The validation of column-switching LC/MS as a high-throughput approach for direct analysis of di(2-ethylhexyl) phthalate released from PVC medical devices in intravenous solution

被引:24
作者
Inoue, K
Higuchi, T
Okada, F
Iguchi, H
Yoshimura, Y
Sato, A
Nakazawa, H
机构
[1] Hoshi Univ, Dept Analyt Chem, Fac Pharmaceut Sci, Shinagawa Ku, Tokyo 1428501, Japan
[2] Yokohama City Univ Med, Dept Pharm, Kanazawa Ku, Yokohama, Kanagawa, Japan
[3] Showa Univ, Dept Oral Biomat & Technol, Sch Dent, Shinagawa Ku, Tokyo 1428555, Japan
关键词
di(2-ethylhexyl)phthalate; DEHP; PVC medical devices; intravenous (IV) solution; column-switching; LC/MS;
D O I
10.1016/S0731-7085(03)00019-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Health Canada reported recently that medical devices containing di(2-ethylhexyl) phthalate (DEHP) should not be used in the clinical treatment of infants, young boys, pregnant women, and nursing mothers. The risk assessment of DEHP released from PVC medical devices is an important issue for hospitalized patients. In this study, a simple, accurate, low-contamination and high-throughput analytical technique for the determination of DEHP in intravenous (IV) solution was developed using column-switching liquid chromatography/mass spectrometry (LC/MS) with an extraction mini-column. The sample preparation for on-line extraction involved simply mixing IV solution with internal standard as DEHP-d(4) in LC glass vials. The IV fat emulsion drug sample cannot be analyzed directly, hence this sample spiked with DEHP-d4 solution was extracted by hexane and measured by column-switching LC/MS yielding an average recovery of 92.2% (C.V. = 7.8%, n = 5). A linear response was found for a variety of drugs tested within the validated range of 0.1 or 0.5-10 mug/ml with correlation coefficients (r) greater than 0.99. These results suggest that this method can assay background exposure to DEHP released from PVC medical devices in the patients. The method was applied to various IV solution samples to establish the first screening method for DEHP released from medical devices with respect to their safety. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1145 / 1152
页数:8
相关论文
共 19 条
[1]   Determination of bis(2-ethylhexyl) phthalate in water by high-performance liquid chromatography with direct on-column preconcentration [J].
Baram, GI ;
Azarova, IN ;
Gorshkov, AG ;
Vereshchagin, AL ;
Lang, B ;
Kiryukhina, ED .
JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 55 (08) :750-754
[2]  
Berset JD, 2001, J AOAC INT, V84, P383
[3]   Quantitative detection of eight phthalate metabolites in human urine using HPLC-APCI-MS/MS [J].
Blount, BC ;
Milgram, KE ;
Silva, MJ ;
Malek, NA ;
Reidy, JA ;
Needham, LL ;
Brock, JW .
ANALYTICAL CHEMISTRY, 2000, 72 (17) :4127-4134
[4]  
EBIHARA M, 2000, IYAKUHIN SOGO SAYO K, V24, P3
[5]  
GIAM CS, 1975, ANAL CHEM, V47, P2319, DOI 10.1021/ac60363a017
[6]   Investigation of the release behavior of diethylhexyl phthalate from the polyvinyl-chloride tubing for intravenous administration [J].
Hanawa, T ;
Muramatsu, E ;
Asakawa, K ;
Suzuki, M ;
Tanaka, M ;
Kawano, K ;
Seki, T ;
Juni, K ;
Nakajima, S .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2000, 210 (1-2) :109-115
[7]   MASS-SPECTROMETRY OF DI-(2-ETHYLHEXYL)PHTHALATE METABOLITES [J].
HARVAN, DJ ;
HASS, JR ;
ALBRO, PW ;
FRIESEN, MD .
BIOMEDICAL MASS SPECTROMETRY, 1980, 7 (06) :242-246
[8]   The clinical effects of plasticizers, antioxidants, and other contaminants in medical polyvinylchloride tubing during respiratory and non-respiratory exposure [J].
Hill, SS ;
Shaw, BR ;
Wu, AHB .
CLINICA CHIMICA ACTA, 2001, 304 (1-2) :1-8
[9]   Determination of phthalates in water samples using polystyrene solid-phase extraction and liquid chromatography quantification [J].
Jara, S ;
Lysebo, C ;
Greibrokk, T ;
Lundanes, E .
ANALYTICA CHIMICA ACTA, 2000, 407 (1-2) :165-171
[10]   Evaluation of model solvent systems for assessing the accumulation of container extractables in drug formulations [J].
Jenke, DR .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2001, 224 (1-2) :51-60