Medium- to Long-Term Outcomes after Reverse Total Shoulder Arthroplasty with a Standard Long Stem

被引:7
作者
Buelhoff, Matthias [1 ]
Zeifang, Felix [2 ]
Welters, Caroline [3 ]
Renkawitz, Tobias [1 ]
Schiltenwolf, Marcus [1 ]
Tross, Anna-K [1 ]
机构
[1] Heidelberg Univ Hosp, Clin Orthopaed Surg, Schlierbacher Landstr 200a, D-69118 Heidelberg, Germany
[2] Ethianum Clin Heidelberg, Vossstr 6, D-69115 Heidelberg, Germany
[3] Munich Municipal Hosp Grp, Clin Dermatol, Thalkirchnerstr 48, D-80337 Munich, Germany
关键词
shoulder; reverse arthroplasty; polyethylene wear; humeral loosening; glenoid loosening; scapular notching; PATIENTS YOUNGER; GLENOHUMERAL OSTEOARTHRITIS; CUFF; PROSTHESIS; DESIGN; SURVIVORSHIP; ARTHRITIS; COMPONENT; REVISION; SCORE;
D O I
10.3390/jcm11092274
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Long-term clinical and radiographic outcome data after standard cemented long-stem reverse shoulder arthroplasty (RSA) remain underreported. The aim of this study is to report on medium- to long-term data of patients over 60 years of age. Methods: The same type of RSA (Aequalis Reverse II, Memphis, TN, USA) was implanted in 27 patients with a mean age of 73 years (range 61-84). Indications for RSA were cuff tear arthropathy (CTA) in 25 cases and osteoarthritis (OA) in two cases. Pre- and postoperative Constant Score was assessed and component loosening, polyethylene wear, scapular notching and revision rates were recorded at a mean clinical follow-up (FU) of 127.6 months (SD +/- 33.7; range 83-185). Results: The mean-adjusted CS (aCS) improved from 30.0 (range 10-59) to 95.0 (range 33-141) points (p < 0.001). Glenoid loosening was found in two (9.1%) and stem loosening was found in three (13.6%) cases. Polyethylene wear was observed in four (18.2%) cases. Scapular notching appeared in 15 (68.2%) cases but was not associated with poor aCS (p = 0.423), high levels of pain (p = 0.798) or external rotation (p = 0.229). Revision surgery was necessary in three (11.1%) cases. Conclusions: RSA with a cemented standard long stem leads to improvement in forward elevation, abduction and pain after a mean FU of 10 years. However, external rotation does not improve with this prosthetic design. Moreover, scapular notching is observed in the majority of cases, and revision rates (11.1%) as well as humeral loosening rates (13.6%) remain a concern. Level of evidence: Level 4, retrospective cohort study.
引用
收藏
页数:11
相关论文
共 43 条
[1]   Reverse shoulder arthroplasty in 41 patients with cuff tear arthropathy with a mean follow-up period of 5 years [J].
Al-Hadithy, Nawfal ;
Domos, Peter ;
Sewell, Mathew D. ;
Pandit, Ravi .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2014, 23 (11) :1662-1668
[2]  
Australian Orthopaedic Association, 2020, National joint replacement registry annual reports
[3]   Influence of glenoid component design and humeral component retroversion on internal and external rotation in reverse shoulder arthroplasty: A cadaver study [J].
Berhouet, J. ;
Garaud, P. ;
Favard, L. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2013, 99 (08) :887-894
[4]  
Boehm D, 2004, UNFALLCHIRURG, V107, P397, DOI 10.1007/s00113-004-0757-3
[5]   Grammont reverse prosthesis: Design, rationale, and biomechanics [J].
Boileau, P ;
Watkinson, DJ ;
Hatzidakis, AM ;
Balg, F .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (01) :147S-161S
[6]   Neer Award 2005: The Grammont reverse shoulder prosthesis: Results in cuff tear arthritis, fracture sequelae, and revision arthroplasty [J].
Boileau, Pascal ;
Watkinson, Duncan ;
Hatzidakis, Armodios M. ;
Hovorka, Istvan .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2006, 15 (05) :527-540
[7]   Bony increased-offset-reverse shoulder arthroplasty: 5 to 10 years' follow-up [J].
Boileau, Pascal ;
Morin-Salvo, Nicolas ;
Bessiere, Charles ;
Chelli, Mikael ;
Gauci, Marc-Olivier ;
Lemmex, Devin B. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2020, 29 (10) :2111-2122
[8]   Reverse total shoulder arthroplasty after failed rotator cuff surgery [J].
Boileau, Pascal ;
Gonzalez, Jean-Francois ;
Chuinard, Christopher ;
Bicknell, Ryan ;
Walch, Gilles .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2009, 18 (04) :600-606
[9]   Factors Influencing Appropriate Implant Selection and Position in Reverse Total Shoulder Arthroplasty [J].
Callegari, Jonathan ;
Haidamous, Georges ;
Laedermann, Alexandre ;
Phillips, Cameron ;
Tracy, Shane ;
Denard, Patrick .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2021, 52 (02) :157-166
[10]   Revision Reverse Shoulder Arthroplasty [J].
Chalmers, Peter N. ;
Boileau, Pascal ;
Romeo, Anthony A. ;
Tashjian, Robert Z. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2019, 27 (12) :426-436