Artificial iris implantation in various iris defects and lens conditions

被引:49
作者
Mayer, Christian [1 ]
Tandogan, Tamer [2 ,3 ]
Hoffmann, Andrea E. [1 ]
Khoramnia, Ramin [2 ,3 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Eye Clin, Munich, Germany
[2] Heidelberg Univ, David J Apple Int Lab Ocular Pathol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
[3] Heidelberg Univ, Int Vis Correct Res Ctr, Dept Ophthalmol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
关键词
INTRAOCULAR-LENS; POSTTRAUMATIC ANIRIDIA; CATARACT;
D O I
10.1016/j.jcrs.2017.06.003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We describe 6 surgical techniques used to implant a silicone iris prosthesis: sector-shaped iris segments that require suturing, injector-assisted sulcus fixation, injector-assisted capsular bag fixation with an intraocular lens (IOL) and capsular tension ring, folded iris tissue implanted with a forceps and sutured to the sclera with a scleral-fixated 10L, "sandwich" or "backpack" implantation with an IOL, and open-sky implantation with a perforating keratoplasty. The results of the techniques performed in 51 patients are discussed. Sector-shaped iris segments required longer surgery because of the complexity of intracameral sutures and Carried risks for knots to loosen and sutures to cut through residual iris tissue. The combined implantation of an iris and IOL proved complex and lengthy but solved lens and iris abnormalities in 1 procedure and provided long-lasting stable conditions. The procedures that implanted a complete iris in pseudophakic eyes were shorter and required smaller incisions, a sutureless approach, and injector-assisted implantation.
引用
收藏
页码:724 / 731
页数:8
相关论文
共 18 条
  • [1] Beekhuis WH, 1998, CORNEA, V17, P338, DOI 10.1097/00003226-199805000-00017
  • [2] Iris sutures
    Brauweiler, P.
    [J]. OPHTHALMOLOGE, 2011, 108 (08): : 710 - 713
  • [3] Burger D S, 1993, J Am Optom Assoc, V64, P176
  • [4] Prosthetic iris implantation for congenital, traumatic, or functional iris deficiencies
    Burk, SE
    Da Mata, AP
    Snyder, ME
    Cionni, RJ
    Cohen, JS
    Osher, RH
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) : 1732 - 1740
  • [5] Dalens H, 1980, Bull Soc Ophtalmol Fr, V80, P741
  • [6] Visual acuity in children with coloboma - Clinical features and a new phenotypic classification system
    Hornby, SJ
    Adolph, S
    Gilbert, CE
    Dandona, L
    Foster, A
    [J]. OPHTHALMOLOGY, 2000, 107 (03) : 511 - 520
  • [7] Intraocular lens implantation for patients with coloboma of the iris
    Li, Juanjuan
    Li, Yan
    Hu, Zhulin
    Kong, Lei
    [J]. EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2014, 7 (06) : 1595 - 1598
  • [8] Safety of an artificial iris in a phakic eye
    Magnus, Jessy
    Trau, Rene
    Mathysen, Danny G. P.
    Tassignon, Marie-Jose
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (06) : 1097 - 1100
  • [9] Surgical management of iris defects with prosthetic iris devices
    Mavrikakis, I
    Mavrikakis, E
    Syam, PP
    Bell, J
    Casey, JH
    Casswell, AG
    Brittain, GP
    Liu, C
    [J]. EYE, 2005, 19 (02) : 205 - 209
  • [10] Phacoemulsification and endocapsular implantation of an artificial iris intraocular lens in traumatic cataract and aniridia
    Mavrikakis, I
    Casey, JMH