A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study

被引:38
作者
Schoot, Tessa S. [1 ]
Weenk, Mariska [2 ]
van de Belt, Tom H. [3 ]
Engelen, Lucien J. L. P. G. [3 ]
van Goor, Harry [2 ]
Bredie, Sebastian J. H. [1 ,4 ]
机构
[1] Radboud Univ Nijmegen, Dept Internal Med, Med Ctr, POB 9101, NL-6533 AR Nijmegen, Netherlands
[2] Radboud Univ Nijmegen, Dept Surg, Med Ctr, Nijmegen, Netherlands
[3] Radboud Univ Nijmegen, Radboud REshape Innovat Ctr, Med Ctr, Nijmegen, Netherlands
[4] Radboud Univ Nijmegen, Radboud REshape Innovat Ctr, POB 9101, NL-6533 AR Nijmegen, Netherlands
关键词
hypertension; cuffless blood pressure monitor; wearable device; cardiovascular risk management; patient empowerment; HYPERTENSION; PROTOCOL;
D O I
10.2196/jmir.5414
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Cuffless blood pressure (BP) monitoring devices, based on pulse transit time, are being developed as an easy-to-use, more convenient, fast, and relatively cheap alternative to conventional BP measuring devices based on cuff occlusion. Thereby they may provide a great alternative to BP self-measurement. Objective: The objective of our study was to evaluate the performance of the first release of the Checkme Health Monitor (Viatom Technology), a cuffless BP monitor, in a real-life setting. Furthermore, we wanted to investigate whether the posture of the volunteer and the position of the device relative to the heart level would influence its outcomes. Methods: Study volunteers fell into 3 BP ranges: high (> 160 mmHg), normal (130-160 mmHg), and low (< 130 mmHg). All requirements for test environment, observer qualification, volunteer recruitment, and BP measurements were met according to the European Society of Hypertension International Protocol (ESH-IP) for the validation of BP measurement devices. After calibrating the Checkme device, we measured systolic BP with Checkme and a validated, oscillometric reference BP monitor (RM). Measurements were performed in randomized order both in supine and in sitting position, and with Checkme at and above heart level. Results: We recruited 52 volunteers, of whom we excluded 15 (12 due to calibration failure with Checkme, 3 due to a variety of reasons). The remaining 37 volunteers were divided into low (n= 14), medium (n= 13), and high (n= 10) BP ranges. There were 18 men and 19 women, with a mean age of 54.1 (SD 14.5) years, and mean recruitment systolic BP of 141.7 (SD 24.7) mmHg. BP results obtained by RM and Checkme correlated well. In the supine position, the difference between the RM and Checkme was > 5 mmHg in 17 of 37 volunteers (46%), of whom 9 of 37 (24%) had a difference > 10 mmHg and 5 of 37 (14%) had a difference > 15 mmHg. Conclusions: BP obtained with Checkme correlated well with RM BP, particularly in the position (supine) in which the device was calibrated. These preliminary results are promising for conducting further research on cuffless BP measurement in the clinical and outpatient settings.
引用
收藏
页数:10
相关论文
共 13 条
[1]   Oscillometric blood pressure: a review for clinicians [J].
Alpert, Bruce S. ;
Quinn, David ;
Gallick, David .
JOURNAL OF THE AMERICAN SOCIETY OF HYPERTENSION, 2014, 8 (12) :930-938
[2]  
Association for the Advancement of Medical Instrumentation American National Standards Institute., 2008, ANSIAAMIISO810601200
[3]  
Association for the Advancement of Medical Instrumentation American National Standards Institute., 2013, ANSIAAMIISO810602201
[4]   Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method [J].
Gesche, Heiko ;
Grosskurth, Detlef ;
Kuechler, Gert ;
Patzak, Andreas .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (01) :309-315
[5]   Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring [J].
Kim, Chang-Sei ;
Carek, Andrew M. ;
Mukkamala, Ramakrishna ;
Inan, Omer T. ;
Hahn, Jin-Oh .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (11) :2657-2664
[6]   A Blood Pressure Monitoring Method for Stroke Management [J].
Ma, Heather Ting .
BIOMED RESEARCH INTERNATIONAL, 2014, 2014
[7]   Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure [J].
Mase, Michela ;
Mattei, Walter ;
Cucino, Roberta ;
Faes, Luca ;
Nollo, Giandomenico .
JOURNAL OF ELECTROCARDIOLOGY, 2011, 44 (02) :201-207
[9]  
Nichols WW, 2011, THEORETICAL EXPT CLI, DOI DOI 10.1109/TBME.2015.2440291
[10]   European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults [J].
O'Brien, Eoin ;
Atkins, Neil ;
Stergiou, George ;
Karpettas, Nikos ;
Parati, Gianfranco ;
Asmar, Roland ;
Imai, Yutaka ;
Wang, Jiguang ;
Mengden, Thomas ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :23-38