Optimization and validation of an ion chromatographic method for the simultaneous determination of sodium, ammonium and potassium in exhaled breath condensate
被引:15
作者:
Svensson, S
论文数: 0引用数: 0
h-index: 0
机构:Sahlgrens Univ Hosp, Occupat & Environm Med, SE-41266 Gothenburg, Sweden
Svensson, S
Isacsson, AC
论文数: 0引用数: 0
h-index: 0
机构:Sahlgrens Univ Hosp, Occupat & Environm Med, SE-41266 Gothenburg, Sweden
Isacsson, AC
Ljungkvist, G
论文数: 0引用数: 0
h-index: 0
机构:Sahlgrens Univ Hosp, Occupat & Environm Med, SE-41266 Gothenburg, Sweden
Ljungkvist, G
论文数: 引用数:
h-index:
机构:
Torén, K
Olin, AC
论文数: 0引用数: 0
h-index: 0
机构:Sahlgrens Univ Hosp, Occupat & Environm Med, SE-41266 Gothenburg, Sweden
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
|
2005年
/
814卷
/
01期
关键词:
ion chromatography;
exhaled breath condensate;
sodium;
ammonium;
potassium;
chemometrics;
D O I:
10.1016/j.jchromb.2004.10.018
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
An ion chromatographic method with conductivity detection for the simultaneous quantification of sodium, ammonium and potassium in exhaled breath condensate (EBC) was developed and validated. A factorial design was used to optimize the chromatographic conditions. which resulted in baseline separations of the cations within 6 min. The method requires no pre-treatment of EBC samples. The optimized method was used for the intra-individual screening of cations in EBC of 10 healthy volunteers. The LOQs were low (0.3. 0.1 and 0.2 muM for sodium, ammonium and potassium, respectively), compared with levels detected in healthy volunteers. The responses were linear with good precision, and samples could be stored for at least 10 weeks at refrigerating conditions. (C) 2004 Elsevier B.V. All rights reserved.