Comparison of a new video-optical intubation stylet versus the conventional malleable stylet in simulated difficult tracheal intubation

被引:16
作者
Biro, P
Weiss, M
Gerber, A
Pasch, T
机构
[1] Univ Zurich Hosp, Inst Anasthesiol, CH-8091 Zurich, Switzerland
[2] Univ Zurich, Childrens Hosp, Div Anaesthesia, CH-8032 Zurich, Switzerland
关键词
intubation; tracheal; difficult; equipment; intubation stylet;
D O I
10.1046/j.1365-2044.2000.01519.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Handling and efficacy of a new video-optical intubation stylet were assessed in a simulated difficult tracheal intubation setting and compared with a conventional malleable stylet. Forty-five anaesthetists performed 10 tracheal intubations using both techniques. Laryngoscopy was performed by the observer, who created a grade 3 view according the classification by Cormack and Lehane. The time taken to place the tracheal tube and the final tracheal tube positions were documented. Mean (SD) intubation time for the video-optical stylet was 20.4 (7.7) s and for the malleable stylet 10.2 (3.3) s (p < 0.01). With the video-optical stylet the trachea was correctly intubated in all 225 attempts; with the malleable stylet 14 (19.6%) oesophageal and 44 (19.6%) endobronchial intubations occurred (p < 0.01). The video-optical intubation stylet enabled us to recognise inappropriate tracheal tube positions and to correct them immediately. This equipment can be considered a reliable and effective tool for management of the difficult airway.
引用
收藏
页码:886 / 889
页数:4
相关论文
共 13 条
[1]   Use of the fibrescope video camera system for difficult tracheal intubation [J].
Aoyama, K ;
Takenaka, I ;
Sata, T ;
Shigematsu, A .
BRITISH JOURNAL OF ANAESTHESIA, 1996, 77 (05) :662-664
[2]   OPTICAL STYLET - AID TO INTUBATION AND TEACHING [J].
BERCI, G ;
KATZ, R .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1979, 88 (06) :828-831
[3]  
CAPLAN RA, 1993, ANESTHESIOLOGY, V78, P597
[4]   DIFFICULT TRACHEAL INTUBATION IN OBSTETRICS [J].
CORMACK, RS ;
LEHANE, J .
ANAESTHESIA, 1984, 39 (11) :1105-1111
[5]   INCREASED USE OF GUM-ELASTIC BOUGIE IN CLINICAL-PRACTICE [J].
DOGRA, S ;
FALCONERR, R ;
LATTO, IP .
ANAESTHESIA, 1990, 45 (11) :997-998
[6]  
Goldberg J S, 1990, J Clin Anesth, V2, P21, DOI 10.1016/0952-8180(90)90044-4
[7]   Clinical assessment of a plastic optical fiber stylet for human tracheal intubation [J].
Gravenstein, D ;
Melker, RJ ;
Lampotang, S .
ANESTHESIOLOGY, 1999, 91 (03) :648-653
[8]  
Moncure K, 1998, ANESTHESIOLOGY, V89, pU1079
[9]   Swift conversion from laryngoscopic to fiberoptic intubation with a new, handy fiberoptic stylet [J].
Saruki, N ;
Saito, S ;
Sato, J ;
Kon, N ;
Tozawa, R .
ANESTHESIA AND ANALGESIA, 1999, 89 (02) :526-528
[10]   New "seeing" stylet-scope and method for the management of the difficult airway [J].
Shikani, AH .