Left Atrial Pressure Monitoring With an Implantable Wireless Pressure Sensor After Implantation of a Left Ventricular Assist Device

被引:29
|
作者
Hubbert, Laila [1 ,2 ]
Baranowski, Jacek [2 ,3 ]
Delshad, Baz [2 ,4 ]
Ahn, Henrik [2 ,4 ]
机构
[1] Linkoping Univ, Dept Cardiol, Linkoping, Sweden
[2] Linkoping Univ, Dept Med & Hlth Sci, Linkoping, Sweden
[3] Linkoping Univ, Dept Clin Physiol, Linkoping, Sweden
[4] Linkoping Univ, Dept Thorac & Vasc Surg, Linkoping, Sweden
关键词
mechanical circulatory support; LVAD; left atrial pressure; MEMS; QUALITY-OF-LIFE; MECHANICAL CIRCULATORY SUPPORT; END-ORGAN FUNCTION; CONTINUOUS-FLOW; TRIAL;
D O I
10.1097/MAT.0000000000000451
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
After implantation of a continuous-flow left ventricular assist device (LVAD), left atrial pressure (LAP) monitoring allows for the precise management of intravascular volume, inotropic therapy, and pump speed. In this case series of 4 LVAD recipients, we report the first clinical use of this wireless pressure sensor for the long-term monitoring of LAP during LVAD support. A wireless microelectromechanical system pressure sensor (Titan, ISS Inc., Ypsilanti, MI) was placed in the left atrium in four patients at the time of LVAD implantation. Titan sensor LAP was measured in all four patients on the intensive care unit and in three patients at home. Ramped speed tests were performed using LAP and echocardiography in three patients. The left ventricular end-diastolic diameter (cm), flow (L/min), power consumption (W), and blood pressure (mm Hg) were measured at each step. Measurements were performed over 36, 84, 137, and 180 days, respectively. The three discharged patients had equipment at home and were able to perform daily recordings. There were significant correlations between sensor pressure and pump speed, LV and LA size and pulmonary capillary wedge pressure, respectively (r = 0.92-0.99, p < 0.05). There was no device failure, and there were no adverse consequences of its use.
引用
收藏
页码:E60 / E65
页数:6
相关论文
共 50 条
  • [1] Right Atrial Pressure Waveform Predicts Right Ventricular Failure After Left Ventricular Assist Device Implantation
    Samura, Takaaki
    Yoshioka, Daisuke
    Asanoi, Hidetsugu
    Toda, Koichi
    Miyagawa, Shigeru
    Yoshikawa, Yasushi
    Hata, Hiroki
    Kainuma, Satoshi
    Kawamura, Takuji
    Kawamura, Ai
    Sakata, Yasushi
    Sawa, Yoshiki
    ANNALS OF THORACIC SURGERY, 2019, 108 (05): : 1361 - 1368
  • [2] Safety and Impact of Direct Left Atrial Pressure Monitoring in Patients Undergoing Continuous-Flow Left Ventricular Assist Device Implantation
    Randhawa, V. K.
    Soltesz, E. G.
    Faulkenberg, K. D.
    Wang, Q.
    Wolski, K. E.
    Tong, M. Z.
    Unai, S.
    West, L. M.
    Buda, T.
    Tang, W.
    Estep, J. D.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2020, 39 (04): : S443 - S444
  • [3] Right ventricular failure after left ventricular assist device implantation:: The need for an implantable right ventricular assist device
    Furukawa, K
    Motomura, T
    Nosé, Y
    ARTIFICIAL ORGANS, 2005, 29 (05) : 369 - 377
  • [4] Changes in Pulmonary Artery Pressure Utilizing Remote Monitoring Data after Left Ventricular Assist Device Implantation
    Kilic, Ahmet
    Katz, Jason N.
    Joseph, Susan
    Brisco-Bacik, Meredith A.
    Uriel, Nir
    Lima, Brian
    Agarwal, Rahul
    Bharmi, Rupinder
    Lee, Sangjin
    JOURNAL OF CARDIAC FAILURE, 2017, 23 (08) : S30 - S30
  • [5] Usefulness of Analyzing Right Atrial Pressure Waveform to Predict Right Ventricular Failure After Left Ventricular Assist Device Implantation
    Samura, T.
    Asanoi, H.
    Toda, K.
    Yoshioka, D.
    Miyagawa, S.
    Yoshikawa, Y.
    Hata, H.
    Saito, S.
    Domae, K.
    Matsuura, R.
    Sawa, Y.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (04): : S180 - S180
  • [6] An implantable Fabry-Perot pressure sensor fabricated on left ventricular assist device for heart failure
    Zhou, Ming-Da
    Yang, Chuan
    Liu, Zhiwen
    Cysyk, Joshua P.
    Zheng, Si-Yang
    BIOMEDICAL MICRODEVICES, 2012, 14 (01) : 235 - 245
  • [7] Timing and Trends of Right Atrial Pressure and Risk of Right Heart Failure After Left Ventricular Assist Device Implantation
    Gulati, Gaurav
    Sutaria, Nilay
    Vest, Amanda R.
    Denofrio, David D.
    KawaborI, Masashi
    Couper, Gregory
    Kiernan, Michael S.
    JOURNAL OF CARDIAC FAILURE, 2020, 26 (05) : 394 - 401
  • [8] Development of an Inlet Pressure Sensor for Control in a Left Ventricular Assist Device
    Fritz, Bryan
    Cysyk, Joshua
    Newswanger, Ray
    Weiss, William
    Rosenberg, Gerson
    ASAIO JOURNAL, 2010, 56 (03) : 180 - 185
  • [9] An Implantable Intravascular Pressure Sensor for a Ventricular Assist Device
    Brancato, Luigi
    Keulemans, Grim
    Verbelen, Tom
    Meyns, Bart
    Puers, Robert
    MICROMACHINES, 2016, 7 (08):
  • [10] A novel implantable left atrial sensor for pressure monitoring in heart failure
    Di Mario, Carlo
    Sievert, Kolja
    Meucci, Francesco
    Sievert, Horst
    EUROINTERVENTION, 2020, 16 (05) : 432 - 433