Automated segmentation of an intensity calibration phantom in clinical CT images using a convolutional neural network

被引:10
作者
Uemura, Keisuke [1 ,2 ]
Otake, Yoshito [1 ]
Takao, Masaki [3 ]
Soufi, Mazen [1 ]
Kawasaki, Akihiro [1 ]
Sugano, Nobuhiko [2 ]
Sato, Yoshinobu [1 ]
机构
[1] Nara Inst Sci & Technol, Div Informat Sci, Grad Sch Sci & Technol, Ikoma, Nara, Japan
[2] Osaka Univ, Grad Sch Med, Dept Orthopaed Med Engn, Suita, Osaka, Japan
[3] Osaka Univ, Dept Orthopaed, Grad Sch Med, Suita, Osaka, Japan
基金
日本学术振兴会;
关键词
Artificial intelligence; Bone mineral density; Deep learning; Quantitative computed tomography; Phantom segmentation; U-net; BONE; ASSOCIATION; STRENGTH; DENSITY;
D O I
10.1007/s11548-021-02345-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose In quantitative computed tomography (CT), manual selection of the intensity calibration phantom's region of interest is necessary for calculating density (mg/cm(3)) from the radiodensity values (Hounsfield units: HU). However, as this manual process requires effort and time, the purposes of this study were to develop a system that applies a convolutional neural network (CNN) to automatically segment intensity calibration phantom regions in CT images and to test the system in a large cohort to evaluate its robustness. Methods This cross-sectional, retrospective study included 1040 cases (520 each from two institutions) in which an intensity calibration phantom (B-MAS200, Kyoto Kagaku, Kyoto, Japan) was used. A training dataset was created by manually segmenting the phantom regions for 40 cases (20 cases for each institution). The CNN model's segmentation accuracy was assessed with the Dice coefficient, and the average symmetric surface distance was assessed through fourfold cross-validation. Further, absolute difference of HU was compared between manually and automatically segmented regions. The system was tested on the remaining 1000 cases. For each institution, linear regression was applied to calculate the correlation coefficients between HU and phantom density. Results The source code and the model used for phantom segmentation can be accessed at https://github.com/keisuke-uemura/CT-Intensity-Calibration-Phantom-Segmentation. The median Dice coefficient was 0.977, and the median average symmetric surface distance was 0.116 mm. The median absolute difference of the segmented regions between manual and automated segmentation was 0.114 HU. For the test cases, the median correlation coefficients were 0.9998 and 0.999 for the two institutions, with a minimum value of 0.9863. Conclusion The proposed CNN model successfully segmented the calibration phantom regions in CT images with excellent accuracy.
引用
收藏
页码:1855 / 1864
页数:10
相关论文
共 23 条
  • [21] Vertebral Bone Mineral Density Measured by Quantitative Computed Tomography With and Without a Calibration Phantom: A Comparison Between 2 Different Software Solutions
    Therkildsen, Josephine
    Thygesen, Jesper
    Winther, Simon
    Svensson, My
    Hauge, Ellen-Margrethe
    Bottcher, Morten
    Ivarsen, Per
    Jorgensen, Hanne Skou
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2018, 21 (03) : 367 - 374
  • [22] The distribution of bone mineral density in the femoral heads of unstable intertrochanteric fractures
    Uemura, Keisuke
    Takao, Masaki
    Otake, Yoshito
    Hamada, Hidetoshi
    Sakai, Takashi
    Sato, Yoshinobu
    Sugano, Nobuhiko
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY, 2018, 26 (02):
  • [23] A cross-sectional study on the age-related cortical and trabecular bone changes at the femoral head in elderly female hip fracture patients
    Whitmarsh, Tristan
    Otake, Yoshito
    Uemura, Keisuke
    Takao, Masaki
    Sugano, Nobuhiko
    Sato, Yoshinobu
    [J]. SCIENTIFIC REPORTS, 2019, 9 (1)