Direct electrothermal atomic absorption spectrometry determination of nickel in sea water using multiple hot injection and Zeeman correction

被引:10
作者
Bermejo-Barrera, P [1 ]
Moreda-Pineiro, J [1 ]
Moreda-Pineiro, A [1 ]
Bermejo-Barrera, A [1 ]
机构
[1] Univ Santiago de Compostela, Fac Chem, Dept Analyt Chem Nutr & Bromatol, Santiago De Compostela 15706, Spain
关键词
nickel; sea water; ETAAS; multiple injection; hot injection; Zeeman correction;
D O I
10.1016/S0039-9140(97)00159-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Methods for the direct determination of Ni in sea water samples by ETAAS were developed using Zeeman effect background correction system (ZEBC) and a multi-injection technique. A mass of palladium nitrate of 2.5 mu g (for an injection volume of 100 mu l) was used as chemical modifier. The optimum pyrolysis and atomization temperatures were 1700 and 2100 degrees C, respectively. The characteristic mass (m(0)) and characteristic concentration (C-0), precision and accuracy were studied for different injection volumes (20, 100 and 200 mu l). For an injection volume of 100 mu l (five 20 mu l aliquot) of sample the accuracy analysis of different certified materials (saline and non saline water) was agreeable. The total time of the proposed procedure is 6 min. A m(0) and C-0 of 34.5 pg and 0.3 mu g l(-1), respectively were obtained for this injection volume (100 mu l). Finally, interferences from major and minor components of sea water was studied. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:807 / 815
页数:9
相关论文
共 35 条