Exact and approximate Fourier rebinning algorithms for the solution of the data truncation problem in 3-D PET

被引:10
作者
Ben Bouallegue, Faycal
Crouzet, Jean-Francois
Comtat, Claude
Fourcade, Marjolaine
Mohammadi, Bijan
Mariano-Goulart, Denis
机构
[1] Univ Montpellier 2, Math & Modeling Inst, F-34095 Montpellier 5, France
[2] CEA, Dept Med Res, SHFJ, F-91401 Orsay, France
[3] Montpellier Med Univ, Biophys & Nucl Med Dept, Nucl Med Serv, CHU Lapeyronie, F-34295 Montpellier 5, France
关键词
Fourier rebinning; image reconstruction; medical imaging; positron emission tomography (PET); reprojection; 3-DIMENSIONAL IMAGE-RECONSTRUCTION; PERFORMANCE;
D O I
10.1109/TMI.2007.897362
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper presents an extended 3-D exact rebinning formula in the Fourier space that leads to an iterative reprojection algorithm (iterative FOREPROJ), which enables the estimation of unmeasured oblique projection data on the basis of the whole set of measured data. In first approximation, this analytical formula also leads to an extended Fourier rebinning equation that is the basis for an approximate reprojection algorithm (extended FORE). These algorithms were evaluated on numerically simulated 3-D positron emission tomography (PET) data for the solution of the truncation problem, i.e., the estimation of the missing portions in the oblique projection data, before the application of algorithms that require complete projection data such as some rebinning methods (FOREX) or 3-D reconstruction algorithms (3DRP or direct Fourier methods). By taking advantage of all the 3-D data statistics, the iterative FOREPROJ reprojection provides a reliable alternative to the classical FOREPROJ method, which only exploits the low-statistics nonoblique data. It significantly improves the quality of the external reconstructed slices without loss of spatial resolution. As for the approximate extended FORE algorithm, it clearly exhibits limitations due to axial interpolations, but will require clinical studies with more realistic measured data in order to decide on its pertinence.
引用
收藏
页码:1001 / 1009
页数:9
相关论文
共 27 条
[1]   FULLY 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY [J].
COLSHER, JG .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) :103-115
[2]  
Comtat C, 1999, IEEE NUCL SCI CONF R, P1260, DOI 10.1109/NSSMIC.1999.842786
[3]  
DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717
[4]   A PERFORMANCE STUDY OF 3D RECONSTRUCTION ALGORITHMS FOR POSITRON EMISSION TOMOGRAPHY [J].
DEFRISE, M ;
GEISSBUHLER, A ;
TOWNSEND, DW .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :305-320
[5]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[6]   A fast rebinning algorithm for 3D positron emission tomography using John's equation [J].
Defrise, M ;
Liu, XA .
INVERSE PROBLEMS, 1999, 15 (04) :1047-1065
[7]   A FACTORIZATION METHOD FOR THE 3D X-RAY TRANSFORM [J].
DEFRISE, M .
INVERSE PROBLEMS, 1995, 11 (05) :983-994
[8]  
Edholm P. R., 1986, Proceedings of the SPIE - The International Society for Optical Engineering, V671, P8, DOI 10.1117/12.966672
[9]   3D RECONSTRUCTION FOR A MULTIRING PET SCANNER BY SINGLE-SLICE REBINNING AND AXIAL DECONVOLUTION [J].
ERLANDSSON, K ;
ESSER, PD ;
STRAND, SE ;
VANHEERTUM, RL .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :619-629
[10]   Nonuniform fast Fourier transforms using min-max interpolation [J].
Fessler, JA ;
Sutton, BP .
IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2003, 51 (02) :560-574