Correlation analysis of CT-based rectal planning dosimetric parameters with in vivo dosimetry of MOSkin and PTW 9112 detectors in Co-60 source HDR intracavitary cervix brachytherapy

被引:4
作者
Jamalludin, Z. [1 ,2 ]
Malik, R. A. [2 ]
Ung, N. M. [2 ,3 ]
机构
[1] Univ Malaya, Med Ctr, Med Phys Unit, Kuala Lumpur 59100, Malaysia
[2] Univ Malaya, Med Ctr, Dept Clin Oncol, Kuala Lumpur 59100, Malaysia
[3] Univ Malaya, Fac Med, Clin Oncol Unit, Kuala Lumpur 50603, Malaysia
关键词
MOSkin; In vivo dosimetry; Rectal ICRU; Rectal D-2cc; Co-60 HDR brachytherapy; DOSE-RATE BRACHYTHERAPY; VOLUME PARAMETERS; RISK-FACTORS; CANCER; COMPLICATIONS; RADIOTHERAPY; TOMOGRAPHY; CARCINOMA;
D O I
10.1007/s13246-021-01026-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intracavitary cervical brachytherapy delivers high doses of radiation to the target tissue and a portion of these doses will also hit the rectal organs due to their close proximity. Rectal dose can be evaluated from dosimetric parameters in the treatment planning system (TPS) and in vivo (IV) dose measurement. This study analyzed the correlation between IV rectal dose with selected volume and point dose parameters from TPS. A total of 48 insertions were performed and IV dose was measured using the commercial PTW 9112 semiconductor diode probe. In 18 of 48 insertions, a single MOSkin detector was attached on the probe surface at 50 mm from the tip. Four rectal dosimetric parameters were retrospectively collected from TPS; (a) PTW 9112 diode maximum reported dose (RPmax) and MOSkin detector, (b) minimum dose to 2 cc (D-2cc), (c) ICRU reference point (ICRUr), and (d) maximum dose from additional points (R-max). The IV doses from both detectors were analyzed for correlation with these dosimetric parameters. This study found a significantly high correlation between IV measured dose from RPmax (r = 0.916) and MOSkin (r = 0.959) with TPS planned dose. The correlation between measured RPmax with both D-2cc and R-max revealed high correlation of r > 0.7, whereas moderate correlation (r = 0.525) was observed with ICRUr. There was no significant correlation between MOSkin IV measured dose with D-2cc, ICRUr and R-max. The non-significant correlation between parameters was ascribable to differences in both detector position within patients, and dosimetric volume and point location determined on TPS, rather than detector uncertainties.
引用
收藏
页码:773 / 783
页数:11
相关论文
共 28 条
[1]   In-vivo rectal dose measurements with diodes to avoid misadministrations during intracavitary high dose rate brachytherapy for carcinoma of the cervix [J].
Alecu, R ;
Alecu, M .
MEDICAL PHYSICS, 1999, 26 (05) :768-770
[2]   In vivo dosimetry with semiconductors in medium dose rate (MDR) brachytherapy for cervical cancer [J].
Allahverdi, Mahmoud ;
Jaberi, Ramin ;
Aghili, Mehdi ;
Ghahremani, Fatemeh ;
Geraily, Ghazale .
JAPANESE JOURNAL OF RADIOLOGY, 2013, 31 (03) :160-165
[3]   Clinical application of MOSkin dosimeters to rectal wall in vivo dosimetry in gynecological HDR brachytherapy [J].
Carrara, M. ;
Romanyukha, A. ;
Tenconi, C. ;
Mazzeo, D. ;
Cerrotta, A. ;
Borroni, M. ;
Cutajar, D. ;
Petasecca, M. ;
Lerch, M. ;
Bucci, J. ;
Richetti, A. ;
Presilla, S. ;
Fallai, C. ;
Gambarini, G. ;
Pignoli, E. ;
Rosenfeld, A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 41 :5-12
[4]   Semiconductor real-time quality assurance dosimetry in brachytherapy [J].
Carrara, Mauro ;
Cutajar, Dean ;
Alnaghy, Saree ;
Espinoza, Anthony ;
Romanyukha, Anna ;
Presilla, Stefano ;
Tenconi, Chiara ;
Cerrotta, Annamaria ;
Fallai, Carlo ;
Safavi-Naeini, Mitra ;
Petasecca, Marco ;
Kejda, Alannah ;
Lerch, Michael ;
Corde, Stephanie ;
Jackson, Michael ;
Howie, Andrew ;
Bucci, Joseph ;
Rosenfeld, Anatoly B. .
BRACHYTHERAPY, 2018, 17 (01) :133-145
[5]  
Cheng G, 2015, INT J RADIAT ONCOL, V93, pE253
[6]   Unique role of proximal rectal dose in late rectal complications for patients with cervical cancer undergoing high-dose-rate intracavitary brachytherapy [J].
Cheng, JCH ;
Peng, LC ;
Chen, YH ;
Huang, DYC ;
Wu, JK ;
Jian, JJM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04) :1010-1018
[7]  
Choi P, 1992, Clin Oncol (R Coll Radiol), V4, P186, DOI 10.1016/S0936-6555(05)81087-8
[8]   Correlation of dose-volume parameters, endoscopic and clinical rectal side effects in cervix cancer patients treated with definitive radiotherapy including MRI-based brachytherapy [J].
Georg, Petra ;
Kirisits, Christian ;
Goldner, Gregor ;
Doerr, Wolfgang ;
Hammer, Johan ;
Poetzi, Regina ;
Berger, Daniel ;
Dimopoulos, Johannes ;
Georg, Dietmar ;
Poetter, Richard .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (02) :173-180
[9]   Dependency of semiconductor dosimeter responses, used in MDR/LDR brachytherapy, on factors which are important in clinical conditions [J].
Ghahramani, Fatemeh ;
Allahverdi, Mahmoud ;
Jaberi, Ramin .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2008, 13 (01) :29-33
[10]   In vivo diode dosimetry vs. computerized tomography and digitally reconstructed radiographs for critical organ dose calculation in high-dose-rate brachytherapy of cervical cancer [J].
Hassouna, Ashraf H. ;
Bahadur, Yasir A. ;
Constantinescu, Camelia ;
El Sayed, Mohamed E. ;
Naseem, Hussain ;
Naga, Adly F. .
BRACHYTHERAPY, 2011, 10 (06) :498-502