A diamond target for megavoltage cone-beam CT

被引:18
|
作者
Sawkey, D. [1 ]
Lu, M. [1 ]
Morin, O. [1 ]
Aubin, M. [1 ]
Yom, S. S. [1 ]
Gottschalk, A. R. [1 ]
Bani-Hashemi, A. [2 ]
Faddegon, B. A. [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, Ctr Comprehens Canc, San Francisco, CA 94143 USA
[2] Siemens Healthcare Sector, Concord, CA 94520 USA
关键词
biomedical equipment; computerised tomography; diamond; dosimetry; linear accelerators; Monte Carlo methods; X-ray imaging; X-RAY-BEAMS; MONTE-CARLO; COMPUTED-TOMOGRAPHY; SIMULATION; SYSTEM; IMPLEMENTATION; LINE;
D O I
10.1118/1.3302831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A sintered diamond target was used in place of a graphite target as part of an imaging beam line (an unflattened beam from a graphite target) installed on a linear accelerator. The diamond target, with a greater density than the graphite target, permitted imaging at the lower beam energy (4 MV) required with the graphite target and the higher beam energy (6 MV) conventionally used with the tungsten/stainless steel target and stainless steel flattening filter. Images of phantoms and patients were acquired using the different beam lines and compared. The beam spectra and dose distributions were determined using Monte Carlo simulation. Results: The diamond target allowed use of the same beam energy as for treatment, simplifying commissioning and quality assurance. Images acquired with the diamond target at 4 MV were similar to those obtained with the graphite target at 4 MV. The slight reduction in low energy photons due to the higher-Z sintering material in the diamond target had minimal effect on image quality. Images acquired at 6 MV with the diamond target showed a small decrease in contrast-to-noise ratio, resulting from a decrease in the fraction of photons in the beam in the energy range to which the detector is most sensitive. Conclusions: The diamond target provides images of a similar quality to the graphite target. Diamond allows use of the higher beam energy conventionally used for treatment, provides a higher dose rate for the same beam current, and potentially simplifies installation and maintenance of the beam line.
引用
收藏
页码:1246 / 1253
页数:8
相关论文
共 50 条
  • [41] An interval subdividing based method for geometric calibration of cone-beam CT
    Tan, Shengqi
    Cong, Peng
    Liu, Ximing
    Wu, Zhifang
    NDT & E INTERNATIONAL, 2013, 58 : 49 - 55
  • [42] Development and applicability of a quality control phantom for dental cone-beam CT
    Pauwels, Ruben
    Stamatakis, Harry
    Manousaridis, Giorgos
    Walker, Adrian
    Michielsen, Koen
    Bosmans, Hilde
    Bogaerts, Ria
    Jacobs, Reinhilde
    Horner, Keith
    Tsiklakis, Kostas
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (04): : 245 - 260
  • [43] Interleaved acquisition for cross scatter avoidance in dual cone-beam CT
    Giles, William
    Bowsher, James
    Li, Hao
    Yin, Fang-Fang
    MEDICAL PHYSICS, 2012, 39 (12) : 7719 - 7728
  • [44] Quantitative cone-beam CT reconstruction with polyenergetic scatter model fusion
    Mason, Jonathan H.
    Perelli, Alessandro
    Nailon, William H.
    Davies, Mike E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (22)
  • [45] Stable and Robust Geometric Self-Calibration for Cone-Beam CT Using Mutual Information
    Muders, Jens
    Hesser, Juergen
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (01) : 202 - 217
  • [46] Evaluation of Dose Calculation Based On Cone-Beam CT
    Abe, T.
    Tateoka, K.
    Fujimoto, K.
    Nakazawa, T.
    Saitou, Y.
    Nakata, A.
    Yano, M.
    Sakata, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 259 - 259
  • [47] The contribution of Cone-Beam CT in the assessment of orofacial pain
    Langlais, R. -P.
    Mah, P.
    Goulet, J. -P.
    DOULEUR ET ANALGESIE, 2009, 22 (02): : 112 - 120
  • [48] Dedicated Cone-Beam CT System for Extremity Imaging
    Carrino, John A.
    Al Muhit, Abdullah
    Zbijewski, Wojciech
    Thawait, Gaurav K.
    Stayman, J. Webster
    Packard, Nathan
    Senn, Robert
    Yang, Dong
    Foos, David H.
    Yorkston, John
    Siewerdsen, Jeffrey H.
    RADIOLOGY, 2014, 270 (03) : 816 - 824
  • [49] Lesions of the jaws presenting as radiolucencies on cone-beam CT
    MacDonald, D.
    CLINICAL RADIOLOGY, 2016, 71 (10) : 972 - 985
  • [50] Noise suppression in scatter correction for cone-beam CT
    Zhu, Lei
    Wang, Jing
    Xing, Lei
    MEDICAL PHYSICS, 2009, 36 (03) : 741 - 752