Validation of reflex testing rules and establishment of a new workflow for body fluid cell analysis using a Sysmex XN-550 automatic hematology analyzer

被引:14
作者
Cho, Y-U [1 ,2 ]
You, E.
Jang, S.
Park, C-J
机构
[1] Univ Ulsan, Dept Lab Med, Coll Med, Seoul, South Korea
[2] Asan Med Ctr, Seoul, South Korea
关键词
body fluid; reflex testing rules; Sysmex XN-550; validation; workflow; CEREBROSPINAL-FLUID; ASCITIC FLUIDS; PERFORMANCE EVALUATION; LEUKOCYTE COUNTS; BLOOD-CELLS; MODE; XE-5000; DIFFERENTIATION; XN-1000;
D O I
10.1111/ijlh.12774
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: We developed and validated reflex testing rules for the microscopic examination (ME) of body fluids (BFs) on the Sysmex XN-550 (Sysmex Corporation) instrument. Methods: We assessed the detection limits, precision, linearity, and carryover. To develop the reflex testing rules (derivation arm), we tested 515 samples and then validated the rules using another 507 samples (validation arm). Results: All analytical performances were acceptable, and the carryover was negligible. There was agreement between the automated count and ME of red blood cells (r=.98) and total nucleated cells (TNCs) (r=.98), as well as the differential counts of neutrophils (r=.90) and lymphocytes (r=.84). We developed reflex testing rules: TNCs < 10/mu L, cell 2/cell 1 ratio <= 0.7, HF-BF cells > 7.9/100 white blood cells, LY-X >= 85 or LY-Y >= 90, and eosinophils > 2.5%. In the validation arm, implementation of the rules resulted in 126 rule-negative samples (24.9%) that were well correlated between the 2 methods. We propose a new workflow for BF cell analysis based on automated counting. Conclusion: The Sysmex XN-550 can be a suitable alternative to ME for BF cell analysis, especially for screening samples and subsequent automatic reporting under the rational use of laboratory-specific rules.
引用
收藏
页码:258 / 267
页数:10
相关论文
共 29 条
[1]  
[Anonymous], SYSMEX J INT
[2]  
[Anonymous], BODY FLUID ANAL
[3]  
Aulesa C, 2003, Lab Hematol, V9, P214
[4]   ICSH guidelines for the verification and performance of automated cell counters for body fluids [J].
Bourner, G. ;
De la Salle, B. ;
George, T. ;
Tabe, Y. ;
Baum, H. ;
Culp, N. ;
Keng, T. B. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (06) :598-612
[5]  
Brown Wendy, 2003, Lab Hematol, V9, P155
[6]   Reflex Testing Rules for Cell Count and Differentiation of Nucleated Elements in Pleural and Ascitic Fluids on Sysmex XE-5000 [J].
Buoro, Sabrina ;
Esposito, Sara Appassiti ;
Vavassori, Mauro ;
Mecca, Tommaso ;
Ottomano, Cosimo ;
Dominoni, Paola ;
Seghezzi, Michela ;
Candiago, Elisabetta ;
Farina, Claudio ;
Gianatti, Andrea ;
Crippa, Alberto ;
Lippi, Giuseppe .
JALA, 2016, 21 (02) :297-304
[7]   Cell Population Data and reflex testing rules of cell analysis in pleural and ascitic fluids using body fluid mode on Sysmex XN-9000 [J].
Buoro, Sabrina ;
Mecca, Tommaso ;
Azzara, Giovanna ;
Seghezzi, Michela ;
Dominoni, Paola ;
Crippa, Alberto ;
Ottomano, Cosimo ;
Lippi, Giuseppe .
CLINICA CHIMICA ACTA, 2016, 452 :92-98
[8]   Body fluid cellular analysis using the Sysmex XN-2000 automatic hematology analyzer: focusing on malignant samples [J].
Cho, Y. -U. ;
Chi, H. -S. ;
Park, S. H. ;
Jang, S. ;
Kim, Y. -J. ;
Park, C. -J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (03) :346-356
[9]   Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100 [J].
de Jonge, Robert ;
Brouwer, Rob ;
van Rijn, Marcel ;
van Acker, Bernadette A. C. ;
Otten, Henk J. A. M. ;
Lindemans, Jan .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (11) :1367-1371
[10]   Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids [J].
de Jonge, Robert ;
Brouwer, Rob ;
de Graaf, Marieke T. ;
Luitwieler, Ronald L. ;
Fleming, Cherina ;
de Frankrijker-Merkestijn, Magda ;
Smitt, Peter A. E. Sillevis ;
Boonstra, Joke G. ;
Lindemans, Jan .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (05) :665-675