Commissioning an image-guided localization system for radiotherapy

被引:32
作者
Phillips, MH [1 ]
Singer, K [1 ]
Miller, E [1 ]
Stelzer, K [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Radiat Oncol, Seattle, WA 98195 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2000年 / 48卷 / 01期
关键词
image-guided radiation; radiation target localization; positioning; IMRT; conformal therapy; radiosurgery;
D O I
10.1016/S0360-3016(00)00581-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use. Methods and Materials: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system. Results: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system. Conclusion: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault. (C) 2000 Elsevier Science Inc.
引用
收藏
页码:267 / 276
页数:10
相关论文
共 28 条
[1]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[2]  
BOVA FJ, 1998, MED DOSIM, V23, P21
[3]   Quality assurance system to correct for errors arising from couch rotation in linac-based stereotactic radiosurgery [J].
Brezovich, IA ;
Pareek, PN ;
Plott, WE ;
Jennelle, RLS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :883-890
[4]  
Buatti J M, 1998, Med Dosim, V23, P201, DOI 10.1016/S0958-3947(98)00010-7
[5]   Preliminary experience with frameless stereotactic radiotherapy [J].
Buatti, JM ;
Bova, FJ ;
Friedman, WA ;
Meeks, SL ;
Marcus, RB ;
Mickle, JP ;
Ellis, TL ;
Mendenhall, WM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :591-599
[6]   FROM RADIOTHERAPY TO STEREOTAXIC RADIOSURGERY - PHYSICAL AND DOSIMETRICAL CONSIDERATIONS [J].
CHIEREGO, G ;
FRANCESCON, P ;
COLOMBO, F ;
POZZA, F .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :214-218
[7]   FINE-TUNING OF LINEAR-ACCELERATOR ACCESSORIES FOR STEREOTAXIC RADIOTHERAPY [J].
ENGLER, MJ ;
CURRAN, BH ;
TSAI, JS ;
STERNICK, ES ;
SELLES, WD ;
WAZER, DE ;
MASON, WP ;
SAILOR, T ;
MACKIE, TR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (04) :1001-1008
[8]   THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM [J].
FRIEDMAN, WA ;
BOVA, FJ .
SURGICAL NEUROLOGY, 1989, 32 (05) :334-342
[9]  
Gibbs F. A, 1992, INT J RADIAT ONCOL, V25, P117
[10]   RELOCATABLE FRAME FOR STEREOTAXIC EXTERNAL BEAM RADIOTHERAPY [J].
GILL, SS ;
THOMAS, DGT ;
WARRINGTON, AP ;
BRADA, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (03) :599-603