Comparative Dose of Contrast-Enhanced Spectral Mammography (CESM), Digital Mammography, and Digital Breast Tomosynthesis

被引:46
作者
Phillips, Jordana [1 ,2 ]
Mihai, Georgeta [1 ,2 ]
Hassonjee, Sarah Esaa [1 ,2 ]
Raj, Sean D. [1 ,2 ]
Palmer, Matthew R. [1 ,2 ]
Brook, Alexander [1 ,2 ]
Zhang, Da [1 ,2 ]
机构
[1] Harvard Med Sch, Dept Radiol, 330 Brookline Ave,4th Fl, Boston, MA 02215 USA
[2] Beth Israel Deaconess Med Ctr, 330 Brookline Ave,4th Fl, Boston, MA 02215 USA
关键词
contrast-enhanced digital mammography; contrast-enhanced mammography; contrast-enhanced spectral mammography; radiation dose; DOSIMETRY PROTOCOLS; RADIATION-EXPOSURE; PERFORMANCE; MRI; UK;
D O I
10.2214/AJR.17.19036
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to provide a more accurate estimation of the radiation dose of contrast-enhanced spectral mammography (CESM) relative to that of 2D digital mammography and tomosynthesis using phantom and patient data and an accepted dosimetry protocol that eliminates vendor-specific average glandular dose (AGD) estimates while including breast density. MATERIALS AND METHODS. Patient and phantom AGD estimation was performed using two vendors (system 1 and system 2) in five imaging modes, including 2D, 3D, and CESM imaging. Patient AGD was retrospectively estimated from 45 patients who underwent mammography with all imaging modes during 2012-2016. Patient and phantom AGD were estimated using accepted European and International Atomic Energy Agency protocols for dosimetry and were compared across imaging modes using a paired t test with Bonferroni correction. RESULTS. Phantom data showed that the imaging modes with the lowest to highest AGDs were system 1 2D, followed by system 2 2D and system 2 3D, which had comparable values (p = 0.6), followed by system 1 CESM, and then by system 2 2D plus 3D. One hundred eighty views in 45 patients showed that the system 1 CESM AGD was 1.8 times greater than the system 1 21) AU) (p < 0.001), 1.2 times greater than the system 2 21) AO) (p < 0.001), 1.2 times greater than the system 2 3D AGD (p < 0.001), and 0.6 times less than the system 2 2D plus 3D AGD (p < 0.001). CONCLUSION. The CESM dose for system 1 is within an acceptable range as compared with other commonly performed mammographic examinations and should not preclude its use as a diagnostic breast imaging tool.
引用
收藏
页码:839 / 846
页数:8
相关论文
共 25 条
[1]   Contrast-enhanced Spectral Mammography: Technique, Indications, and Clinical Applications [J].
Bhimani, Chandni ;
Matta, Danielle ;
Roth, Robyn G. ;
Liao, Lydia ;
Tinney, Elizabeth ;
Brill, Kristin ;
Germaine, Pauline .
ACADEMIC RADIOLOGY, 2017, 24 (01) :84-88
[2]   Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data [J].
Bouwman, R. W. ;
van Engen, R. E. ;
Young, K. C. ;
den Heeten, G. J. ;
Broeders, M. J. M. ;
Schopphoven, S. ;
Jeukens, C. R. L. P. N. ;
Veldkamp, W. J. H. ;
Dance, D. R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20) :7893-7907
[3]   Diagnostic performance of dual-energy contrast-enhanced subtracted mammography in dense breasts compared to mammography alone: interobserver blind-reading analysis [J].
Cheung, Yun-Chung ;
Lin, Yu-Ching ;
Wan, Yung-Liang ;
Yeow, Kee-Min ;
Huang, Pei-Chin ;
Lo, Yung-Feng ;
Tsai, Hsiu-Pei ;
Ueng, Shir-Hwa ;
Chang, Chee-Jen .
EUROPEAN RADIOLOGY, 2014, 24 (10) :2394-2403
[4]   Estimation of mean glandular dose for contrast enhanced digital mammography: factors for use with the UK, European and IAEA breast dosimetry protocols [J].
Dance, D. R. ;
Young, K. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) :2127-2137
[5]   Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols [J].
Dance, D. R. ;
Young, K. C. ;
van Engen, R. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (14) :4361-4372
[6]   Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol [J].
Dance, DR ;
Skinner, CL ;
Young, KC ;
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3225-3240
[7]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[8]  
Dance DR, 2011, PHYS MED BIOL
[9]   Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study [J].
Dromain, Clarisse ;
Thibault, Fabienne ;
Diekmann, Felix ;
Fallenberg, Eva M. ;
Jong, Roberta A. ;
Koomen, Marcia ;
Hendrick, R. Edward ;
Tardivon, Anne ;
Toledano, Alicia .
BREAST CANCER RESEARCH, 2012, 14 (03)
[10]   Dual-energy contrast-enhanced digital mammography: initial clinical results [J].
Dromain, Clarisse ;
Thibault, Fabienne ;
Muller, Serge ;
Rimareix, Francoise ;
Delaloge, Suzette ;
Tardivon, Anne ;
Balleyguier, Corinne .
EUROPEAN RADIOLOGY, 2011, 21 (03) :565-574