Analysis of Implant Strength After Implantoplasty in Three Implant-Abutment Connection Designs: An In Vitro Study

被引:28
作者
Alexandre Gehrke, Sergio [1 ,2 ]
Aramburu Junior, Jaime Sarda [1 ,3 ]
Dedavid, Berenice Anina [4 ]
Shibli, Jamil Awad [5 ]
机构
[1] Biotecnos Technol & Sci, Santa Maria, RS, Brazil
[2] Catholic Univ Murcia, Master Course Lab & Clin Res, Murcia, Spain
[3] Fac Itapiranga, Canarias, Spain
[4] Pontificia Univ Catolica Rio Grande do Sul, Dept Mat Engn, Postgrad Program, Porto Alegre, RS, Brazil
[5] Univ Guarulhos, Dent Res Div, Dept Periodontol & Oral Implantol, Postgrad Program Implantol, Guarulhos, SP, Brazil
关键词
dental implant; fracture mode; fracture strength; implant connection; implantoplasty; 3-YEAR CLINICAL-TRIAL; SHAPED ORAL IMPLANTS; PERI-IMPLANTITIS; SURGICAL THERAPY; RESECTIVE SURGERY; EUROPEAN WORKSHOP; BONE; PREVALENCE; CONSENSUS; SEVERITY;
D O I
10.11607/jomi.4399
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of this in vitro study was to assess the resistance to static fatigue of implants with different connections before and after implantoplasty. Materials and Methods: Sixty conical implants and 60 abutments were used; 4-mm-diameter versions were available for each model. Three groups (n = 20) were established based on the following implant connections: external hexagon (group 1), internal hexagon (group 2), and Morse taper (group 3). The implants of each group were submitted to a compressive load before (n = 10) and after the implantoplasty (n = 10). The wear was performed in a mechanical lathe machine using a carbide bur, and the final dimensions of each sample were measured. All groups were subjected to quasi-static loading at a 30-degree angle to the implant axis in a universal testing machine and 5 mm out of the implant support. Results: After the implantoplasty, the mean final diameter was 3.13 +/- 0.033 mm for group 1, 3.23 +/- 0.023 mm for group 2, and 3.25 +/- 0.03 mm for group 3. The mean fracture strengths for the groups before and after the implantoplasty were, respectively, 773.1 +/- 13.16 N and 487.1 +/- 93.72 N in group 1; 829.4 +/- 14.12 N and 495.7 +/- 85.24 N in group 2; and 898.1 +/- 19.25 N and 717.6 +/- 77.25 N in group 3. Conclusion: Resistance to loading decreased significantly after implantoplasty, and varied among the three implant connection designs.
引用
收藏
页码:E65 / E70
页数:6
相关论文
共 38 条
[1]  
ALBREKTSSON T, 1994, PROCEEDINGS OF THE 1ST EUROPEAN WORKSHOP ON PERIODONTOLOGY, P365
[2]  
[Anonymous], ORAL HLTH DENT MANAG
[3]   Significance of keratinized mucosa in maintenance of dental implants with different surfaces [J].
Chung, Dyeus M. ;
Oh, Tae-Ju ;
Shotwell, Jeffrey L. ;
Misch, Carl E. ;
Wang, Hom-Lay .
JOURNAL OF PERIODONTOLOGY, 2006, 77 (08) :1410-1420
[4]   Surgical treatment of peri-implantitis [J].
Claffey, Noel ;
Clarke, Emily ;
Polyzois, Ioannis ;
Renvert, Stefan .
JOURNAL OF CLINICAL PERIODONTOLOGY, 2008, 35 :316-332
[5]  
Deppe H, 2007, INT J ORAL MAX IMPL, V22, P79
[6]   Prevalence of subjects with progressive bone loss at implants [J].
Fransson, C ;
Lekholm, U ;
Jemt, T ;
Berglundh, T .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) :440-446
[7]   Severity and pattern of peri-implantitis-associated bone loss [J].
Fransson, Christer ;
Tomasi, Cristiano ;
Pikner, Solveig Sunden ;
Grondahl, Kerstin ;
Wennstrom, Jan L. ;
Leyland, Alastair H. ;
Berglundh, Tord .
JOURNAL OF CLINICAL PERIODONTOLOGY, 2010, 37 (05) :442-448
[8]   Influence of bone insertion level of the implant on the fracture strength of different connection designs: an in vitro study [J].
Gehrke, Sergio Alexandre ;
dos Santos Vianna, Miriam Souza ;
Dedavid, Berenice Anina .
CLINICAL ORAL INVESTIGATIONS, 2014, 18 (03) :715-720
[9]   Anti-infective surgical therapy of peri-implantitis. A 12-month prospective clinical study [J].
Heitz-Mayfield, L. J. A. ;
Salvi, G. E. ;
Mombelli, A. ;
Faddy, M. ;
Lang, N. P. .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (02) :205-210
[10]  
International Organization for Standardization, 2007, ISO 14801 DENT IMPL