Errors in ultrasound digital image distance measurements

被引:27
作者
Goldstein, A [1 ]
机构
[1] Wayne State Univ, Sch Med, Dept Radiol, Detroit, MI USA
关键词
digital image; digital caliper measurement; digital distance measurement; pixelation errors; edge shifting; image distance measurement protocols;
D O I
10.1016/S0301-5629(00)00249-0
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Digital image distance measurements have two sources of error: they are the inherent image pixelation error and modality-specific registration errors. A theoretical analysis was performed to determine the pixelation error uncertainty in ultrasound (US) clinical and phantom digital image distance measurements. For horizontal or vertical phantom image distance measurements, the average absolute image pixelation error was found to be 0.5 pixel width. In the clinical horizontal case, the combined absolute distance measurement error due to pixel size and operator cursor placement errors was found to be 1.5 pixel widths (for a cursor placement error of 1 pixel width). For image clinical distance measurements, a formula is given for the dependence of the measurement error on the inclination of the measured structure in the pixel matrix, A protocol is recommended that minimizes all sources of error in clinical image distance measurements, (C) 2000 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1125 / 1132
页数:8
相关论文
共 7 条
  • [1] THE ACCURACY OF DIMENSIONAL MEASUREMENT IN ULTRASONIC REAL-TIME SCANNING
    CHIVERS, RC
    BACON, RA
    BLOOMBERG, TJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (647) : 978 - 981
  • [2] DALE G, 1980, INVESTIGATIVE ULTRAS, V1, P33
  • [3] GOLDSTEIN A, 1983, J ULTRAS MED, V2, P195
  • [4] GOLDSTEIN A, 1988, J ULTRAS MED, V7, P487
  • [5] The effect of acoustic velocity on phantom measurements
    Goldstein, L
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (07) : 1133 - 1143
  • [6] THE INFLUENCE OF ULTRASOUND SCANNER BEAM WIDTH ON FEMUR LENGTH MEASUREMENTS
    JAGO, JR
    WHITTINGHAM, TA
    HESLOP, R
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) : 699 - 703
  • [7] Liquid or solid ultrasonically tissue-mimicking materials with very low scatter
    Madsen, EL
    Frank, GR
    Dong, F
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (04) : 535 - 542