Effect of Window and Hole Pattern Cut-Outs on Design Optimization of 3D Printed Braces

被引:0
|
作者
Rizza, Robert [1 ]
Liu, XueCheng [2 ,3 ,4 ]
Anewenter, Vince [5 ]
机构
[1] Milwaukee Sch Engn, Dept Mech Engn, Milwaukee, WI 53202 USA
[2] Dept Orthopaed Surg, Milwaukee, WI USA
[3] Childrens Hosp Wisconsin, Musculoskeletal Funct Assessment Ctr, Milwaukee, WI USA
[4] Med Coll Wisconsin, Dept Orthopaed Surg, Milwaukee, WI USA
[5] Rapid Prototyping Ctr, Milwaukee Sch Engn, Milwaukee, WI USA
来源
FRONTIERS IN REHABILITATION SCIENCES | 2022年 / 3卷
关键词
3D printing; scoliosis braces; hole patterns; cut-outs; brace design; SCOLIOSIS; SIMULATION;
D O I
10.3389/fresc.2022.889905
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
BackgroundThere are many different Thoracic Lumbar Sacral Orthosis style brace designs available in the market for the correction of scoliosis deformity. Hole cut out patterns, are commonly used in brace designs. These cut-outs may be subdivided into two groups: hole patterns and windows. Hole patterns are an array of holes which are implemented to lighten the weight of a brace and allow for the skin to breathe. Windows provide space for spinal derotation and/or breathing. From an examination of the literature, it appears that a systematic analysis of the effect of these cut-outs on the structural integrity and functionality of the brace has not been undertaken. Furthermore, there is a lack of understanding on the effect of spacing, size and geometry of the cut-outs on the mechanical behavior of the brace. Method of ApproachIn this study, Finite Element Analysis is employed to examine the mechanical response of the brace to these cut-outs. Geometry for the Thoracic Lumbar Sacral Orthosis was obtained by scanning an existing brace using an optical scan and converted into a Computer Aided Design model. A systematic approach was undertaken where cut-out geometry, spacing and size was varied. The deformation and stress in the thickness of the brace was ascertained from the Finite Element Analysis. An appropriate factor of safety for the structural analysis was determined using a standardized approach and used to quantify the structural integrity of the brace due to the cut-out. Various geometries were analyzed for the hole patterns including circle, triangle, diamond, and hexagon. For the window, the geometries considered were circle, trapezoidal and the "bib" geometry. ResultsIt was found that linear hole patterns where the holes are aligned do not provide a desirable structural factor safety. Furthermore, among all the possible geometries, the hexagonal cut-out was the best structurally while reducing the weight of the brace the most. The optimal spacing was found to be 12 mm, and the optimal hole surface area was found to be 78.54 mm(2). For the windows in the abdominal area, the "bib" shape provided the best structural integrity and generated the lowest amount of deformation. An increase in the size of this window had a small effect on the stress but an almost negligible effect on the deformation. ConclusionsA hexagonal hole pattern should be used with a spacing of 12 mm and each hole should have a surface area of 78.54 mm(2). Windows in the abdominal area should be of "bib" shape. The size of the window cut-outs does not affect the brace stress and deformation significantly. Thus, the size of these windows should be based on the functional aspects of the brace, i.e., the minimum required size needed to permit the patient to breathe comfortably as in the case of the abdominal window or to allow for proper derotation, as in the case of the derotation window.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Computational optimization of a 3D printed collimator
    Islam, Fahima
    Lin, Jiao
    Huegle, Thomas
    Lumsden, Ian
    Anderson, David
    Elliott, Amy
    Haberl, Bianca
    Granroth, Garrett
    JOURNAL OF NEUTRON RESEARCH, 2020, 22 (2-3) : 155 - 168
  • [2] Design, optimization, simulation, and implementation of a 3D printed soft robotic peristaltic pump
    Totuk, Onat Halis
    Mistikoglu, Selcuk
    Guvenc, Mehmet Ali
    ENGINEERING RESEARCH EXPRESS, 2024, 6 (04):
  • [3] Production of patient-specific electron beam aperture cut-outs using a low-cost, multi-purpose 3D printer
    Michiels, Steven
    Mangelschots, Bram
    De Roover, Robin
    Devroye, Cedric
    Depuydt, Tom
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 756 - 760
  • [4] A 3D visualization tool for the design and customization of spinal braces
    Fortin, D.
    Cheriet, F.
    Beausejour, M.
    Debanne, P.
    Joncas, J.
    Labelle, H.
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2007, 31 (08) : 614 - 624
  • [5] Effect of fiber reinforcement on the open hole tensile strength of 3D printed composites
    Prajapati, Ashish R.
    Dave, Harshit K.
    Raval, Harit K.
    MATERIALS TODAY-PROCEEDINGS, 2021, 46 : 8629 - 8633
  • [6] Design, evaluation, and optimization of 3D printed truss scaffolds for bone tissue engineering
    Shirzad, M.
    Zolfagharian, A.
    Matbouei, A.
    Bodaghi, M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 120
  • [7] 3D Printed Fabric: Techniques for Design and 3D Weaving Programmable Textiles
    Takahashi, Haruki
    Kim, Jeeeun
    PROCEEDINGS OF THE 32ND ANNUAL ACM SYMPOSIUM ON USER INTERFACE SOFTWARE AND TECHNOLOGY (UIST 2019), 2019, : 43 - 51
  • [8] Open Hole Tension of 3D Printed Aligned Discontinuous Composites
    Krajangsawasdi, Narongkorn
    Hamerton, Ian
    Woods, Benjamin K. S.
    Ivanov, Dmitry S.
    Longana, Marco L.
    MATERIALS, 2022, 15 (23)
  • [9] Design of a 3D printed concrete bridge by testing
    Salet, Theo A. M.
    Ahmed, Zeeshan Y.
    Bos, Freek P.
    Laagland, Hans L. M.
    VIRTUAL AND PHYSICAL PROTOTYPING, 2018, 13 (03) : 222 - 236
  • [10] Exploring Design Approaches for 3D Printed Antennas
    Carvalho, Saul S.
    Reis, Joao R. V.
    Mateus, Artur
    Caldeirinha, Rafael F. S.
    IEEE ACCESS, 2024, 12 : 10718 - 10735