol, Krankenhaus Porz Rhein, Cologne, Germany
[2] Univ Hosp Rostock, Dept Internal Med, Div Cardiol, Rostock, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2016年 / 39卷 / 05期
关键词
pulmonary vein reconnection; cryoballoon ablation; pulmonary vein isolation; atrial fibrillation; ABLATION; CATHETER; EFFICACY; OUTCOMES;
D O I
10.1111/pace.12834
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Second-generation cryoballoon (CB-2) is associated with improved outcomes for pulmonary vein isolation (PVI) compared to first generation (CB-1). However, data regarding the predictors of pulmonary vein (PV) electrical reconnection are limited. In this study, we aimed to characterize the predilection sites and predictors of reconnection in patients with recurrent atrial tachyarrhythmia (ATa) after PVI using CB-1 and CB-2. Methods: A total of 59 patients (mean age: 62 11 years and 66% male) with recurrent ATa after previous CB-PVI, using either a 28-mm CB-1 or CB-2, underwent repeat ablation. PV reisolation was performed by irrigated radiofrequency ablation using three-dimensional electroanatomical mapping systems. Results: Electrical PV reconnection was detected in 10 of 11 (91%) of CB-1 patients compared to 41 of 48 (85%) of CB-2 patients. Time to redo procedure after index CB-1 was 8.9 +/- 10.2 months and 11.2 +/- 7.0 months in CB-2. Bonus freeze was applied in all patients with CB-1 and 41% of the patients with CB-2. Superior quadrants of both superior PVs and inferior quadrants of the both inferior PVs exhibited higher predilection for conduction compared to other quadrants (P < 0.001). Multivariate binary logistic analysis revealed that right inferior PV (RIPV; odds ratio [OR]: 1.52, 95% confidence interval [CI]: 1.09-2.13, P = 0.014) and minimum temperature (OR: 1.09, 95% CI: 1.03-1.15, P = 0.004) were the independent predictors of electrical reconnection after CB-2 ablation. Conclusion: Conduction gaps after CB-1 and CB-2 were higher in inferior PVs compared to superior PVs. The RIPV and minimum CB temperature were independent predictors of PV electrical reconnection after CB-2.
引用
收藏
页码:434 / 442
页数:9
相关论文
共 50 条
  • [1] Spontaneous and Adenosine-Induced Pulmonary Vein Reconnection After Cryoballoon Ablation with the Second-Generation Device
    Ciconte, Giuseppe
    Chierchia, Gian-Battista
    De Asmundis, Carlo
    Sieira, Juan
    Conte, Giulio
    Julia, Justo
    Di Giovanni, Giacomo
    Wauters, Kristel
    Baltogiannis, Giannis
    Saitoh, Yukio
    Mugnai, Giacomo
    Catanzariti, Domenico
    Tondo, Claudio
    Brugada, Pedro
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (08) : 845 - 851
  • [2] Predictors of hemoptysis in the setting of pulmonary vein isolation using the second-generation cryoballoon
    Wei, Hui-Qiang
    Guo, Xiao-Gang
    Zhou, Gong-Bu
    Sun, Qi
    Yang, Jian-Du
    Luo, Bin
    Ma, Jian
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (07) : 958 - 965
  • [3] Improved In Vivo Performance of Second-Generation Cryoballoon for Pulmonary Vein Isolation
    Coulombe, Nicolas
    Paulin, Jaime
    Su, Wilber
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (08) : 919 - 925
  • [4] Pulmonary vein isolation with real-time pulmonary vein potential recording using second-generation cryoballoon: Procedural and biophysical predictors of acute pulmonary vein reconnection
    Wei, Hui-Qiang
    Guo, Xiao-Gang
    Zhou, Gong-Bu
    Sun, Qi
    Liu, Xu
    Yang, Jian-Du
    Luo, Bin
    Ma, Jian
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (01): : 14 - 21
  • [5] Improved Procedural Efficacy of Pulmonary Vein Isolation Using the Novel Second-Generation Cryoballoon
    Fuernkranz, Alexander
    Bordignon, Stefano
    Schmidt, Boris
    Gunawardene, Melanie
    Schulte-Hahn, Britta
    Urban, Verena
    Bode, Frank
    Nowak, Bernd
    Chun, Julian K. R.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (05) : 492 - 497
  • [6] The slope of the initial temperature drop predicts acute pulmonary vein isolation using the second-generation cryoballoon
    Deubner, Nikolas
    Greiss, Harald
    Akkaya, Ersan
    Zaltsberg, Sergey
    Hain, Andreas
    Berkowitsch, Alexander
    Guettler, Norbert
    Kuniss, Malte
    Neumann, Thomas
    EUROPACE, 2017, 19 (09): : 1470 - 1477
  • [7] Circulatory Dynamics During Pulmonary Vein Isolation Using the Second-Generation Cryoballoon
    Kajiyama, Takatsugu
    Miyazaki, Shinsuke
    Watanabe, Tomonori
    Yamao, Kazuya
    Kusa, Shigeki
    Igarashi, Miyako
    Nakamura, Hiroaki
    Hachiya, Hitoshi
    Iesaka, Yoshito
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2017, 6 (10):
  • [8] Incidence and characteristics of phrenic nerve palsy following pulmonary vein isolation with the second-generation as compared with the first-generation cryoballoon in 360 consecutive patients
    Fuernkranz, Alexander
    Bordignon, Stefano
    Schmidt, Boris
    Perrotta, Laura
    Dugo, Daniela
    De lazzari, Manuel
    Schulte-Hahn, Britta
    Nowak, Bernd
    Chun, Julian K. R.
    EUROPACE, 2015, 17 (04): : 574 - 578
  • [9] Pulmonary vein isolation in patients with a left common pulmonary vein: Comparison between second-generation cryoballoon and radiofrequency ablation
    Yamaguchi, Masao
    Miyazaki, Shinsuke
    Kajiyama, Takatsugu
    Hada, Masahiro
    Nakamura, Hiroaki
    Hachiya, Hitoshi
    Iesaka, Yoshito
    JOURNAL OF CARDIOLOGY, 2019, 73 (3-4) : 292 - 298
  • [10] Autonomic nervous system modulation and clinical outcome after pulmonary vein isolation using the second-generation cryoballoon
    Miyazaki, Shinsuke
    Nakamura, Hiroaki
    Taniguchi, Hiroshi
    Hachiya, Hitoshi
    Kajiyama, Takatsugu
    Watanabe, Tomonori
    Igarashi, Miyako
    Ichijo, Sadamitsu
    Hirao, Kenzo
    Iesaka, Yoshito
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (09) : 1015 - 1020