Comparison of Lesion Detection in the Transverse and Coronal Views on Automated Breast Sonography

被引:17
作者
Chae, Eun Young [1 ]
Cha, Joo Hee [1 ]
Kim, Hak Hee [1 ]
Shin, Hee Jung [1 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiol,Res Inst Radiol, Seoul 138736, South Korea
关键词
automated breast sonography; automated sonography; breast; breast ultrasound; coronal view; BI-RADS; INTEROBSERVER AGREEMENT; ULTRASOUND; MAMMOGRAPHY; CANCER; US; MASSES; VARIABILITY; DIAGNOSIS; LEXICON;
D O I
10.7863/ultra.34.1.125
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives The purpose of our study was to compare the accuracy of detection of lesions in the coronal and transverse views on automated breast sonography. Methods Three breast radiologists independently interpreted automated breast sonograms from 113 women: 14 with negative findings and 99 with known breast lesions (99 benign and 53 malignant findings). The readers were asked to detect the presence or absence of the abnormalities using the transverse and coronal views in different reading sessions. If a lesion was detected, we evaluated its location, sonographic features, including shape, orientation, margin, boundary type, and echo pattern, and final American College of Radiology Breast Imaging Reporting and Data System assessment category. The reading time was also evaluated. Results- Of 456 potential detections (152 lesions with 3 readers), 80.5% of the detections were made with the transverse view, and 67.3% were made with the coronal view. The detection rates for both malignant and benign lesions were significantly higher for the transverse view than for the coronal view (P <.001). The reading time for the coronal view was less than that for the transverse view (mean +/- SD, 3.83 +/- 1.71 versus 5.57 +/- 2.21 minutes). The intraclass correlation coefficient values for lesion location, distance from the nipple, and lesion size for both views all exceeded 0.7. Conclusions Although the coronal view on automated breast sonography showed comparable reliability and a shorter reading time, the detection rate for the coronal view was significantly lower than that for the transverse view in both benign and malignant lesions. It is preferable to reserve the coronal view as an additional method to the transverse view.
引用
收藏
页码:125 / 135
页数:11
相关论文
共 21 条
[1]   Breast Imaging Reporting and Data System Lexicon for US: Interobserver Agreement for Assessment of Breast Masses [J].
Abdullah, Nouf ;
Mesurolle, Benoit ;
El-Khoury, Mona ;
Kao, Ellen .
RADIOLOGY, 2009, 252 (03) :665-672
[2]   Combined screening with ultrasound and mammography vs mammography alone in women at elevated risk of breast cancer [J].
Berg, Wendie A. ;
Blume, Jeffrey D. ;
Cormack, Jean B. ;
Mendelson, Ellen B. ;
Lehrer, Daniel ;
Bohm-Velez, Marcela ;
Pisano, Etta D. ;
Jong, Roberta A. ;
Evans, W. Phil ;
Morton, Marilyn J. ;
Mahoney, Mary C. ;
Larsen, Linda Hovanessian ;
Barr, Richard G. ;
Farria, Dione M. ;
Marques, Helga S. ;
Boparai, Karan .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2008, 299 (18) :2151-2163
[3]   Operator dependence of physician-performed whole-breast US: Lesion detection and characterization [J].
Berg, Wendie A. ;
Blume, Jeffrey D. ;
Cormack, Jean B. ;
Mendelson, Ellen B. .
RADIOLOGY, 2006, 241 (02) :355-365
[4]   Incidental findings on sonography of the breast:: Clinical significance and diagnostic workup [J].
Buchberger, W ;
DeKoekkoek-Doll, P ;
Springer, P ;
Obrist, P ;
Dünser, M .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (04) :921-927
[5]   Breast cancers initially detected by hand-held ultrasound: detection performance of radiologists using automated breast ultrasound data [J].
Chang, Jung Min ;
Moon, Woo Kyung ;
Cho, Nariya ;
Park, Jeong Seon ;
Kim, Seung Ja .
ACTA RADIOLOGICA, 2011, 52 (01) :8-14
[6]   Differentiating benign from malignant solid breast masses: Comparison of two-dimensional and three-dimensional US [J].
Cho, Nariya ;
Moon, Woo Kyung ;
Cha, Joo Hee ;
Kim, Sun Mi ;
Han, Boo-Kyung ;
Kim, Eun-Kyung ;
Kim, Mi Hye ;
Chung, Soo Young ;
Choi, Hye-Young ;
Im, Jung-Gi .
RADIOLOGY, 2006, 240 (01) :26-32
[7]   Breast cancer detection using automated whole breast ultrasound and mammography in radiographically dense breasts [J].
Kelly, Kevin M. ;
Dean, Judy ;
Comulada, W. Scott ;
Lee, Sung-Jae .
EUROPEAN RADIOLOGY, 2010, 20 (03) :734-742
[8]   Comparison of conventional and automated breast volume ultrasound in the description and characterization of solid breast masses based on BI-RADS features [J].
Kim, Hyunji ;
Cha, Joo Hee ;
Oh, Ha-Yeun ;
Kim, Hak Hee ;
Shin, Hee Jung ;
Chae, Eun Young .
BREAST CANCER, 2014, 21 (04) :423-428
[9]   The Clinical Utility of Automated Breast Volume Scanner: A Pilot Study of 139 Cases [J].
Kim, Young Wook ;
Kim, Soon Kwang ;
Youn, Hyun Jo ;
Choi, Eun Jung ;
Jung, Sung Hoo .
JOURNAL OF BREAST CANCER, 2013, 16 (03) :329-334
[10]   Breast-cancer screening with ultrasonography [J].
Kopans, DB .
LANCET, 1999, 354 (9196) :2096-2097