共 26 条
Accuracy of Scheimpflug Holladay Equivalent Keratometry Readings after Corneal Refractive Surgery in the Absence of Clinical History
被引:3
作者:
Bialer, Omer Y.
[1
,3
]
Kaiserman, Igor
[2
,4
]
Bahar, Irit
[1
,3
]
机构:
[1] Rabin Med Ctr, Dept Ophthalmol, IL-49100 Petah Tiqwa, Israel
[2] Barzilai Govt Hosp, Dept Ophthalmol, Ashqelon, Israel
[3] Tel Aviv Univ, Sackler Fac Med, IL-69978 Tel Aviv, Israel
[4] Ben Gurion Univ Negev, Fac Med, IL-84105 Beer Sheva, Israel
关键词:
Scheimpflug camera;
Equivalent keratometry reading;
Intraocular lens power calculation;
Cataract surgery;
Pentacam;
Corneal refractive surgery;
LENS POWER CALCULATION;
MYOPIC PHOTOREFRACTIVE KERATECTOMY;
REQUIRING CATARACT-SURGERY;
AXIAL LENGTH;
EYES;
PENTACAM;
FORMULA;
ADJUSTMENT;
D O I:
10.1159/000363140
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Background and Objective: To identify the most accurate combination of Pentacam's equivalent keratometry readings (EKR) and intraocular lens power formula when the clinical history is unavailable. Patients and Methods: A total of 18 patients underwent cataract surgery after refractive surgery. The Pentacam 4.5- and 3.0-mm EKR were combined with the SRK II, SRK/T, Hoffer-Q, and Holladay I and II formulas. Results: The smallest deviation from the predicted value was achieved by combining the 4.5 EKR with the Holladay II formula (mean arithmetic deviation, -0.2 +/- 0.4 dpt). Conclusion: The 4.5-mm EKR + Holladay II formula can accurately calculate intraocular lens power in patients with previous refractive surgery. (C) 2014 S. Karger AG, Basel
引用
收藏
页码:217 / 223
页数:7
相关论文
共 26 条