Fast reconstruction with uniform noise properties in half scan computed tomography

被引:9
作者
Pan, XC [1 ]
机构
[1] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
computed tomography; halfscan; fan-beam projection; image reconstruction; helical CT;
D O I
10.1118/1.1288239
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The hybrid algorithms developed recently for the reconstruction of fan-beam images possess computational and noise properties superior to those of the fan-beam filtered backprojection (FFBP) algorithm. However, the hybrid algorithms cannot be applied directly to a halfscan fan-beam sinogram because they require knowledge of a fullscan fan-beam sinogram. In this work, we developed halfscan-hybrid algorithms for image reconstruction in halfscan computed tomography (CT). Numerical evaluation indicates that the proposed halfscan-hybrid algorithms are computationally more efficient than are the widely used halfscan-FFBP algorithms. Also, the results of quantitative studies demonstrated clearly that the noise levels in images reconstructed by use of the halfscan-hybrid algorithm are generally lower and spatially more uniform than are those in images reconstructed by use of the halfscan-FFBP algorithm. Such reduced and uniform image noise levels may be translated into improvement of the accuracy and precision of lesion detection and parameter estimation in noisy CT images without increasing the radiation dose to the patient. Therefore, the halfscan-hybrid algorithms may have significant implication for image reconstruction in conventional and helical CT. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00109-7].
引用
收藏
页码:2031 / 2036
页数:6
相关论文
共 24 条
[1]  
[Anonymous], 1976, COMPUT VISION GRAPH
[2]  
[Anonymous], 2006, J ICRU
[3]   THE RADON-TRANSFORM AND ITS APPLICATIONS [J].
BARRETT, HH .
PROGRESS IN OPTICS, 1984, 21 :217-286
[4]   COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[5]   Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies [J].
Gies, M ;
Kalender, WA ;
Wolf, H ;
Suess, C ;
Madsen, MT .
MEDICAL PHYSICS, 1999, 26 (11) :2235-2247
[6]   RECONSTRUCTION OF FAN-BEAM DATA BY FILTERING THE BACK-PROJECTION [J].
GULLBERG, GT .
COMPUTER GRAPHICS AND IMAGE PROCESSING, 1979, 10 (01) :30-47
[7]   RECONSTRUCTION ALGORITHM FOR FAN BEAM WITH A DISPLACED CENTER-OF-ROTATION [J].
GULLBERG, GT ;
CRAWFORD, CR ;
TSUI, BMW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (01) :23-29
[8]   A simple theorem relating noise anti patient dose in computed tomography [J].
Harpen, MD .
MEDICAL PHYSICS, 1999, 26 (11) :2231-2234
[9]  
HERMAN GT, 1980, IMAGE RECONSTRUCTION
[10]  
HORN BK, 1977, P IEEE, V67, P1616