An Active Timing Controlled-gate Fluorescence Lifetime Imaging Microscopy Using an Ultrafast Picosecond Diode Laser

被引:0
作者
Liu, Yang [1 ]
Zhou, Yan [1 ]
Liu, Yuliang [1 ]
机构
[1] Chinese Acad Sci, Inst Semicond, Optoelect Syst Lab, Beijing 100083, Peoples R China
来源
OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS VI | 2014年 / 9268卷
关键词
active time-gated; FLIM; lifetime determination; low-jitter; time-resolved fluorescence; IN-VIVO;
D O I
10.1117/12.2071342
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report a time-gated fluorescence lifetime imaging microscopy (FLIM) using an active timing-controlled device and an ultrafast picosecond diode laser. This system could utilize more precise synchronization between laser excitation pulse and ICCD gate combining low timing-jitter diode laser with high precise timing control unit. Two flexible trigger signals for laser source and intensifier gate could adjust time sequence with picosecond precision compared with conventional passive delay adjustment. The importance of the adjustable repetition rate of this laser source is discussed with respect to noise reduction and precision in the lifetime determination, illustrating a further significant advantage over conventional mode-locked solid-state lasers. This FLIM system instrument could satisfy a wide temporal dynamic range demand for various biological applications.
引用
收藏
页数:6
相关论文
共 12 条
[1]   High frame rate fluorescence lifetime imaging [J].
Agronskaia, AV ;
Tertoolen, L ;
Gerritsen, HC .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2003, 36 (14) :1655-1662
[2]  
Chang C.-W., 2009, IMPROVING PRECISION, P7370
[3]   Fluorescence lifetime imaging microscopy [J].
Chang, Ching-Wei ;
Sud, Dhruv ;
Mycek, Mary-Ann .
DIGITAL MICROSCOPY, 3RD EDITION, 2007, 81 :495-+
[4]   Enhancing precision in time-domain fluorescence lifetime imaging [J].
Chang, Ching-Wei ;
Mycek, Mary-Ann .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (05)
[5]  
Elson D, 2002, OPT PHOTONICS NEWS, V13, P26, DOI 10.1364/OPN.13.11.000026
[6]   Fluorescence lifetime system for microscopy and multiwell plate imaging with a blue picosecond diode laser [J].
Elson, DS ;
Siegel, J ;
Webb, SED ;
Lévêque-Fort, S ;
Lever, MJ ;
French, PMW ;
Lauritsen, K ;
Wahl, M ;
Erdmann, R .
OPTICS LETTERS, 2002, 27 (16) :1409-1411
[7]   Fluorescence lifetime imaging using a compact, low-cost, diode-based all-solid-state regenerative amplifier [J].
Mendez, E ;
Elson, DS ;
Koeberg, M ;
Dunsby, C ;
Bradley, DDC ;
French, PMW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (05) :1264-1267
[8]  
Sud D., 2008, WIDE FIELD TIME DOMA
[9]   Image restoration for fluorescence lifetime imaging microscopy (FLIM) [J].
Sud, Dhruv ;
Mycek, Mary-Ann .
OPTICS EXPRESS, 2008, 16 (23) :19192-19200
[10]   Fluorescence lifetime imaging microscopy: in vivo application to diagnosis of oral carcinoma [J].
Sun, Yinghua ;
Phipps, Jennifer ;
Elson, Daniel S. ;
Stoy, Heather ;
Tinling, Steven ;
Meier, Jeremy ;
Poirier, Brian ;
Chuang, Frank S. ;
Farwell, D. Gregory ;
Marcu, Laura .
OPTICS LETTERS, 2009, 34 (13) :2081-2083