A study of ageing in timing RPCs

被引:4
作者
Lopes, L
Ferreira-Marques, R
Fonte, P [1 ]
Piedade, AP
Policarpo, A
机构
[1] Univ Coimbra, Dept Fis, LIP Combra, P-3004516 Coimbra, Portugal
[2] Univ Coimbra, Dept Fis, CFRM, P-3004516 Coimbra, Portugal
[3] Univ Coimbra, Dept Engn Mecan, ICEMS, P-3030201 Coimbra, Portugal
关键词
RPC; TOF; ageing;
D O I
10.1016/j.nima.2004.07.013
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The present study concerns the search for ageing effects in 0.3 mm gap timing RPCs. Counters were made with glass and aluminium electrodes and kept under a continuous gas flow of tetrafluorethane with 10% sulphur hexafluoride, 5% isobutane and water at 10% relative humidity. After 500 days of operation at the saturation current an accumulated charge of 800 mC was collected, equivalent to over 8 years of continuous operation at 300 Hz/cm(2). No evidence was found of any long-term increase of dark current, suggesting that the ageing effect in timing RPCs is less severe than in streamer-mode glass RPCs. However, a visual inspection of the electrodes confirmed the existence of a deposit over the glass electrodes that was analysed by Electron Probe Microanalysis. (C) 2004 Elsevier B.V. All rights reserved.