The effect of surgical alignment and soft tissue conditions on the kinematics and wear of a fixed bearing total knee replacement

被引:23
作者
Johnston, Helena [1 ]
Abdelgaied, Abdellatif [1 ]
Pandit, Hemant [2 ]
Fisher, John [1 ]
Jennings, Louise M. [1 ]
机构
[1] Univ Leeds, Inst Med & Biol Engn, Leeds, W Yorkshire, England
[2] Univ Leeds, Fac Med & Hlth, Leeds Inst Rheumat & Musculoskeletal Med, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Total knee replacement; Wear; Surgical alignment; Kinematics; Simulation; Contact mechanics; COMPUTER-ASSISTED NAVIGATION; ANTERIOR CRUCIATE LIGAMENT; TIBIAL COMPONENT; POLYETHYLENE WEAR; FEMORAL ROLLBACK; Q-ANGLE; ARTHROPLASTY; MALALIGNMENT; FAILURE; INSTRUMENTATION;
D O I
10.1016/j.jmbbm.2019.103386
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As life expectancy and activity levels of patients increase so does the demand on total knee replacements (TKRs). Abnormal mechanics and wear of TKRs can lead to implant loosening and revision. Component alignment after surgery varies due to the presurgical alignment, the accuracy of the surgical instrumentation and due to patient factors, such as the soft tissue balance. This study experimentally investigated the effect of variation in component alignment and the soft tissue conditions on the kinematics and wear of a fixed bearing TKR. DePuy Sigma fixed bearing TKRs with moderately cross-linked UHMWPE were used. Different alignment conditions were simulated in the coronal, sagittal and transverse planes in an ISO force-controlled simulation system. Three different soft tissue conditions were simulated using virtual springs to represent a stiff knee, a preserved PCL and a resected PCL. Four different alignment conditions were studied; ideal alignment, 4 degrees tibial and femoral varus joint line, 14 degrees rotational mismatch and 10 degrees posterior tibial slope. The varus joint line alignment resulted in similar kinematics and lower wear rate compared to ideal alignment. The rotational mismatch alignment resulted in significantly higher tibial rotation and abduction-adduction as well as a significantly higher wear rate than ideal alignment. The posterior tibial slope alignment resulted in significantly higher wear than the ideal alignment and dislocated under the lower tension soft tissue conditions. Component alignment and the soft tissue conditions had a significant effect on the kinematics and wear of the TKR investigated in this study. The surgical alignment of the TKR is an important factor in the clinical outcome of the joint as factors such as increased tibial rotation can lead to anterior knee pain and instability and increased wear can lead to aseptic loosening and early failure resulting in revision.
引用
收藏
页数:11
相关论文
共 70 条
  • [1] A comparison between electromechanical and pneumatic-controlled knee simulators for the investigation of wear of total knee replacements
    Abdelgaied, Abdellatif
    Fisher, John
    Jennings, Louise M.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 237 (07) : 643 - 651
  • [2] The effect of insert conformity and material on total knee replacement wear
    Abdelgaied, Abdellatif
    Brockett, Claire L.
    Liu, Feng
    Jennings, Louise M.
    Jin, Zhongmin
    Fisher, John
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (01) : 98 - 106
  • [3] Knee adduction moment and development of chronic knee pain in elders
    Amin, S
    Luepongsak, N
    McGibbon, CA
    LaValley, MP
    Krebs, DE
    Felson, DT
    [J]. ARTHRITIS & RHEUMATISM-ARTHRITIS CARE & RESEARCH, 2004, 51 (03): : 371 - 376
  • [4] Computer assisted navigation in total knee arthroplasty - Comparison with conventional methods
    Anderson, KC
    Buehler, KC
    Markel, DC
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (07) : 132 - 138
  • [5] [Anonymous], 2014, IMPLANTS SURG WEAR 3
  • [6] [Anonymous], 2009, IMPLANTS SURG WEAR 1
  • [7] Babazadeh P, 2009, ORTHOP REV, V1
  • [8] Making sense of knee arthroplasty kinematics: News you can use
    Banks, SA
    Harman, MK
    Bellemans, J
    Hodge, WA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A : 64 - 72
  • [9] Investigation of wear of knee prostheses in a new displacement/force-controlled simulator
    Barnett, PI
    McEwen, HMJ
    Auger, DD
    Stone, MH
    Ingham, E
    Fisher, J
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H1) : 51 - 61
  • [10] Comparison of wear in a total knee replacement under different kinematic conditions
    Barnett, PI
    Fisher, J
    Auger, DD
    Stone, MH
    Ingham, E
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (10-12) : 1039 - 1042