In vitro validation of a self-driving aortic-turbine venous-assist device for Fontan patients

被引:19
作者
Pekkan, Kerem [1 ]
Aka, Ibrahim Basar [1 ]
Tutsak, Ece [1 ]
Ermek, Erhan [1 ]
Balim, Haldun [1 ]
Lazoglu, Ismail [1 ]
Turkoz, Riza [2 ]
机构
[1] Koc Univ, Dept Mech Engn, Rumeli Feneri Campus, Istanbul, Turkey
[2] Acibadem Univ, Dept Cardiovasc Surg, Sch Med, Istanbul, Turkey
基金
欧洲研究理事会;
关键词
pediatric ventricle assist device; venous hemodynamics; congenital heart disease; single-ventricle physiology; hemodynamics; mock-up circulation tests; Fontan; MECHANICAL CIRCULATORY SUPPORT; LEFT-HEART SYNDROME; CAVOPULMONARY ASSIST; TRICUSPID-ATRESIA; SINGLE-VENTRICLE; SHUNT CONFIGURATIONS; SURGICAL REPAIR; UNITED-STATES; PULSATILE; MODEL;
D O I
10.1016/j.jtcvs.2018.02.088
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Palliative repair of single ventricle defects involve a series of open-heart surgeries where a single-ventricle (Fontan) circulation is established. As the patient ages, this paradoxical circulation gradually fails, because of its high venous pressure levels. Reversal of the Fontan paradox requires an extra subpulmonic energy that can be provided through mechanical assist devices. The objective of this study was to evaluate the hemodynamic performance of a totally implantable integrated aortic-turbine venous-assist (iATVA) system, which does not need an external drive power and maintains low venous pressure chronically, for the Fontan circulation. Methods: Blade designs of the co-rotating turbine and pump impellers were developed and 3 prototypes were manufactured. After verifying the single-ventricle physiology at a pulsatile in vitro circuit, the hemodynamic performance of the iATVA system was measured for pediatric and adult physiology, varying the aortic steal percentage and circuit configurations. The iATVA system was also tested at clinical off-design scenarios. Results: The prototype iATVA devices operate at approximately 800 revolutions per minute and extract up to 10% systemic blood from the aorta to use this hydrodynamic energy to drive a blood turbine, which in turn drives a mixed-flow venous pump passively. By transferring part of the available energy from the single-ventricle outlet to the venous side, the iATVA system is able to generate up to approximately 5 mm Hg venous recovery while supplying the entire caval flow. Conclusions: Our experiments show that a totally implantable iATVA system is feasible, which will eliminate the need for external power for Fontan mechanical venous assist and combat gradual postoperative venous remodeling and Fontan failure.
引用
收藏
页码:292 / +
页数:17
相关论文
共 63 条
  • [21] Infection in ventricular assist devices: Prevention and treatment
    Holman, WL
    Rayburn, BK
    McGiffin, DC
    Foley, BA
    Benza, RL
    Bourge, RC
    Pinderski, LJ
    Kirklin, JK
    [J]. ANNALS OF THORACIC SURGERY, 2003, 75 (06) : S48 - S57
  • [22] Fontan Conversion Templates: Patient-Specific Hemodynamic Performance of the Lateral Tunnel Versus the Intraatrial Conduit With Fenestration
    Hong, Haifa
    Dur, Onur
    Zhang, Haibo
    Zhu, Zhongqun
    Pekkan, Kerem
    Liu, Jinfen
    [J]. PEDIATRIC CARDIOLOGY, 2013, 34 (06) : 1447 - 1454
  • [23] Rescue cardiac transplantation for failing staged palliation in patients with hypoplastic left heart syndrome
    Jacobs, Jeffrey Phillip
    Quintessenza, James Anthony
    Chai, Paul J.
    Lindberg, Harald L.
    Asante-Korang, Alfred
    McCormack, Jorge
    Dadlani, Gul
    Boucek, Robert J.
    [J]. CARDIOLOGY IN THE YOUNG, 2006, 16 (06) : 556 - 562
  • [24] Seven-year clinical experience with the extracardiac pedicled pericardial Fontan operation
    Kavarana, MN
    Pagni, S
    Recto, MR
    Sobczyk, WL
    Yeh, T
    Mitchell, M
    Austin, EH
    [J]. ANNALS OF THORACIC SURGERY, 2005, 80 (01) : 37 - 43
  • [25] DIMINISHED VENOUS VASCULAR CAPACITANCE IN PATIENTS WITH UNIVENTRICULAR HEARTS AFTER THE FONTAN OPERATION
    KELLEY, JR
    MACK, GW
    FAHEY, JT
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (03) : 158 - 163
  • [26] Improving results of the Fontan procedure in patients with heterotaxy syndrome
    Kim, Soo-Jin
    Kim, Woong-Han
    Lim, Hong Gook
    Lee, Chang-Ha
    Lee, Jae Young
    [J]. ANNALS OF THORACIC SURGERY, 2006, 82 (04) : 1245 - 1251
  • [27] Pulsatile venous waveform quality in Fontan circulation-clinical implications, venous assists options and the future
    Kocyildirim, Ergin
    Dur, Onur
    Soran, Ozlem
    Tuzun, Egemen
    Miller, Matthew W.
    Housler, Greggory J.
    Wearden, Peter D.
    Fossum, Theresa W.
    Morell, Victor O.
    Pekkan, Kerem
    [J]. ANATOLIAN JOURNAL OF CARDIOLOGY, 2012, 12 (05) : 420 - 426
  • [28] KREUTZER G, 1973, J THORAC CARDIOV SUR, V66, P613, DOI 10.1016/s0022-5223(19)40598-9
  • [29] An Artificial Right Ventricle for Failing Fontan: In Vitro and Computational Study
    Lacour-Gayet, Francois G.
    Lanning, Craig J.
    Stoica, Serban
    Wang, Rui
    Rech, Bryan A.
    Goldberg, Steven
    Shandas, Robin
    [J]. ANNALS OF THORACIC SURGERY, 2009, 88 (01) : 170 - 176
  • [30] Lucke L, 2014, IEEE ENG MED BIO, P286, DOI 10.1109/EMBC.2014.6943585