Absolute calibration of ACIS X-ray CCDs using calculable, undispersed synchrotron radiation

被引:15
作者
Bautz, MW [1 ]
Pivovaroff, MJ [1 ]
Kissel, SE [1 ]
Prigozhin, GY [1 ]
Isobe, T [1 ]
Jones, SE [1 ]
Ricker, GR [1 ]
Thornagel, R [1 ]
Kraft, S [1 ]
Scholze, F [1 ]
Ulm, G [1 ]
机构
[1] MIT, Ctr Space Res, Cambridge, MA 02139 USA
来源
X-RAY OPTICS, INSTRUMENTS, AND MISSIONS III | 2000年 / 4012卷
关键词
CCDs; X-rays; ACIS; Chandra;
D O I
10.1117/12.391582
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
We have performed a series of measurements with the Physikalisch-Technische Bundesanstalt beamline of the electron storage ring BESSY I which provide the basis for the absolute calibration of the Advanced CCD Imaging Spectrometer (ACIS). ACIS is a prime focal plane instrument aboard the recently-launched Chandra X-ray Observatory. We have achieved an absolute detection efficiency knowledge accurate to better than 5% over the 0.3 - 4 keV band. We describe our measurement and analysis techniques, including our detector response modeling and pileup corrections. We summarize a variety of external and internal consistency checks which provide the basis for our error estimates. We discuss the factors limiting the accuracy of our measurements.
引用
收藏
页码:53 / 67
页数:15
相关论文
共 20 条
[1]  
Arnaud KA, 1996, ASTR SOC P, V101, P17
[2]  
ARNOLD G, 1992, REV SCI INSTRUM, V63, P1
[3]   X-ray CCD calibration for the AXAF CCD imaging spectrometer [J].
Bautz, M ;
Pivovaroff, M ;
Baganoff, F ;
Isobe, T ;
Jones, S ;
Kissel, S ;
LaMarr, B ;
Manning, H ;
Prigozhin, G ;
Ricker, G ;
Nousek, J ;
Grant, C ;
Nishikida, K ;
Scholze, F ;
Thornagel, R ;
Ulm, G .
X-RAY OPTICS, INSTRUMENTS, AND MISSIONS, 1998, 3444 :210-224
[4]   X-ray CCD calibration for the AXAF CCD imaging spectrometer [J].
Bautz, M ;
Kissel, S ;
Prigozhin, G ;
Jones, S ;
Isobe, T ;
Manning, H ;
Pivovaroff, M ;
Ricker, G ;
Woo, J .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY VII, 1996, 2808 :170-181
[5]   X-ray CCD response functions, front to back [J].
Bautz, MW ;
Prigozhin, GY ;
Pivovaroff, MJ ;
Jones, SE ;
Kissel, SE ;
Ricker, GR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 436 (1-2) :40-52
[6]   Soft-x-ray CCD imagers for AXAF [J].
Burke, BE ;
Gregory, JA ;
Bautz, MW ;
Prigozhin, GY ;
Kissel, SE ;
Kosicki, BB ;
Loomis, AH ;
Young, DJ .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1997, 44 (10) :1633-1642
[7]  
GARMIRE G, 2000, UNPUB
[8]  
Gendreau K.C., 1995, THESIS MIT
[9]   X-RAY INTERACTIONS - PHOTOABSORPTION, SCATTERING, TRANSMISSION, AND REFLECTION AT E=50-30,000 EV, Z=1-92 [J].
HENKE, BL ;
GULLIKSON, EM ;
DAVIS, JC .
ATOMIC DATA AND NUCLEAR DATA TABLES, 1993, 54 (02) :181-342
[10]  
JONES S, 2000, UNPUB