In vitro studies on the effect of delaminated a-C:H film fragments on bone marrow cell cultures

被引:27
作者
Bruinink, A [1 ]
Schroeder, A
Francz, G
Hauert, R
机构
[1] EMPA, Swiss Fed Labs Mat Testing & Res, St Gallen, Switzerland
[2] Hesper Ltd, Basel, Switzerland
[3] EMPA, Swiss Fed Labs Mat Testing & Res, Dubendorf, Switzerland
[4] Univ Basel, URZ, Basel, Switzerland
关键词
particles; bone marrow cells; a-C : H; titanium; biocompatibility;
D O I
10.1016/j.biomaterials.2004.09.030
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Amorphous hydrogenated carbon (a-C:H) films have many outstanding properties required for a protective coating material on load bearing medical implants. Recently, titanium doped a-C:H films have been evaluated regarding their effects on bone marrow cell cultures. But many materials that are well-tolerated in bulk form are able to induce toxic reaction if present particulate form. In order to further assess biocompatibility aspects of these two coatings, film delamination has been mimicked in exposure to fluids. In the present study, particles from a-C:H, a-C:H/Ti and a-C:H-a-C:H/Ti bilayer films were added to bone marrow cell cultures in vitro. The results showed that plain a-C:H and to a certain extent a-CH/Ti particles were inert. Both kinds of particles did not significantly stimulate the osteoclast-related enzyme tartrate resistant acid phosphatase (TRAP). A slight increase in cell proliferation and total culture TRAP was found in cultures treated by a-C:H-a-C:H/Ti bilayer films. Latter effect can probably be traced back by the relative high percentage of small particles of a size of around 2 mum. However, if corrected by the cell number also no differences between particle-treated and untreated control cultures could be found, indicating the absence of a toxic effect from delaminated a-C:H coatings. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:3487 / 3494
页数:8
相关论文
共 57 条
  • [21] HAUERT R, 2004, IN PRESS TRIBOL INT
  • [22] Hirayama T, 2001, J Orthop Sci, V6, P53, DOI 10.1007/s007760170025
  • [23] Jacobs JJ, 2001, CLIN ORTHOP RELAT R, P71
  • [24] KOBZIK L, 1993, J IMMUNOL, V151, P2753
  • [25] Reduction of wear in total hip replacement prostheses by amorphous diamond coatings
    Lappalainen, R
    Selenius, M
    Anttila, A
    Konttinen, YT
    Santavirta, SS
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 66B (01) : 410 - 413
  • [26] Liao HH, 1999, J BIOMED MATER RES, V47, P220, DOI 10.1002/(SICI)1097-4636(199911)47:2<220::AID-JBM12>3.3.CO
  • [27] 2-0
  • [28] Lu Luo, 1993, Bio-Medical Materials and Engineering, V3, P223
  • [29] MANIATOPOULOS C, 1988, CELL TISSUE RES, V254, P317
  • [30] MCCOLL IR, 1993, DIAM RELAT MATER, V3, P83