A nano-carbon electrode optimized for adsorptive stripping voltammetry: Application to detection of the stimulant selegiline in authentic saliva

被引:35
作者
dos Santos, Wallans T. P. [1 ,2 ]
Amin, Hatem M. A. [2 ]
Compton, Richard G. [2 ]
机构
[1] Univ Fed Vales Jequitinhonha & Mucuri, Dept Pharm, Campus JK, BR-39100000 Diamantina, MG, Brazil
[2] Univ Oxford, Dept Chem, Phys & Theoret Chem Lab, South Parks Rd, Oxford OX1 3QZ, England
关键词
Adsorptive stripping voltammetry; Carbon Black; Doping control; Saliva; Selegiline; ORAL FLUID; CYCLIC VOLTAMMETRY; NANOTUBES; PLASMA; PHARMACOKINETICS; VOLUNTEERS; URINE; DRUGS;
D O I
10.1016/j.snb.2018.10.037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A multilayer Carbon Black (CB) modified Glassy Carbon Electrode (GCE) is proposed, developed and optimized for Adsorptive Stripping Voltammetry (AdSV). The thick (up to 1000 monolayers) multilayer CB modification offers a significant enhancement in sensitivity for AdSV but not diffusive voltammetry. The detection of the stimulant selegiline, which is a prohibited drug for athletes in sports competitions, is used as model for application of the proposed electrochemical method in saliva samples. The analytical performance of the CB-GCE for detection of seligeline in authentic saliva samples revealed a limit of detection of 0.12 mu mol L-1 which is sufficiently low to allow a possible application of this fast method in the doping control of this drug.
引用
收藏
页码:433 / 439
页数:7
相关论文
共 21 条
[1]   An electrochemical immunoassay for the screening of celiac disease in saliva samples [J].
Adornetto, Gianluca ;
Fabiani, Laura ;
Volpe, Giulia ;
De Stefano, Alessia ;
Martini, Sonia ;
Nenna, Raffaella ;
Lucantoni, Federica ;
Bonamico, Margherita ;
Tiberti, Claudio ;
Moscone, Danila .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (23) :7189-7196
[2]   Non-triangular potential sweep cyclic voltammetry of reversible electron transfer: Experiment meets theory [J].
Amin, Hatem M. A. ;
Uchida, Yuki ;
Batchelor-McAuley, Christopher ;
Katelhon, Enno ;
Compton, Richard G. .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 815 :24-29
[3]  
[Anonymous], 2017, 2017 PROH LIST INT S
[4]   Optimising carbon electrode materials for adsorptive stripping voltammetry Korbua [J].
Chaisiwamongkhol, Korbua ;
Batchelor-McAuley, Christopher ;
Sokolov, Stanislav V. ;
Holter, Jennifer ;
Young, Neil P. ;
Compton, Richard G. .
APPLIED MATERIALS TODAY, 2017, 7 :60-66
[5]   BOXER DISQUALIFIED FOR TAKING SELEGILINE [J].
COLOSIMO, C ;
ALBANESE, A .
LANCET, 1995, 346 (8975) :647-647
[6]   A computational and experimental study of the cyclic voltammetry response of partially blocked electrodes. Part II: Randomly distributed and overlapping blocking systems [J].
Davies, TJ ;
Brookes, BA ;
Fisher, AC ;
Yunus, K ;
Wilkins, SJ ;
Greene, PR ;
Wadhawan, JD ;
Compton, RG .
JOURNAL OF PHYSICAL CHEMISTRY B, 2003, 107 (26) :6431-6444
[7]   Pharmacokinetics study of mazindol in plasma, oral fluid, and urine of volunteers [J].
de Oliveira, Marcella Herbstrith ;
Lukasewicz Ferreira, Pamela Cristina ;
Carlos, Graciela ;
Salazar, Fernanda Rodrigues ;
Bergold, Ana Maria ;
Pechansky, Flavio ;
Limberger, Renata Pereira ;
Froehlich, Pedro Eduardo .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 2016, 72 (08) :945-951
[8]   Saliva as an alternative specimen to plasma for drug bioanalysis: A review [J].
Elmongy, Hatem ;
Abdel-Rehim, Mohamed .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2016, 83 :70-79
[9]  
Ghani NTA, 2012, INT J ELECTROCHEM SC, V7, P7235
[10]  
International Programme on Chemical Safety International Chemical Safety Cards, 2017, CARBON BLACK