3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry

被引:153
作者
Mans, Anton [1 ]
Remeijer, Peter [1 ]
Olaciregui-Ruiz, Igor [1 ]
Wendling, Markus [1 ]
Sonke, Jan-Jakob [1 ]
Mijnheer, Ben [1 ]
van Herk, Marcel [1 ]
Stroom, Joep C. [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
关键词
Portal dosimetry; Volumetric-modulated arc therapy; In vivo dosimetry; VIVO EPID DOSIMETRY; CONE-BEAM CT; TREATMENT DELIVERY; DOSE VERIFICATION; IMRT TREATMENTS; PROSTATE IMRT; CANCER; RADIOTHERAPY; ALGORITHM; GUIDANCE;
D O I
10.1016/j.radonc.2009.12.020
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: To demonstrate the feasibility of back-projection portal dosimetry for accurate 3D dosimetric verification of volumetric-modulated arc therapy (VMAT), pre-treatment as well as in vivo. Materials and methods: Several modifications to our existing approach were implemented to make the method applicable to VMAT: (i) gantry angle-resolved data acquisition, (ii) calculation of the patient transmission, (iii) compensation for detector 'flex' and (iv) 3D dose reconstruction and evaluation. Results: Planned and EPID-(Electronic Portal Image Detector)-reconstructed close distributions show good agreement for pre-treatment verification of two prostate, a stereotactic lung and a head-and-neck VMAT plan and for in vivo verification of VMAT treatments of prostate and lung cancer. Averaged over pre-treatment verifications, planned and measured isocentre dose ratios were -1.2% (range [-4.7%,1.8%]). 3D gamma analysis (3% maximum dose, 3 mm) revealed mean gamma <gamma(mean)> = 0.37 [0.34,0.39], maximum 1% gamma (gamma(1%)) = 0.72 [0.66,0.81] and percentage of points with -gamma <= 1 < P(gamma <= 1)> = 99% [97%,100%.] For in vivo verification, the average isocentre dose ratio was -1.2% [-0.8%,-1.7%], <gamma(mean)> = 0.52 [0.40,0.64], <gamma(1%)> = 0.92 [0.76,1.08] and < P(gamma <= 1)> = 96% [93%,100%]. Conclusions: Our portal dosimetry method was successfully adapted for verification of VMAT treatments, pre-treatment as well as in vivo. (C) 2010 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 94 (2010) 181-187
引用
收藏
页码:181 / 187
页数:7
相关论文
共 23 条
[1]   Treatment of lung cancer using volumetric modulated arc therapy and image guidance: A case study [J].
Bedford, James L. ;
Hansen, Vibeke Nordmark ;
Mcnair, Helen A. ;
Aitken, Alexandra H. ;
Brock, Juliet E. C. ;
Warrington, Alan P. ;
Brada, Michael .
ACTA ONCOLOGICA, 2008, 47 (07) :1438-1443
[2]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[3]   Development and evaluation of an efficient approach to volumetric arc therapy planning [J].
Bzdusek, Karl ;
Friberger, Henrik ;
Eriksson, Kjell ;
Hardemark, Bjorn ;
Robinson, David ;
Kaus, Michael .
MEDICAL PHYSICS, 2009, 36 (06) :2328-2339
[4]   Accurate technique for complete geometric calibration of cone-beam computed tomography systems [J].
Cho, YB ;
Moseley, DJ ;
Siewerdsen, JH ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (04) :968-983
[5]   Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? [J].
Guckenberger, Matthias ;
Richter, Anne ;
Krieger, Thomas ;
Wilbert, Juergen ;
Baier, Kurt ;
Flentje, Michael .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :259-265
[6]   RapidArc volumetric modulated therapy planning for prostate cancer patients [J].
Kjaer-Kristoffersen, Flemming ;
Ohlhues, Lars ;
Medin, Joakim ;
Korreman, Stine .
ACTA ONCOLOGICA, 2009, 48 (02) :227-232
[7]   Dosimetric verification of RapidArc treatment delivery [J].
Korreman, Stine ;
Medin, Joakim ;
Kjaer-Kristoffersen, Flemming .
ACTA ONCOLOGICA, 2009, 48 (02) :185-191
[8]   Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT [J].
McCurdy, B. M. C. ;
Greer, P. B. .
MEDICAL PHYSICS, 2009, 36 (07) :3028-3039
[9]   Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification [J].
McDermott, L. N. ;
Wendling, M. ;
van Asselen, B. ;
Stroom, J. ;
Sonke, J. -J. ;
van Herk, M. ;
Mijnheer, B. J. .
MEDICAL PHYSICS, 2006, 33 (10) :3921-3930
[10]   3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT [J].
McDermott, Leah N. ;
Wendling, Markus ;
Nijkamp, Jasper ;
Mans, Anton ;
Sonke, Jan-Jakob ;
Mijnheer, Ben J. ;
van Herk, Marcel .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :35-42