The Italian affair: The employment of parallel-plate ionization chambers for dose measurements in high dose-per-pulse IORT electron beams

被引:27
作者
Cella, L. [1 ,2 ]
Liuzzi, R. [1 ,2 ]
Salvatore, M. [2 ]
机构
[1] CNR, Inst Biostruct & Bioimages, I-80131 Naples, Italy
[2] Univ Naples Federico 2, Dept Diagnost Imaging & Radiat Oncol, I-80131 Naples, Italy
关键词
free-electron fraction; ion recombination; ionization chamber; IORT; RECOMBINATION CORRECTION; COLLECTION EFFICIENCY; RADIATION; DOSIMETRY;
D O I
10.1118/1.3432601
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: At the high dose-per-pulse rates used by some intraoperative radiotherapy (IORT) units, the employment of ionization chambers for dose measurements needs an appropriate correction (k(sat)) for ion recombination. Through a revision of the existing literature, the authors compared different methods for the determination of the recombination correction factor and their impact on clinical dosimetry. Methods: A dosimetric characterization of IORT electron beams from a Linac Hitesys Novac7 (Aprilia-Latina, Italy) was performed. Dose-to-water (D(w)) values were measured with dose-per-pulse independent chemical dosimeters (operated by the Italian Primary Standard Dosimetry Laboratory, ENEA) and compared to doses obtained by two different parallel-plate ionization chamber models (Markus and Advanced Markus, PTW, Freiburg, Germany). For dose measurements using ionization chambers, the authors applied two different methods for the determination of the ion recombination correction factor (k(sat)), as suggested in previous articles. The first method is based on the experimental estimation of the free-electron fraction values p; the second one is based on the "nonstandard" two-voltage analysis including the free-electron component. Results: For a Markus type chamber, there is a good agreement between the results on k(sat) and those reported in the literature for both methods, while for the Advanced Markus chamber, no data are available for comparison. Conclusions: Comparing values of D(w) obtained by dose-per-pulse independent dosimeters and by ionization chambers measurements corrected using the two different approaches, the authors can conclude that the k(sat) factor determination is very critical and that only an experimental protocol using ionization chamber intercalibration can be considered reliable. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3432601]
引用
收藏
页码:2918 / 2924
页数:7
相关论文
共 17 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM radiation therapy committee task group no. 72 [J].
Beddar, A. Sam ;
Biggs, Peter J. ;
Chang, Sha ;
Ezzell, Gary A. ;
Faddegon, Bruce A. ;
Hensley, Frank W. ;
Mills, Michael D. .
MEDICAL PHYSICS, 2006, 33 (05) :1476-1489
[3]  
Boag J.W., 1987, The Dosimetry of Ionizing Radiation
[5]   The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation [J].
Boag, JW ;
Hochhauser, E ;
Balk, OA .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05) :885-897
[6]  
BRAHME A, 1981, STRAHLENTHER ONKOL, V157, P258
[7]   On measuring the output of an IORT mobile dedicated accelerator [J].
De Angelis, C. ;
Soriani, A. ;
Benassi, M. ;
Onori, S. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :221-225
[8]   Ion recombination correction for very high dose-per-pulse high-energy electron beams [J].
Di Martino, F ;
Giannelli, M ;
Traino, AC ;
Lazzeri, M .
MEDICAL PHYSICS, 2005, 32 (07) :2204-2210
[9]   Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse [J].
Fiandra, Christian ;
Ragona, Riccardo ;
Ricardi, Umberto ;
Anglesio, Silvia ;
Giglioli, Francesca Romana .
MEDICAL PHYSICS, 2008, 35 (12) :5463-5470
[10]   THE SIGNIFICANCE OF THE LIFETIME AND COLLECTION TIME OF FREE-ELECTRONS FOR THE RECOMBINATION CORRECTION IN THE IONOMETRIC DOSIMETRY OF PULSED RADIATION [J].
HOCHHAUSER, E ;
BALK, OA ;
SCHNEIDER, H ;
ARNOLD, W .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1994, 27 (03) :431-440