Extending the throughput of Biacore 4000 biosensor to accelerate kinetic analysis of antibody-antigen interaction

被引:16
|
作者
Kamat, Vishal [1 ]
Rafique, Ashique [1 ]
机构
[1] Regeneron Pharmaceut, Therapeut Prot, Biomol HTS Ctr, 777 Old Saw Mill River Rd, Tarrytown, NY 10591 USA
关键词
Biacore; Surface plasmon resonance; Label-free optical biosensor; Monoclonal antibody; High throughput; Antibody-antigen interaction; Binding kinetic; Affinity; A100;
D O I
10.1016/j.ab.2017.04.020
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The surface plasmon resonance (SPR) biosensors are being routinely used in different stages of drug discovery and development. However, the lack of high throughput SPR biosensors continues to be a primary bottleneck for the rapid kinetic screening of large panels of monoclonal antibodies (mAbs). To further increase the throughput of the Biacore 4000 biosensor, we have developed three kinetic screening assays to characterize mAb-antigen interactions (i) 16-mAb capture kinetic, (ii) single cycle kinetic (SCK), and (iii) parallel kinetic (PK). The performance of all three kinetic assays was evaluated by characterizing the binding of kinetically diverse human mAbs to four antigens with molecular weights of 1410, 29kD, 38kD, and 48kD and binding affinities ranging from 130pM to 200 nM. The binding rate constants measured using all three kinetic assays were reproducible across multiple experiments and correlated with the values generated using the conventional 8-mAb capture kinetic assay on the Biacore 4000 (R-2 > 0.94). Moreover, the 16-mAb capture assay decreased experiment time and analyte consumption by 35% and 50%, respectively. This work illustrates the significance of the 16-mAb capture kinetic, SCK, and PK assays to increase the throughput of Biacore 4000 and to support rapid kinetic screening of mAbs. (C) 2017 The Authors. Published by Elsevier Inc.
引用
收藏
页码:75 / 86
页数:12
相关论文
共 50 条
  • [31] Parameter estimation and identifiability analysis for a bivalent analyte model of monoclonal antibody-antigen binding
    Nguyen, Kyle
    Li, Kan
    Flores, Kevin
    Tomaras, Georgia D.
    Dennison, S. Moses
    McCarthy, Janice M.
    ANALYTICAL BIOCHEMISTRY, 2023, 679
  • [32] Structural aspects of antibody-antigen interaction revealed through small random peptide libraries
    Slootstra, JW
    Puijk, WC
    Ligtvoet, GJ
    Langeveld, JPM
    Meloen, RH
    MOLECULAR DIVERSITY, 1996, 1 (02) : 87 - 96
  • [33] Integration of antibody-antigen and receptor-ligand reactions to establish a gold strip biosensor for detection of 33 β-lactam antibiotics
    Li, Yue
    Liu, Liqiang
    Xu, Chuanlai
    Kuang, Hua
    Sun, Li
    SCIENCE CHINA-MATERIALS, 2021, 64 (08) : 2056 - 2066
  • [34] Kinetic analysis of a high-affinity antibody/antigen interaction performed by planar waveguide fluorescence immunosensor
    Guo, Hongli
    Zhou, Xiaohong
    Zhang, Yan
    Gu, Chunmei
    Song, Baodong
    Shi, Hanchang
    RSC ADVANCES, 2016, 6 (17): : 13837 - 13845
  • [35] A comparison of binding surfaces for SPR biosensing using an antibody-antigen system and affinity distribution analysis
    Zhao, Huaying
    Gorshkova, Inna I.
    Fu, Gregory L.
    Schuck, Peter
    METHODS, 2013, 59 (03) : 328 - 335
  • [36] Label-free optical detection of protein antibody-antigen interaction on an capped porous anodic alumina layer chip
    Kim, Do-Kyun
    Kerman, Kagan
    Yamamura, Shohei
    Kwon, Young-Soo
    Takamura, Yuzuru
    Tamiya, Eiichi
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2008, 47 (02) : 1351 - 1354
  • [37] Characterization of ELISA Antibody-Antigen Interaction using Footprinting-Mass Spectrometry and Negative Staining Transmission Electron Microscopy
    Lin, Margaret
    Krawitz, Denise
    Callahan, Matthew D.
    Deperalta, Galahad
    Wecksler, Aaron T.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2018, 29 (05) : 961 - 971
  • [38] Design and Simulation of the Microcantilever Biosensor for MITF Antigen and D5 Monoclonal Antibody Interaction Finite Element Analysis, and Experimental
    Akcali, Pelin
    Kelleci, Kubra
    Ozer, Sevil
    CURRENT PROTEIN & PEPTIDE SCIENCE, 2024, 25 (03) : 256 - 266
  • [39] AppA: a web server for analysis, comparison, and visualization of contact residues and interfacial waters of antibody-antigen structures and models
    Nguyen, Minh N.
    Verma, Chandra S.
    Zhong, Pingyu
    NUCLEIC ACIDS RESEARCH, 2019, 47 (W1) : W482 - W489
  • [40] Kinetic analysis of the interaction between HIV-1 protease and inhibitors using optical biosensor technology
    Markgren, PO
    Hämäläinen, M
    Danielson, UH
    ANALYTICAL BIOCHEMISTRY, 2000, 279 (01) : 71 - 78