Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication

被引:22
作者
Dave, Mital H. [1 ]
Koepfer, Nelly [1 ]
Madjdpour, Caveh [1 ]
Frotzler, Angela [1 ]
Weiss, Markus [1 ]
机构
[1] Univ Childrens Hosp, Dept Anaesthesia, CH-8032 Zurich, Switzerland
关键词
Tracheal tube cuff; PEEP; Benchtop trials; Aspiration; TUBE CUFF; GEL LUBRICATION; PREVENT LEAKAGE; HIGH-VOLUME; PRESSURE; ASPIRATION; SURGERY;
D O I
10.1007/s00540-010-0871-z
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Longitudinal folds in tracheal tube (TT) cuffs cause leakage of pooled secretions past the tube cuff, and the most common in vitro method to test the efficacy of a new tube is a benchtop model using an artificial rigid trachea. This study compared the potential of a static and dynamic ventilation benchtop model and cuff lubrication in testing the tracheal sealing properties of a given TT cuff. Static trial Six brands of 7.5 mm internal diameter (ID) cuffed TT (n = 8) with high volume-low pressure cuffs were inflated in an artificial trachea (18 mm ID) without and with lubrication. Dynamic trial The same tube cuffs, without lubrication, were subjected to positive pressure ventilation (PPV) + positive end-expiratory pressure (PEEP) of 5cmH(2)O or to PPV alone (without PEEP) or to PEEP alone (without PPV). Clear water (5 ml) was placed above the tube cuff, and fluid leakage (ml) was measured up to 60 min. Gel lubrication, PEEP alone and PPV + PEEP completely prevented fluid leakage across the tube cuffs in all six TT brands tested within 60 min when compared to the static unlubricated model (0% leak versus 100% leak; P < 0.01). Fluid leakage in the static unlubricated model and the PPV group was 1.38-4.76 ml and 0.23-4.47 ml, respectively. Gel lubrication, PEEP alone, and PPV + PEEP in the benchtop model had a much stronger protective effect than PPV alone on fluid leakage. Studies testing the fluid sealing efficiency of tube cuffs might be more conclusive in a static benchtop model without lubrication than in a dynamic model.
引用
收藏
页码:247 / 252
页数:6
相关论文
共 13 条
  • [1] Leakage of fluid around high-volume, low-pressure cuffs - A comparison of four tracheal tubes
    Asai, T
    Shingu, K
    [J]. ANAESTHESIA, 2001, 56 (01) : 38 - 42
  • [2] Gel lubrication of the tracheal tube cuff reduces pulmonary aspiration
    Blunt, MC
    Young, PJ
    Patil, A
    Haddock, A
    [J]. ANESTHESIOLOGY, 2001, 95 (02) : 377 - 381
  • [3] Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube
    Dullenkopf, A
    Gerber, A
    Weiss, M
    [J]. INTENSIVE CARE MEDICINE, 2003, 29 (10) : 1849 - 1853
  • [4] Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit
    Lucangelo, Umberto
    Zin, Walter A.
    Antonaglia, Vittorio
    Petrucci, Lara
    Viviani, Marino
    Buscema, Giovanni
    Borelli, Massimo
    Berlot, Giorgio
    [J]. CRITICAL CARE MEDICINE, 2008, 36 (02) : 409 - 413
  • [5] Leakage of fluid around low-pressure tracheal tube cuffs
    Oikkonen, M
    Aromaa, U
    [J]. ANAESTHESIA, 1997, 52 (06) : 567 - 569
  • [6] Polyurethane cuffed endotracheal tubes to prevent early postoperative pneumonia after cardiac surgery: A pilot study
    Poelaert, Jan
    Depuydt, Pieter
    De Wolf, Annick
    de Velde, Stijn Van
    Herck, Ingrid
    Blot, Stijn
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2008, 135 (04) : 771 - 776
  • [7] The effect of gel lubrication on cuff leakage of double lumen tubes during thoracic surgery
    Sanjay, PS
    Miller, SA
    Corry, PR
    Russell, GN
    Pennefather, SH
    [J]. ANAESTHESIA, 2006, 61 (02) : 133 - 137
  • [8] ASPIRATION BEYOND ENDOTRACHEAL CUFFS
    SEEGOBIN, RD
    VANHASSELT, GL
    [J]. CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1986, 33 (03) : 273 - 279
  • [9] Evaluation of a new design of tracheal tube cuff to prevent leakage of fluid to the lungs
    Young, PJ
    Ridley, SA
    Downward, G
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1998, 80 (06) : 796 - 799
  • [10] Leakage of fluid past the tracheal tube cuff in a benchtop model
    Young, PJ
    Rollinson, M
    Downward, G
    Henderson, S
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1997, 78 (05) : 557 - 562