Accuracy and precision of a new portable ultrasound scanner, the BME-150A, in residual urine volume measurement: a comparison with the BladderScan BVI 3000

被引:33
作者
Choe, Jin Ho
Lee, Ji Yeon
Lee, Kyu-Sung
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Urol, Seoul, South Korea
[2] Sungkyunkwan Univ, Sch Med, Cheil Gen Hosp, Dept Urol, Seoul, South Korea
关键词
ultrasonography; urinary catheterization; urine;
D O I
10.1007/s00192-006-0219-2
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
The objective of the study was to determine the relative accuracy of a new portable ultrasound unit, BME-150A (TM), and the BladderScan (TM) BVI 3000, as assessed in comparison with the catheterized residual urine volume. We used both of these machines to prospectively measure the residual urine volumes of 89 patients (40 men and 49 women) who were undergoing urodynamic studies. The ultrasound measurements were compared with the postscan bladder volumes obtained by catheterization in the same patients. The ultrasounds were followed immediately (within 5 min) by in-and-out catheterizations while the patients were in a supine position. There were a total of 116 paired measurements made. The BME-150A and the BVI 3000 demonstrated a correlation with the residual volume of 0.92 and 0.94, and a mean difference from the true residual volume of 7.8 and 3.6 ml, respectively. Intraclass correlation coefficients for the accuracy of the two bladder scans were 0.90 for BME-150A and 0.95 for BVI 3000. The difference of accuracy between the two models was not significant (p=0.2421). There were six cases in which a follow-up evaluation of falsely elevated post-void residual urine volume measurements on the ultrasound studies resulted in comparatively low catheterized volumes, with a range of differences from 66 to 275.5 ml. These cases were diagnosed with an ovarian cyst, uterine myoma, or uterine adenomyosis on pelvic ultrasonography. The accuracy of the BME-150A is comparable to that of the BVI 3000 in estimating the true residual urine volumes and is sufficient enough for us to recommend its use as an alternative to catheterization.
引用
收藏
页码:641 / 644
页数:4
相关论文
共 13 条
[1]   The accuracy of portable abdominal ultrasound equipment in measuring postvoid residual volume [J].
Alnaif B. ;
Drutz H.P. .
International Urogynecology Journal, 1999, 10 (4) :215-218
[2]  
CARDENAS DD, 1988, ARCH PHYS MED REHAB, V69, P514
[3]  
Chan H, 1993, J Neurosci Nurs, V25, P309
[4]   THE ACCURACY OF PORTABLE ULTRASOUND SCANNING IN THE MEASUREMENT OF RESIDUAL URINE VOLUME [J].
COOMBES, GM ;
MILLARD, RJ .
JOURNAL OF UROLOGY, 1994, 152 (06) :2083-2085
[5]  
Cooperberg M R, 2000, Urology, V55, P590, DOI 10.1016/S0090-4295(99)00565-8
[6]   ULTRASONIC STUDIES OF BLADDER [J].
HOLMES, JH .
JOURNAL OF UROLOGY, 1967, 97 (04) :654-&
[7]   The accuracy of ultrasonic estimation of bladder volume: A comparison of portable and stationary equipment [J].
Huang, YH ;
Bih, LI ;
Chen, SL ;
Tsai, SJ ;
Teng, CH .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2004, 85 (01) :138-141
[8]   BLADDER VOLUME DETERMINATION USING A DEDICATED, PORTABLE ULTRASOUND SCANNER [J].
IRETON, RC ;
KRIEGER, JN ;
CARDENAS, DD ;
WILLIAMSBURDEN, B ;
KELLY, E ;
SOUCI, T ;
CHAPMAN, WH .
JOURNAL OF UROLOGY, 1990, 143 (05) :909-911
[9]   ACCURACY OF TOTAL BLADDER VOLUME AND RESIDUAL URINE MEASUREMENTS - COMPARISON BETWEEN REAL-TIME ULTRASONOGRAPHY AND CATHETERIZATION [J].
MAINPRIZE, TC ;
DRUTZ, HP .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1989, 160 (04) :1013-1016
[10]   EXPERIENCE WITH PORTABLE ULTRASOUND EQUIPMENT AND MEASUREMENT OF URINE VOLUMES - INTER-USER RELIABILITY AND FACTORS OF PATIENT POSITION [J].
MASSAGLI, TL ;
CARDENAS, DD ;
KELLY, EW .
JOURNAL OF UROLOGY, 1989, 142 (04) :969-971