Scatter and attenuation corrections for a PEM detector using List-Mode OSEM

被引:0
|
作者
Ferreira, Claudia S. [1 ,2 ]
Cao, Liji [3 ]
Bugalho, Ricardo [2 ]
Matela, Nuno [1 ]
Ortigao, Catarina [2 ]
Varela, Joao [2 ]
Almeida, Pedro [1 ]
机构
[1] Univ Lisbon, Fac Ciencias, Inst Biofis & Engn Biomed, Lisbon, Portugal
[2] LIP Lab Instrumentacao & Fis Expt Particulas, Lisbon, Portugal
[3] German Canc Res Ctr, Div Med Phys Radiol, Heidelberg, Germany
来源
2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2013年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Emission Tomography imaging techniques are known for its quantitative potential. As in whole-body Positron Emission Tomography (PET), also in breast dedicated Positron Emission Mammography (PEM) systems, this goal can only be achieved if a set of corrections for known physics processes is implemented thoroughly. Corrections for Compton scattered events, photon attenuation and random events are examples of such important data corrections. In this work, we report the performance of scatter and attenuation corrections, specifically implemented for PEM. These correction methods were tested on data from the ClearPEM scanner, a dual-head planar system. The ClearPEM scanner does not incorporate transmission images for attenuation correction. With no anatomical information available, the PEM emission image was used to obtain the acquisition's characteristics needed as input for both the Monte Carlo (MC) based scatter correction method and the attenuation correction algorithm. A MC scatter correction can accurately model the scatter distribution of a PEM acquisition, but removing this estimate from the acquired data can result in a severe bias and increased noise in low counts PEM studies. To avoid such drawbacks, an approach based on a smoothed imaged of the scatter distribution was used. Due to the typically low counts nature of PEM acquisitions, correction procedures were combined with a 3D-OSEM-LM reconstruction, so reconstructed images could be obtained in a short period of time. Results demonstrated that both corrections perform as expected. The presence of scattered events was strongly reduced. Counts loss due to photon attenuation was clearly compensated, along with the recovery of the correct activity distribution.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] A LIST-MODE OSEM-BASED ATTENUATION AND SCATTER COMPENSATION METHOD FOR SPECT
    Rahman, Md Ashequr
    Laforest, Richard
    Jha, Abhinav K.
    2020 IEEE 17TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2020), 2020, : 646 - 650
  • [2] Projector with Realistic Detector Scatter Modelling for PET List-Mode Reconstruction
    Autret, Awen
    Bert, Julien
    Strauss, Olivier
    Visvikis, Dimitris
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3333 - 3336
  • [3] Novel Scatter Compensation of List-Mode PET Data Using Spatial and Energy Dependent Corrections
    Guerin, Bastien
    El Fakhri, Georges
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2011, 30 (03) : 759 - 773
  • [4] GPU acceleration of MOLAR for HRRT list-mode OSEM reconstructions
    Barker, W. Craig
    Thada, Shanthalaxmi
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3004 - 3008
  • [5] Strategies for Accelerating Forward and Backprojection in List-mode OSEM PET Reconstruction using GPUs
    Dieckmann, William
    Thada, Shanthalaxmi
    Barker, W. Craig
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 4110 - 4113
  • [6] Sampling Lines of Response for the Sensitivity Image in List-Mode OSEM PET Reconstruction
    Dieckmann, William
    Thada, Shanthalaxmi
    Barker, W. Craig
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [7] Fully 3-D List-Mode OSEM Accelerated by Graphics Processing Units
    Pratx, Guillem
    Chinn, Garry
    Habte, Frezghi
    Olcott, Peter
    Levin, Craig
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2196 - 2202
  • [8] Fast Energy Dependent Scatter Correction for List-Mode PET Data
    Alvarez-Gomez, Juan Manuel
    Santos-Blasco, Joaquin
    Moliner Martinez, Laura
    Rodriguez-Alvarez, Maria Jose
    JOURNAL OF IMAGING, 2021, 7 (10)
  • [9] List-Mode Detector Scatter Estimation Method with Potential for Real-Time Implementation in High Resolution PET Imaging
    Ortega Maynez, Leticia
    Koesters, Thomas
    Uhlendorf, Volkmar
    Wuebbeling, Frank
    Schaefers, Klaus
    Reader, Andrew J.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4543 - +
  • [10] Scatter modeling for 3-D PET list-mode EM reconstruction
    Manavaki, R
    Reader, AJ
    Keller, C
    Missimer, J
    Walledge, RJ
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1863 - 1868