Three-year effectiveness and safety of the XEN gel stent as a solo procedure or in combination with phacoemulsification in open-angle glaucoma: a multicentre study

被引:44
作者
Reitsamer, Herbert [1 ]
Vera, Vanessa [2 ]
Ruben, Simon [3 ]
Au, Leon [4 ,5 ]
Vila-Arteaga, Jorge [6 ]
Teus, Miguel [7 ]
Lenzhofer, Markus [1 ]
Shirlaw, Andrew [8 ]
Bai, Zhanying [9 ]
Balaram, Mini [10 ]
Stalmans, Ingeborg [11 ]
机构
[1] SALK Paracelsus Med Univ, Univ Clin Salzburg, Dept Ophthalmol & Optometry, Salzburg, Austria
[2] Unidad Oftalmol Caracas, Dept Ophthalmol, Caracas, Venezuela
[3] Southend Univ Hosp NHS Fdn Trust, Dept Ophthalmol, Westcliff On Sea, England
[4] Manchester Royal Eye Hosp, Dept Eye Res, Manchester, Lancs, England
[5] Univ Manchester, Med Acad Hlth Sci Ctr, Manchester, Lancs, England
[6] Hosp Univ La Fe, Valencia, Spain
[7] Univ Alcal, Hosp Univ Principe Asturias, Madrid, Spain
[8] Allergan, Marlow, Bucks, England
[9] Allergan, Madison, NJ USA
[10] Allergan, Irvine, CA USA
[11] Univ Hosp UZ Leuven, Dept Ophthalmol, Leuven, Belgium
关键词
glaucoma; intraocular pressure; gel stent; implant; MIGS; XEN; IMPLANTATION; SURGERY;
D O I
10.1111/aos.14886
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To assess the 3-year effectiveness and safety of the XEN gel stent implanted ab interno in open-angle glaucoma (OAG). Methods This study was a multicentre, retrospective chart review of consecutive patients with OAG who underwent ab-interno gel stent placement alone or combined with phacoemulsification between 1 January 2014 and 1 October 2015. Outcome measures included mean changes in intraocular pressure (IOP) and IOP-lowering medication count from medicated baseline at 1, 2, 3 (primary outcome) and 4 years (if available) postimplantation. Intraoperative complications, adverse events of special interest (AESIs) and secondary surgical interventions (SSIs) were recorded. Results The safety and effectiveness populations included 212 eyes (primary and secondary) and 174 eyes (primary), respectively. Mean IOP and medication decreased from 20.7 mmHg and 2.5 at baseline (n = 163 primary/first implanted eyes) to 13.9 mmHg and 1.1 medications (n = 76) at 3 years postimplantation, respectively. Mean changes from baseline in IOP (-5.6, -6.2 and -6.6 mmHg) and IOP-lowering medication count (-1.8, -1.6 and -1.4) were statistically significant at 1, 2 and 3 years postimplantation, respectively. Results appeared comparable when implantation was performed with (n = 76) or without (n = 98) phacoemulsification. In primary eyes with 4-year IOP and medication count data (n = 27), mean IOP was 14.0 mmHg on 1.3 medications at 4 years postimplantation. Fifteen (7.1%) eyes had intraoperative complications, 31 (14.6%) experienced 46 postoperative AESIs, and 26 (12.3%) required SSI. Conclusion The gel stent effectively lowered IOP and IOP-lowering medication count over 3 years, with a predictable and acceptable safety profile, when implanted via the traditional ab-interno technique.
引用
收藏
页码:E233 / E245
页数:13
相关论文
共 50 条
  • [31] Standalone XEN45 Gel Stent implantation in the treatment of open-angle glaucoma: A systematic review and meta-analysis
    Lim, Sheng Yang
    Betzler, Bjorn Kaijun
    Yip, Leonard Wei Leon
    Dorairaj, Syril
    Ang, Bryan Chin Hou
    SURVEY OF OPHTHALMOLOGY, 2022, 67 (04) : 1048 - 1061
  • [32] Three-year clinical outcomes of phacoemulsification combined with excisional goniotomy using the kahook dual blade for cataract and open-angle glaucoma in Saudi Arabia
    Albuainain, Abdulrahman
    Al Habash, Ahmed
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2022, 36 (02) : 213 - 217
  • [33] Results of XEN45 Gel Stent Implantation in the Treatment of Primary Open-Angle Glaucoma Using 5, 10 or 20 μg Mitomycin C: A Pilot Study
    Reichel, Felix F.
    Guggenberger, Vanessa
    Faber, Hanna
    Neubauer, Jonas
    Voykov, Bogomil
    JOURNAL OF OPHTHALMOLOGY, 2024, 2024
  • [34] Trabecular microbypass stent combined with phacoemulsification in patients with open-angle glaucoma, 1-year outcome in a Taiwanese population
    Su, Wei-Wen
    TAIWAN JOURNAL OF OPHTHALMOLOGY, 2023, 13 (03) : 360 - 365
  • [35] XEN45 Gel Stent implantation in eyes with primary open angle glaucoma: A study from a single hospital in Taiwan
    Chao, Yu-Jang
    Ko, Yu-Chieh
    Chen, Mei-Ju
    Lo, Kang-Jung
    Chang, Yu-Fan
    Liu, Catherine Jui-Ling
    JOURNAL OF THE CHINESE MEDICAL ASSOCIATION, 2021, 84 (01) : 108 - 113
  • [36] Three-Year Findings of the HORIZON Trial A Schlemm Canal Microstent for Pressure Reduction in Primary Open-Angle Glaucoma and Cataract
    Ahmed, Iqbal Ike K.
    Rhee, Douglas J.
    Jones, Jason
    Singh, Inder Paul
    Radcliffe, Nathan
    Gazzard, Gus
    Samuelson, Thomas W.
    Ong, Jeb
    Singh, Kuldev
    OPHTHALMOLOGY, 2021, 128 (06) : 857 - 865
  • [37] Canaloplasty: Three-year results of circumferential viscodilation and tensioning of Schlemm canal using a microcatheter to treat open-angle glaucoma
    Lewis, Richard A.
    von Wolff, Kurt
    Tetz, Manfred
    Koerber, Norbert
    Kearney, John R.
    Shingleton, Bradford J.
    Samuelson, Thomas W.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (04) : 682 - 690
  • [38] Influence of Mitomycin C on the Therapeutic Success of Stand-Alone Xen45 Gel Stents and Its Combination with Cataract Surgery in Open-Angle Glaucoma Patients
    Denkel, Laura
    Hohberger, Bettina
    Laemmer, Robert
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2021,
  • [39] Influence of Mitomycin C on the Therapeutic Success of Stand-Alone Xen45 Gel Stents and Its Combination with Cataract Surgery in Open-Angle Glaucoma Patients
    Denkel, Laura
    Hohberger, Bettina
    Laemmer, Robert
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2021, 238 (08) : 861 - 867
  • [40] Comparing the efficacy of trabeculectomy and XEN gel microstent implantation for the treatment of primary open-angle glaucoma: a retrospective monocentric comparative cohort study
    Theilig, Theresa
    Rehak, Matus
    Busch, Catharina
    Bormann, Caroline
    Schargus, Marc
    Unterlauft, Jan Darius
    SCIENTIFIC REPORTS, 2020, 10 (01)